Development of DICOM image-based CT low dose simulator using fan-beam transform

被引:5
作者
Horiuchi, Tetsuya [1 ,2 ]
Yamamoto, Shuji [3 ]
Murase, Kenya [1 ]
机构
[1] Osaka Univ, Grad Sch Med, Course Hlth Sci, Dept Med Phys & Engn,Div Med Technol & Sci, Suita, Osaka 5650871, Japan
[2] Natl Ctr Child Hlth & Dev, Dept Radiol, Tokyo, Japan
[3] LISIT Co Ltd, Tokyo, Japan
关键词
Computed tomography; CT; noise; dose reduction; computer simulation; STATISTICAL ITERATIVE RECONSTRUCTION; FILTERED BACK-PROJECTION; MODERN DIAGNOSTIC MDCT; NOISE POWER SPECTRUM; X-RAY CT; COMPUTED-TOMOGRAPHY; MULTIDETECTOR CT; PEDIATRIC CT; CANCER RISKS; RADIATION;
D O I
10.3233/THC-130746
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
BACKGROUND: Projection-data based method is established as an accurate simulation of low-dose CT, but cooperation of the CT manufacturer is required. Image-based method is awaited for independent and efficient research. OBJECTIVE: To develop and verify image-based method for adding noise to DICOM images. METHODS: This technique reproduces projection data by fan-beam transform of the DICOM image data and by creating and reconstructing noise projections into a noise image that can be added to the original image. A cylindrical test phantom and anthropomorphic phantom were used to verify the simulation method for both standard and adaptive iterative reconstruction methods. RESULTS: The simulated images compared favorably with actual images. Standard deviation difference of the simulated images were 1.27% +/- 0.81% (cylindrical phantom) and ranged from 1.61% +/- 0.81% (head), to 5.41% +/- 1.57% (hepatic portal region) (anthropomorphic phantom). The standard deviation difference of the cylindrical phantom with the adaptive iterative technique were 3.41% +/- 1.76% at 50% ASIR and 3.85% +/- 1.82% at 100% ASIR. The simulated noise power spectra of both types of images were comparable to the actual images. CONCLUSION: We developed and validated a DICOM image-based low-dose simulation that allows researchers to perform dose reduction research independently and efficiently.
引用
收藏
页码:441 / 454
页数:14
相关论文
共 27 条
  • [1] Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise
    Boedeker, K. L.
    McNitt-Gray, M. F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) : 4047 - 4061
  • [2] Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta
    Boedeker, K. L.
    Cooper, V. N.
    McNitt-Gray, M. F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) : 4027 - 4046
  • [3] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [4] Brenner David J., 2010, Reviews on Environmental Health, V25, P63
  • [5] Brenner DI, 2002, PEDIATR RADIOL, V32, P42
  • [6] Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative
    Brenner, DJ
    [J]. PEDIATRIC RADIOLOGY, 2002, 32 (04) : 228 - 231
  • [7] Estimates of the cancer risks from pediatric CT radiation are not theoretical: Comment on "Point/Counterpoint: In x-ray computed tomography, technique factors should be selected appropriate to size. Against the Proposition" [Med. Phys. 28, 1543-1545 (2001)]
    Brenner, DJ
    Elliston, CD
    Hall, EJ
    Berdon, WE
    [J]. MEDICAL PHYSICS, 2001, 28 (11) : 2387 - 2388
  • [8] Addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain
    Britten, AJ
    Crotty, M
    Kiremidjian, H
    Grundy, A
    Adam, EJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) : 323 - 328
  • [9] Feasibility study of computed tomography colonography using limited bowel preparation at normal and low-dose levels study
    Florie, Jasper
    van Gelder, Rogier E.
    Schutter, Michiel P.
    van Randen, Adrienne
    Venema, Henk W.
    de Jager, Steven
    van der Hulst, Victor P. M.
    Prent, Anna
    Bipat, Shandra
    Bossuyt, Patrick M. M.
    Baak, Lubbertus C.
    Stoker, Jaap
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (12) : 3112 - 3122
  • [10] Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients
    Frush, DP
    Slack, CC
    Hollingsworth, CL
    Bisset, GS
    Donnelly, LF
    Hsieh, J
    Lavin-Wensell, T
    Mayo, JR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (05) : 1107 - 1113