Determination of major nonmetallic impurities in magnesium by glow discharge mass spectrometry with a fast flow source using sintered and pressed powder samples

被引:18
作者
Plotnikov, Alexei [1 ]
Pfeifer, Jens [1 ]
Richter, Silke [1 ]
Kipphardt, Heinrich [1 ]
Hoffmann, Volker [2 ]
机构
[1] BAM Fed Inst Mat Res & Testing, D-12489 Berlin, Germany
[2] Leibniz Inst Solid State & Mat Res Dresden, D-01171 Dresden, Germany
关键词
Glow discharge mass spectrometry; Magnesium matrix; Nonmetallic impurities; Fast flow source; Electrical parameters; Calibration samples; ION-SOURCE; GAS-FLOW; OXYGEN; ALLOY; RESOLUTION;
D O I
10.1007/s00216-014-8185-x
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Fast flow glow discharge mass spectrometry with a Grimm-type ion source providing a high sputter rate was used for the determination of major nonmetallic impurities in magnesium. The analytical signal was found to be strongly influenced by the electrical discharge parameters. For calibration by standard addition, synthetic standard samples were produced in two different ways-namely, by pressing and by sintering doped metal powders. The observed sensitivity of the calibration curves was shown to depend on the particle size of the powder. For the magnesium powders, the mass fractions of oxygen, nitrogen, boron, and silicon were determined to be about 0.01 kga <...kg(-1) (relative standard deviation approximately 10-20 %), 2,700 mga <...kg(-1), 150 mga <...kg(-1), and 300 mga <...kg(-1), respectively.
引用
收藏
页码:7463 / 7471
页数:9
相关论文
共 22 条