In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair

被引:22
|
作者
Torsello, Giovanni Federico [1 ,2 ]
Herten, Monika [2 ,3 ]
Mueller, Markus [4 ]
Andre-Frank [4 ]
Torsello, Giovanni B. [2 ]
Austermann, Martin [2 ]
机构
[1] Charite Univ Med Berlin, Charite Campus Virchow Klinikum, Dept Diagnost & Intervent Radiol & Nucl Med, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] St Franziskus Hosp Munster, Dept Vasc Surg, Munster, Germany
[3] Essen Univ Hosp, Dept Orthoped & Trauma Surg, Essen, Germany
[4] Univ Hosp Munster, Dept Hand Trauma & Reconstruct Surg, Biomech Lab, Munster, Germany
关键词
bridging stent-graft; displacement force; fenestrated stent-graft; in vitro experiment; in vitro model; material fatigue; permeability; stent fracture; ENDOGRAFTS; COMPONENTS; OUTCOMES; SECURE;
D O I
10.1177/1526602819842569
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To evaluate the safety and integrity of a new stent-graft as a potential bridging device in fenestrated stent-grafts using an in vitro fenestrated model. Materials and Methods: Polyester test sheets with ten 6-mm- or 8-mm-diameter fenestrations were used to simulate a fenestrated main body endoprosthesis. In total, 50 Viabahn balloon-expandable (VBX) stent-grafts of varying lengths (29 and 39 mm) and diameters (6, 7, and 8 mm) were implanted in the fitting fenestrations. After release, the 6- and 7-mm-diameter stent-grafts were flared with a 10x20-mm angioplasty balloon; a 12x20-mm balloon was used in the 8-mm-diameter devices. Safety of the devices was defined as absence of fractures detected on radiography or computed tomography (CT), as well as material failure detected by microscopy and water permeability testing. The forces (in Newtons, N) needed for perpendicular dislocation (pullout force) and axial dislocation (shear stress force) were also evaluated. Results: Forty VBX stent-grafts were subjected to digital radiographic imaging and multiplanar CT. None showed any stent fracture. Subsequent microscopy indicated no damage to the fabric or separation of the graft after flaring. Ten VBX stent-grafts underwent water permeability testing after flaring; no water passed through the graft wall during a 10-minute period under an intraluminal pressure at 120 mm Hg. Testing of 25 VBX stent-grafts revealed initial pullout forces between 11.3 and 31 N. Shear stress tests showed that the average force needed to dislocate the stent-grafts by 50% of their diameter ranged between 5.75 and 6.91 N (mean 6.1 +/- 0.5 N) for the 6-mm stents and between 3.31 and 5.4 N (mean 4.4 +/- 0.8) for the 8-mm stents. Conclusion: This preliminary study demonstrated the applicability of the VBX as a bridging stent-graft in a simulated fenestration model. A comparison with other stent-grafts and clinical assessment are required.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 50 条
  • [21] Aortic aneurysms' treatment with endovascular stent-graft
    Rahman, Ali
    Colak, Mehmet Cengiz
    Ozdemir, Huseyin
    Ozguler, Murat
    Cil, Barbaros E.
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 17 (03): : 167 - 172
  • [22] Fenestrated Endovascular Aortic Repair With Chimney Graft for Thoracoabdominal Aneurysm
    Gaines, Sara
    Williamson, Ashley J.
    Park, Julie
    Babrowski, Trissa A.
    Milner, Ross
    VASCULAR AND ENDOVASCULAR SURGERY, 2022, 56 (04) : 444 - 447
  • [23] Bursting Fracture of Balloon-Expandable Stent and Pseudoaneurysm Formation in the Left Common Iliac Artery and Endovascular Treatment With Graft Stent
    Kim, Min
    Lee, Dae In
    Lee, Ju-Hee
    Lee, Sang Yeub
    Bae, Jang-Whan
    Hwang, Kyung-Kuk
    Kim, Dong-Woon
    Cho, Myeong-Chan
    Kim, Sangmin
    VASCULAR AND ENDOVASCULAR SURGERY, 2018, 52 (07) : 556 - 560
  • [24] Outcome of Fenestrated Stent-Graft Repair for Abdominal Aortic Aneurysms: Does the Number of Fenestrations Matter?
    Hoksbergen, Arjan W. J.
    JOURNAL OF ENDOVASCULAR THERAPY, 2017, 24 (02) : 237 - 238
  • [25] One-year results of a balloon expandable endoprosthesis as a bridging stent for branched endovascular aortic repair
    Usai, Marco V.
    Gargiulo, Mauro
    Haulon, Stephan
    Tielliu, Ignace
    Boeckler, Dittmar
    Verhagen, Hence
    Mendez Fernandez, Alba
    Austermann, Martin J.
    JOURNAL OF VASCULAR SURGERY, 2023, 78 (06) : 1376 - +
  • [26] Treatment for thoracoabdominal aortic aneurysm by fenestrated endovascular aortic repair with physician-modified stent graft
    Yang, Xin
    Dai, Xiang-Chen
    Zhu, Jie-Chang
    Luo, Yu-Dong
    Fan, Hai-Lun
    Feng, Zhou
    Zhang, Yi-Wei
    Hu, Fan-Guo
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2018, 46 (05) : 2014 - 2022
  • [27] Comparing Mobile C-Arm with a Hybrid Operating Room for Imaging in Fenestrated Stent-Graft Endovascular Abdominal Aortic Aneurysm Repair
    Yazar, Ozan
    Mees, Barend M. E.
    Provoost, An-Lies
    Ketting, Shirley
    de Haan, Michiel W.
    Schurink, Geert Willem H.
    ANNALS OF VASCULAR SURGERY, 2020, 68 : 261 - 269
  • [28] Endovascular repair of aortic pathologies involving the aortic arch using castor stent-graft combined with in-vitro fenestration technology
    Luo, Zeng-Rong
    Zhang, Jia-Xin
    Huang, Zhong-Yao
    Chen, Liang-Wan
    BMC CARDIOVASCULAR DISORDERS, 2023, 23 (01)
  • [29] Thoracic endovascular stent-graft therapy in aortic dissection
    Akin, Ibrahim
    Kische, Stephan
    Rehders, Tim C.
    Ince, Hueseyin
    Nienaber, Christoph A.
    CURRENT OPINION IN CARDIOLOGY, 2010, 25 (06) : 552 - 559
  • [30] Early results of a low-profile stent-graft for thoracic endovascular aortic repair
    El Beyrouti, Hazem
    Lescan, Mario
    Doemland, Marco
    Mustafi, Migdat
    Jungmann, Florian
    Jorg, Tobias
    Halloum, Nancy
    Dorweiler, Bernhard
    PLOS ONE, 2020, 15 (11):