ASSESSMENT OF SPATIAL UNCERTAINTIES IN THE RADIOTHERAPY PROCESS WITH THE NOVALIS SYSTEM

被引:31
作者
Hayashi, Naoki [1 ,2 ]
Obata, Yasunori [2 ]
Uchiyama, Yukio [3 ]
Mori, Yoshimasa [1 ]
Hashizume, Chisa [1 ]
Kobayashi, Tatsuya [1 ]
机构
[1] Nagoya Kyoritsu Hosp, Nagoya Radiosurg Ctr, Nagoya, Aichi, Japan
[2] Nagoya Univ, Grad Sch Med Sci, Nagoya, Aichi 4648601, Japan
[3] Gifu Univ Med Sci, Sch Hlth Sci, Gifu, Japan
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2009年 / 75卷 / 02期
关键词
Spatial uncertainty; Image-guided radiotherapy; Geometrical accuracy; Patient setup; Stereotactic radiotherapy; PATIENT SET-UP; GEOMETRICAL UNCERTAINTIES; VERIFICATION PROCEDURE; PROSTATE; ACCURACY; MARGINS; MARKER; CANCER; LOCALIZATION; IRRADIATION;
D O I
10.1016/j.ijrobp.2009.02.080
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study was to evaluate the accuracy of a new version of the ExacTrac X-ray (ETX) system with statistical analysis retrospectively in order to determine the tolerance of systematic components of spatial uncertainties with the Novalis system. Methods and Materials: Three factors of geometrical accuracy related to the ETX system were evaluated by phantom studies. First, location dependency of the detection ability of the infrared system was evaluated. Second, accuracy of the automated calculation by the image fusion algorithm in the patient registration software was evaluated. Third, deviation of the coordinate scale between the ETX isocenter and the mechanical isocenter was evaluated. From the values of these examinations and clinical experiences, the total spatial uncertainty with the Novalis system was evaluated. Results: As to the location dependency of the detection ability of the infrared system, the detection errors between the actual position and the detected position were 1% in translation shift and 0.1 degrees in rotational angle, respectively. As to the accuracy of patient verification software, the repeatability and the coincidence of the calculation value by image fusion were good when the contrast of the X-ray image was high. The deviation of coordinates between the ETX isocenter and the mechanical isocenter was 0.313 +/- 0.024 mm, in a suitable procedure. Conclusions: The spatial uncertainty will be less than 2 mm when suitable treatment planning, optimal patient setup, and daily quality assurance for the Novalis system are achieved in the routine workload. (C) 2009 Elsevier Inc.
引用
收藏
页码:549 / 557
页数:9
相关论文
共 23 条
[1]  
[Anonymous], 1993, J ICRU
[2]   MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS [J].
BALTER, JM ;
SANDLER, HM ;
LAM, K ;
BREE, RL ;
LICHTER, AS ;
TENHAKEN, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01) :113-118
[3]   Target margins for random geometrical treatment uncertainties in conformal radiotherapy [J].
Bel, A ;
vanHerk, M ;
Lebesque, JV .
MEDICAL PHYSICS, 1996, 23 (09) :1537-1545
[4]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[5]   A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES [J].
BEL, A ;
VANHERK, M ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :253-260
[6]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[7]   A quality control study of the accuracy of patient positioning in irradiation of pelvic fields [J].
Creutzberg, CL ;
Althof, VGM ;
deHoog, M ;
Visser, AG ;
Huizenga, H ;
Wijnmaalen, A ;
Levendag, PC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (03) :697-708
[8]  
FUKUGAMI S, 2006, KATAROU KAI MAGAZINE, V11, P90
[9]   Assessment of secondary patient motion induced by automated couch movement during on-line 6 dimensional repositioning in prostate cancer treatment [J].
Linthout, Nadine ;
Verellen, Dirk ;
Tournel, Koen ;
Reynders, Truus ;
Duchateau, Michael ;
Storme, Guy .
RADIOTHERAPY AND ONCOLOGY, 2007, 83 (02) :168-174
[10]   Optimal marker placement in photogrammetry patient positioning system [J].
Liu, HS ;
Yu, Y ;
Schell, MC ;
O'Dell, WG ;
Ruo, R ;
Okunieff, P .
MEDICAL PHYSICS, 2003, 30 (02) :103-110