Metaphyseal anchoring short stem hip arthroplasty provides a more physiological load transfer: a comparative finite element analysis study

被引:29
作者
Yan, Shuang G. [1 ,2 ]
Chevalier, Yan [2 ]
Liu, Fanxiao [2 ,3 ]
Hua, Xingyi [1 ]
Schreiner, Anna [4 ]
Jansson, Volkmar [2 ]
Schmidutz, Florian [2 ,4 ]
机构
[1] Anhui Med Univ, Affiliated Hosp 1, Dept Orthopaed Surg, 1 Baicao Rd, Hefei 230088, Peoples R China
[2] Univ Munich LMU, Dept Orthopaed Surg Phys Med & Rehabil, Munich, Germany
[3] Shandong First Med Univ, Shandong Prov Hosp, Dept Orthopaed, 324 Rd Jing Wu Wei Qi, Jinan 250021, Shandong, Peoples R China
[4] Eberhard Karls Univ Tuebingen, BG Trauma Ctr Tuebingen, Schnarrenbergstr 95, D-72076 Tubingen, Germany
关键词
Stress shielding; Finite element analysis; Total hip arthroplasty; SHA; Stem; METHA SHORT STEM; PRIMARY STABILITY; DXA; AGE;
D O I
10.1186/s13018-020-02027-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Short stem total hip arthroplasty (SHA) preserves femoral bone stock and is supposed to provide a more natural load transfer compared to standard stem total hip arthroplasty (THA). As comparative biomechanical reference data are rare we used a finite element analysis (FEA) approach to compare cortical load transfer after implantations of a metaphyseal anchoring short and standard stem in native biomechanical femora. Methods The subject specific finite element models of biomechanical femora, one native and two with implanted metaphyseal anchoring SHA (Metha, B. Braun Aesculap) and standard THA (CLS, Zimmer-Biomet), were generated from computed tomography datasets. The loading configuration was performed with an axial force of 1400 N. Von Mises stress was used to investigate the change of cortical stress distribution. Results Compared to the native femur, a considerable reduction of cortical stress was recorded after implantation of SHA and standard THA. The SHA showed less reduction proximally with a significant higher metaphyseal cortical stress compared to standard THA. Moreover, the highest peak stresses were observed metaphyseal for the SHA stem while for the standard THA high stress pattern was observed more distally. Conclusions Both, short and standard THA, cause unloading of the proximal femur. However, the metaphyseal anchoring SHA features a clearly favorable pattern in terms of a lower reduction proximally and improved metaphyseal loading, while standard THA shows a higher proximal unloading and more distal load transfer. These load patterns implicate a reduced stress shielding proximally for metaphyseal anchoring SHA stems and might be able to translate in a better bone preservation.
引用
收藏
页数:10
相关论文
共 50 条
[11]   'Pre-launch' finite element analysis of a short-stem total hip arthroplasty system consisting of two implant types [J].
Lerch, Matthias ;
Windhagen, Henning ;
Kurtz, Agnes-Elisabeth ;
Budde, Stefan ;
Behrens, Bernd-Arno ;
Bouguecha, Anas ;
Almohallami, Amer .
CLINICAL BIOMECHANICS, 2019, 61 :31-37
[12]   Comparison of two metaphyseal-fitting (short) femoral stems in primary total hip arthroplasty: study protocol for a prospective randomized clinical trial with additional biomechanical testing and finite element analysis [J].
Tatani, I. ;
Panagopoulos, A. ;
Diamantakos, I. ;
Sakellaropoulos, G. ;
Pantelakis, Sp ;
Megas, P. .
TRIALS, 2019, 20 (1)
[13]   Comparison of two metaphyseal-fitting (short) femoral stems in primary total hip arthroplasty: study protocol for a prospective randomized clinical trial with additional biomechanical testing and finite element analysis [J].
I. Tatani ;
A. Panagopoulos ;
I. Diamantakos ;
G. Sakellaropoulos ;
Sp Pantelakis ;
P. Megas .
Trials, 20
[14]   Effect of stem position and length on bone-stem constructs after cementless hip arthroplasty A FINITE ELEMENT ANALYSIS [J].
Kwak, D-K ;
Bang, S-H ;
Lee, S-J ;
Park, J-H ;
Yoo, J-H .
BONE & JOINT RESEARCH, 2021, 10 (04) :250-258
[15]   Development and Finite Element (FE) analysis of a novel short hip stem concept [J].
Jahnke, Alexander ;
Harz, Torben ;
Ulloa, Carlos Alfonso Fonseca ;
Ishaque, Bernd Alexander ;
Rickert, Markus .
JOURNAL OF ORTHOPAEDICS, 2023, 46 :117-123
[16]   Biomechanical study of the resurfacing hip arthroplasty - Finite element analysis of the femoral component [J].
Watanabe, Y ;
Shiba, N ;
Matsuo, S ;
Higuchi, F ;
Tagawa, Y ;
Inoue, A .
JOURNAL OF ARTHROPLASTY, 2000, 15 (04) :505-511
[17]   Biomechanical Behavior of an Hydroxyapatite-Coated Traditional Hip Stem and a Short One of Similar Design: Comparative Study Using Finite Element Analysis [J].
Gomez-Vallejo, Jesus ;
Roces-Garcia, Jorge ;
Moreta, Jesus ;
Donaire-Hoyas, Daniel ;
Gayoso, Oscar ;
Marques-Lopez, Fernando ;
Albareda, Jorge .
ARTHROPLASTY TODAY, 2021, 7 :167-176
[18]   Comparison of manual rasping and robotic milling for short metaphyseal-fitting stem implantation in total hip arthroplasty: a cadaveric study [J].
Lim, Seung-Jae ;
Kim, Sang-Min ;
Lim, Byung-Ho ;
Moon, Young-Wan ;
Park, Youn-Soo .
COMPUTER AIDED SURGERY, 2013, 18 (1-2) :33-40
[19]   Stress distributions of the short stem and the tapered wedge stem at different alignments: a finite element analysis study [J].
Kaku, Nobuhiro ;
Pramudita, Jonas A. ;
Yamamoto, Kansei ;
Hosoyama, Tsuguaki ;
Tsumura, Hiroshi .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
[20]   Influence of proximal stem geometry and stem-cement interface characteristics on bone and cement stresses in femoral hip arthroplasty: finite element analysis [J].
Massin, P ;
Astoin, E ;
Lavaste, F .
REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2003, 89 (02) :134-143