Accuracy of the dose delivery in prostate cancer patients-using an electronic portal imaging device (EPID)

被引:6
|
作者
Mahdavi, S. R. [1 ]
Gharehbagh, E. Jazayeri [2 ]
Mofid, B. [3 ]
Jafari, A. H. [4 ]
Nikoofar, A. R. [5 ]
机构
[1] Iran Univ Med Sci, Dept Med Phys, Tehran, Iran
[2] Univ Tehran Med Sci, Sch Allied Hlth Sci, Dept Radiol & Radiotherapy Technol, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Dept Radiotherapy, Tehran, Iran
[4] Tehran Univ Medial Sci, Res Ctr Biomed Technol & Robot, Tehran, Iran
[5] Iran Univ Med Sci, Dept Radiotherapy, Tehran, Iran
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2017年 / 15卷 / 01期
关键词
Prostate cancer; EPID; systematic errors; random errors; RADIOPAQUE MARKERS; ORGAN MOTION; TREATMENT VERIFICATION; STANDARD RADIOTHERAPY; ROUTINE RADIOTHERAPY; GUIDED RADIOTHERAPY; PELVIC RADIOTHERAPY; IRRADIATION; POSITION; LOCALIZATION;
D O I
10.18869/acadpub.ijrr.15.1.39
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To correct patient positioning errors (setup errors) during prostate cancer treatment using EPID and fiducial gold markers, to improve the accuracy of the dose delivery in these patients. Materials and Methods: Fifteen patients with localized prostate carcinoma after implantation of fiducial gold markers in their prostate gland underwent the five-field IMRT planning technique. The plan was prepared in accordance with ICRU 50 guidance (PTV to receive 95-107% dose). The software program reconstructed the three-dimensional position of the markers from the different Beams Eye Views (BEV). The discrepancies of the seeds' positions (prostate surrogate) between plan and daily images were calculated three dimensionally. Then, necessary corrections were applied to match the prostate fiducial markers in the portal image with the BEV image in the planned one by moving the couch in the X, Y and Z directions. Results: Data from 15 patients and 469 fractions of radiotherapy were analyzed in this study. Two sets of data were available from EPID software before and after 3D set-up corrections. The mean of the population displacement in Left /Right (L/R), Anterior/Posterior (A/P) and Crania/Caudal (C/C) directions were 0.5,-1.0 and 2.4mm before, and-0.1,-0.5 and 0.9mm after corrections, respectively. The systematic and random errors for the measured populations in the three mentioned directions were 2.4, 2.7 and 2mm and 6.4, 5.9 and 6.1mm before corrections, and 1.1, 2.4 and 1.4mm and 3.8, 3.9 and 3.6mm after corrections, correspondingly. Conclusion: This study provides further evidence that using gold markers in the prostate improves dose delivery to the prostate. Also, it has been demonstrated that the EPID can be a powerful tool in the reduction of treatment setup errors and the quality assurance and verification of complex treatments.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [1] Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients
    Pasma, KL
    Kroonwijk, M
    Quint, S
    Visser, AG
    Heijmen, BJM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (05): : 1297 - 1303
  • [2] In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion
    Kroonwijk, M
    Pasma, KL
    Quint, S
    Koper, PCM
    Visser, AG
    Heijmen, BJM
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) : 125 - 132
  • [3] Dose verifications by use of liquid ionization chamber of an electronic portal imaging device (EPID).
    Tateoka K.
    Oouchi A.
    Nakata K.
    Hareyama M.
    Radiological Physics and Technology, 2008, 1 (2) : 171 - 177
  • [4] Reliability analysis of an Electronic Portal Imaging Device (EPID) of a linear accelerator
    Gouveia, Clesia
    Sobral, Jose
    2023 IEEE 7TH PORTUGUESE MEETING ON BIOENGINEERING, ENBENG, 2023, : 52 - 55
  • [5] Daily organ tracking in intensity-modulated radiotherapy of prostate cancer using an electronic portal imaging device with a dose saving acquisition mode
    Vetterli, Daniel
    Thalmann, Sandrine
    Behrensmeier, Frank
    Kemmerling, Ludger
    Born, Ernst J.
    Mini, Roberto
    Greiner, Richard H.
    Aebersold, Daniel M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 79 (01) : 101 - 108
  • [6] Tracking target position variability using intraprostatic fiducial markers and electronic portal imaging in prostate cancer radiotherapy
    Munoz, F.
    Fiandra, C.
    Franco, P.
    Guarneri, A.
    Ciammella, P.
    De Stefanis, P.
    Rondi, N.
    Moretto, F.
    Badellino, S.
    Iftode, C.
    Ragona, R.
    Ricardi, U.
    RADIOLOGIA MEDICA, 2012, 117 (06): : 1057 - 1070
  • [7] Verification of radiosurgery target point alignment with an electronic portal imaging device (EPID)
    Dong, L
    Shiu, A
    Tung, S
    Boyer, A
    MEDICAL PHYSICS, 1997, 24 (02) : 263 - 267
  • [8] Evaluation of set-up errors and estimation of set-up margin during external beam radiation therapy of prostate cancer using electronic portal imaging device (EPID)
    Khoramian, Daryoush
    Sistani, Soroush
    Farhood, Bagher
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 493 - 500
  • [9] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: II. Evaluation of a dosimetric uncertainty budget and development of a new film-in-EPID absorbed dose calibration methodology
    Renaud, James
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (02) : 1238 - 1247
  • [10] The experience of a developing country using an electronic portal imaging device for the verification of patient positioning and dosimetry in radiotherapy for prostate cancer
    Farhat, Leila
    Fourati, N.
    Mnejja, W.
    Daoud, J.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (03) : 297 - 301