Improvement of the long-term stability of polyimide-coated fused-silica capillaries used in capillary electrophoresis and capillary electrochromatography

被引:37
作者
Baeuml, F [1 ]
Welsch, T [1 ]
机构
[1] Univ Ulm, Dept Analyt & Environm Chem, D-89081 Ulm, Germany
关键词
stability studies; coated capillaries; polyimide coating; buffer composition; thiourea;
D O I
10.1016/S0021-9673(02)00048-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Swelling of the polyimide coating of fused-silica capillaries in acetonitrile-containing buffers was found to be the reason for several problems in capillary electrophoresis (CE) and capillary electrochromatography (CEC). Scanning electron microscopy photographs of the ends of raw fused-silica tubing showed that the coating becomes soft and increases its volume after longer contact with such buffers. As a consequence, separation efficiency can deteriorate, the capillary ends can clog or break off. To prevent swelling of the polyimide coating, fused-silica capillaries used in CE or CEC were heated at 300 degreesC for a longer period of time which improved their long-term stability in comparison to raw fused-silica tubing. (C) 2002 Elsevier Science BV All rights reserved.
引用
收藏
页码:35 / 44
页数:10
相关论文
共 11 条
[1]  
BESSONOV MI, 1987, POLYIMIDES THERMALLY, P97
[2]   INFLUENCE OF THE CAPILLARY EDGE ON THE SEPARATION EFFICIENCY IN CAPILLARY ELECTROPHORESIS [J].
COHEN, N ;
GRUSHKA, E .
JOURNAL OF CHROMATOGRAPHY A, 1994, 684 (02) :323-328
[3]  
Dittmann MM, 1995, LC GC-MAG SEP SCI, V13, P800
[4]  
Dittmann MM, 1997, J CAPILLARY ELECTROP, V4, P201
[5]   ARTIFACTS RELATED TO SAMPLE INTRODUCTION IN CAPILLARY GEL-ELECTROPHORESIS AFFECTING SEPARATION PERFORMANCE AND QUANTITATION [J].
GUTTMAN, A ;
SCHWARTZ, HE .
ANALYTICAL CHEMISTRY, 1995, 67 (13) :2279-2283
[6]   Moderation of the electroosmotic flow in capillary electrophoresis by chemical modification of the capillary surface with tentacle-like oligourethanes [J].
König, S ;
Welsch, T .
JOURNAL OF CHROMATOGRAPHY A, 2000, 894 (1-2) :79-88
[7]  
PAWLOWSKI WJ, 1992, POLYMER AND FIBER SCIENCE : RECENT ADVANCES, P29
[8]  
ROZING G, 1998, INT LAB NEWS, V28, pC8
[9]  
Schmid M, 1999, HRC-J HIGH RES CHROM, V22, P438
[10]   Separation of selected peptides by capillary electroendoosmotic chromatography using 3 μm reversed-phase bonded silica and mixed-mode phases [J].
Walhagen, K ;
Unger, KK ;
Olsson, AM ;
Hearn, MTW .
JOURNAL OF CHROMATOGRAPHY A, 1999, 853 (1-2) :263-275