A solid-state hydroxide ion selective electrode for the measurement of high pH values

被引:9
作者
Ekmekci, G [1 ]
Kalayci, S
Somer, G
机构
[1] Gazi Univ, Gazi Egitim Fak, TR-06500 Ankara, Turkey
[2] Gazi Univ, Fen Edebiyat Fak, TR-06500 Ankara, Turkey
关键词
hydroxide electrode; determination of hydroxide; solid-state membrane; pH electrode;
D O I
10.1016/j.snb.2004.03.012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A solid-state ion selective electrode, which is sensitive to hydroxide ion concentration has been prepared. This electrode with a 10% Cu2S and 90% Ag2Se composition exhibited an analytically useful potential change from 10(-6) to 10(-1) M hydroxide ion, pH 8-13. The slope of linear portion is about 58 mV per decade change in hydroxide concentration. There was no interference of K+, Na+, Ca2+, Zn2+ and Mg2+ cations and Cl-, NO3- and SO42- anions. The response time was 30 s and the lifetime was about 2 years. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:260 / 264
页数:5
相关论文
共 15 条
[1]  
BATES RG, 1996, FUNDAMENTALS ANAL CH
[2]   A new selenite selective membrane electrode and its application [J].
Ekmekçi, G ;
Somer, G .
TALANTA, 1999, 49 (01) :83-89
[3]   Preparation and properties of solid state selenite ion selective electrodes and their applications [J].
Ekmekçi, G ;
Somer, G .
TALANTA, 1999, 49 (01) :91-98
[4]   Selenite-selective membrane electrodes based on ion exchangers and application to anodic slime [J].
Ekmekçi, G ;
Somer, G .
ANALYTICAL SCIENCES, 2000, 16 (03) :307-311
[5]  
Han WS, 2001, ELECTROANAL, V13, P955, DOI 10.1002/1521-4109(200107)13:11<955::AID-ELAN955>3.0.CO
[6]  
2-5
[7]   ENHANCED LIFETIME AND ADHESION OF K+-SENSITIVE, NH4+-SENSITIVE, AND CA-2+-SENSITIVE MEMBRANES ON SOLID-SURFACES USING HYDROXYL-MODIFIED POLYVINYLCHLORIDE MATRICES [J].
HARRISON, DJ ;
CUNNINGHAM, LL ;
LI, XZ ;
TECLEMARIAM, A ;
PERMANN, D .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1988, 135 (10) :2473-2478
[8]   ANALYTICAL STUDY OF LEAD ION-SELECTIVE CERAMIC MEMBRANE ELECTRODE [J].
HIRATA, H ;
HIGASHIYAMA, K .
BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 1971, 44 (09) :2420-+
[9]   Characterization of all solid state hydrogen ion selective electrode based on PVC-SR hybrid membranes [J].
Piao, MH ;
Yoon, JH ;
Gerok-Jeon ;
Shim, YB .
SENSORS, 2003, 3 (06) :192-201
[10]  
PUNGOR E, 1961, ACTA CHIM HUNG, V27, P63