A pneumatic power harvesting ankle-foot orthosis to prevent foot-drop

被引:59
作者
Chin, Robin [1 ]
Hsiao-Wecksler, Elizabeth T. [1 ]
Loth, Eric [2 ]
Kogler, Geza [3 ]
Manwaring, Scott D. [1 ]
Tyson, Serena N. [1 ]
Shorter, K. Alex [1 ]
Gilmer, Joel N. [1 ]
机构
[1] Univ Illinois, Dept Engn Sci & Mech, Urbana, IL 61801 USA
[2] Univ Illinois, Dept Aerosp Engn, Urbana, IL 61801 USA
[3] Georgia Inst Technol, Sch Appl Physiol, Clin Biomech Lab, Atlanta, GA 30332 USA
关键词
GAIT; WALKING;
D O I
10.1186/1743-0003-6-19
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A self-contained, self-controlled, pneumatic power harvesting ankle-foot orthosis (PhAFO) to manage foot-drop was developed and tested. Foot-drop is due to a disruption of the motor control pathway and may occur in numerous pathologies such as stroke, spinal cord injury, multiple sclerosis, and cerebral palsy. The objectives for the prototype PhAFO are to provide toe clearance during swing, permit free ankle motion during stance, and harvest the needed power with an underfoot bellow pump pressurized during the stance phase of walking. Methods: The PhAFO was constructed from a two-part (tibia and foot) carbon composite structure with an articulating ankle joint. Ankle motion control was accomplished through a cam-follower locking mechanism actuated via a pneumatic circuit connected to the bellow pump and embedded in the foam sole. Biomechanical performance of the prototype orthosis was assessed during multiple trials of treadmill walking of an able-bodied control subject (n = 1). Motion capture and pressure measurements were used to investigate the effect of the PhAFO on lower limb joint behavior and the capacity of the bellow pump to repeatedly generate the required pneumatic pressure for toe clearance. Results: Toe clearance during swing was successfully achieved during all trials; average clearance 44 +/- 5 mm. Free ankle motion was observed during stance and plantarflexion was blocked during swing. In addition, the bellow component repeatedly generated an average of 169 kPa per step of pressure during ten minutes of walking. Conclusion: This study demonstrated that fluid power could be harvested with a pneumatic circuit built into an AFO, and used to operate an actuated cam-lock mechanism that controls ankle-foot motion at specific periods of the gait cycle.
引用
收藏
页数:11
相关论文
共 21 条
  • [1] *AM AC ORTH SURG, 1997, ATL ORTH ASS DEV
  • [2] [Anonymous], 1992, GAIT ANAL NORMAL PAT
  • [3] [Anonymous], 1999, HUMAN KINETICS
  • [4] A powered leg orthosis for gait rehabilitation of motor-impaired patients
    Banala, Sai K.
    Kulpe, Alexander
    Agrawal, Sunil K.
    [J]. PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, 2007, : 4140 - +
  • [5] Adaptive control of a variable-impedance ankle-foot orthosis to assist drop-foot gait
    Blaya, JA
    Herr, H
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2004, 12 (01) : 24 - 31
  • [6] Design, implementation and test results of a robust control method for a powered ankle foot orthosis (AFO)
    Boehler, Alexander W.
    Hollander, Kevin W.
    Sugar, Thomas G.
    Shin, Dosun
    [J]. 2008 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-9, 2008, : 2025 - +
  • [7] TECHNIQUE FOR THE DISPLAY OF PRESSURE DISTRIBUTIONS BENEATH THE FOOT
    CAVANAGH, PR
    AE, M
    [J]. JOURNAL OF BIOMECHANICS, 1980, 13 (02) : 69 - &
  • [8] CHIN R, 2008, 8 INT S FLUID POW AS
  • [9] An ankle-foot orthosis powered by artificial pneumatic muscles
    Ferris, DP
    Czerniecki, JM
    Hannaford, B
    [J]. JOURNAL OF APPLIED BIOMECHANICS, 2005, 21 (02) : 189 - 197
  • [10] Greater toe grip and gentler heel strike are the strategies to adapt to slippery surface
    Fong, Daniel Tik-Pui
    Mao, De-Wei
    Li, Jing-Xian
    Hong, Youlian
    [J]. JOURNAL OF BIOMECHANICS, 2008, 41 (04) : 838 - 844