Performance of experimental sample injectors for high-performance liquid chromatography microcolumns

被引:26
作者
Foster, MD [1 ]
Arnold, MA [1 ]
Nichols, JA [1 ]
Bakalyar, SR [1 ]
机构
[1] Rheodyne LP, Rohnert Park, CA 94927 USA
关键词
injection techniques; sample volume; injector dispersion; injector performance; instrumentation;
D O I
10.1016/S0021-9673(99)00957-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An experimental injector for HPLC microcolumns and a 3-nl conductivity detector connected directly to the injector outlet with a 19-nl tube were used to study injector dispersion, guide the design of improved injectors, and suggest appropriate injection techniques. With regard to the small injection volumes required when no on-column concentration technique is used, we show that in some circumstances: (i) there are two volumes to be considered, the sample volume (that which is intended to be injected) and the effective injection volume (that which contains all the sample after it has completely emerged from the injector). Due to dispersion, the latter is often many times the former. An injector performance factor is defined as the ratio of the two volumes. (ii) A smaller sample chamber volume in an injector does not necessarily produce a proportionately smaller effective injection volume, in which case there is a reduction of peak height that degrades sensitivity without a commensurate reduction in peak width that would improve resolution. (iii) Adjusting the geometry of the sample chamber and stator passage can significantly improve injector performance, as illustrated for sample volumes from 2 nl to 1 pi. (iv) In some cases, reducing the diameter of an injector passageway in an attempt to reduce dispersion actually causes performance to worsen. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:231 / 241
页数:11
相关论文
共 24 条
[1]  
Abian J, 1999, J MASS SPECTROM, V34, P244, DOI 10.1002/(SICI)1096-9888(199904)34:4<244::AID-JMS775>3.0.CO
[2]  
2-0
[3]   Choosing sample volume to achieve maximum detection sensitivity and resolution with high-performance liquid chromatography columns of 1.0, 2.1 and 4.6 mm ID [J].
Bakalyar, SR ;
Phipps, C ;
Spruce, B ;
Olsen, K .
JOURNAL OF CHROMATOGRAPHY A, 1997, 762 (1-2) :167-185
[4]   Instrumental requirements for nanoscale liquid chromatography [J].
Chervet, JP ;
Ursem, M ;
Salzmann, JB .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1507-1512
[5]  
CHERVET JP, 1992, LC GC EUR, V5, P33
[6]   ANALYSIS OF PEPTIDE MIXTURES BY CAPILLARY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - A PRACTICAL GUIDE TO SMALL-SCALE SEPARATIONS [J].
DAVIS, MT ;
LEE, TD .
PROTEIN SCIENCE, 1992, 1 (07) :935-944
[7]   A NEW VALVE FOR ZONE-ELECTROPHORETIC SAMPLE TREATMENT COUPLED ONLINE WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
DEBETS, AJJ ;
HUPE, KP ;
BRINKMAN, UAT ;
KOK, WT .
CHROMATOGRAPHIA, 1990, 29 (5-6) :217-222
[8]  
Golay M.J. E., 1958, GAS CHROMATOGR INT S, P36
[9]   EARLY PHASES OF THE DISPERSION OF A SAMPLE INJECTED IN POISEUILLE FLOW [J].
GOLAY, MJE ;
ATWOOD, JG .
JOURNAL OF CHROMATOGRAPHY, 1979, 186 (DEC) :353-370
[10]   Comparison of the responses of triacylglycerols with an evaporative visible light scattering detector used in conventional, micro and capillary liquid chromatography [J].
Héron, S ;
Tchapla, A .
JOURNAL OF CHROMATOGRAPHY A, 1999, 848 (1-2) :95-104