Direct human breath analysis by secondary nano-electrospray ionization ultrahigh-resolution mass spectrometry: Importance of high mass resolution and mass accuracy

被引:29
作者
Li, Xue [1 ,2 ]
Huang, Lei [1 ,2 ]
Zhu, Hui [3 ]
Zhou, Zhen [1 ,2 ]
机构
[1] Jinan Univ, Inst Mass Spectrometry & Atmospher Environm, 601 West Huangpu Ave, Guangzhou 510632, Guangdong, Peoples R China
[2] Guangdong Prov Engn Res Ctr On Line Source Apport, Guangzhou 510632, Guangdong, Peoples R China
[3] Guangzhou Hexin Instrument Co Ltd, Guangzhou 510530, Guangdong, Peoples R China
关键词
VOLATILE ORGANIC-COMPOUNDS; EXHALED BREATH; LUNG-CANCER; AMMONIA; FLOW; AEROSOLS; AIR; MS;
D O I
10.1002/rcm.7794
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
RationaleDirect mass spectrometry (MS)-based methods make it possible to monitor the molecular compositions of hundreds of volatile organic compounds (VOCs) in exhaled human breath in real time. Mass resolution and mass accuracy play important roles for direct MS analysis, especially for the low-concentration isobaric compounds in non-target research. MethodsDirect detection of VOCs in exhaled breath of four healthy subjects (3 males and 1 female aged between 25 to 35years old) has been performed by using secondary nano-electrospray ionization mass spectrometry (Sec-nanoESI-UHRMS) at resolutions (R) of 15,000, 30,000, 60,000 and 120,000. ResultsFor some low-intensity isobaric ions, they could be distinguished only when R60,000, e.g., signals at m/z 96.9591 (sulfate/sulfuric acid), m/z 96.9687 (phosphate/phosphoric acid) and m/z 96.9756 ([C4H2O7S](-)), m/z 234.1161 ([C10H20O3NS](+)) and m/z 234.1338 ([C10H20O5N](+)), m/z 119.0686 (isotope of indole) and m/z 119.0705 (an interfering signal), respectively. At R 120,000, the mass errors were obtained from a set of reference ions, and the values were 0.6 mDa for ions detected in positive detection mode and in the range of -1.0-1.1 mDa for the negative mode. These mass errors were used to exclusively identify unknown compounds detected in the breath samples. By utilizing the present setup, besides the normal VOCs reported previously, we detected non-volatile species (sulfate/sulfuric acid, silicate/silicic acid, phosphate/phosphoric acid and nitrate/nitric acid), dichlorobenzene and an ammonium adduct ([(C2H6SiO)(6)+NH4](+)), which were ascribed to exhaled particles, indoor air pollution and an endogenous source, respectively. ConclusionsFor direct breath analysis, high mass resolution of 60,000 and mass errors of 1.0 mDa (absolute value) covering the mass range of interests (e.g., m/z 50-500) are necessary for the exploration and accurate identification of low-intensity unknown isobaric compounds in non-target research. Copyright (c) 2017 John Wiley & Sons, Ltd.
引用
收藏
页码:301 / 308
页数:8
相关论文
共 33 条
  • [1] Noninvasive detection of lung cancer by analysis of exhaled breath
    Bajtarevic, Amel
    Ager, Clemens
    Pienz, Martin
    Klieber, Martin
    Schwarz, Konrad
    Ligor, Magdalena
    Ligor, Tomasz
    Filipiak, Wojciech
    Denz, Hubert
    Fiegl, Michael
    Hilbe, Wolfgang
    Weiss, Wolfgang
    Lukas, Peter
    Jamnig, Herbert
    Hackl, Martin
    Haidenberger, Alfred
    Buszewski, Boguslaw
    Miekisch, Wolfram
    Schubert, Jochen
    Amann, Anton
    [J]. BMC CANCER, 2009, 9 : 348
  • [2] Benoit F. M., 1985, INT ARCH OCC ENV HEA, V55, P8
  • [3] BREATH ANALYSIS BY ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY
    BENOIT, FM
    DAVIDSON, WR
    LOVETT, AM
    NACSON, S
    NGO, A
    [J]. ANALYTICAL CHEMISTRY, 1983, 55 (04) : 805 - 807
  • [4] Real-time breath analysis with active capillary plasma ionization-ambient mass spectrometry
    Bregy, Lukas
    Sinues, Pablo Martinez-Lozano
    Nudnova, Maryia M.
    Zenobi, Renato
    [J]. JOURNAL OF BREATH RESEARCH, 2014, 8 (02)
  • [5] Analysis of volatile flavour compounds by Proton Transfer Reaction-Mass Spectrometry: fragmentation patterns and discrimination between isobaric and isomeric compounds
    Buhr, K
    van Ruth, S
    Delahunty, C
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 221 (01) : 1 - 7
  • [6] Rapid in vivo fingerprinting of nonvolatile compounds in breath by extractive electrospray ionization quadrupole time-of-flight mass spectrometry
    Chen, Huanwen
    Wortmann, Arno
    Zhang, Weihua
    Zenobi, Renato
    [J]. ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2007, 46 (04) : 580 - 583
  • [7] Costello B. desLacy, 2014, J BREATH RES, V8
  • [8] Respiratory symptoms and peak expiratory flow in children with asthma in relation to volatile organic compounds in exhaled breath and ambient air
    Delfino, RJ
    Gong, H
    Linn, WS
    Hu, Y
    Pellizzari, ED
    [J]. JOURNAL OF EXPOSURE ANALYSIS AND ENVIRONMENTAL EPIDEMIOLOGY, 2003, 13 (05): : 348 - 363
  • [9] Analysis of secondary organic aerosols in air using extractive electrospray ionization mass spectrometry (EESI-MS)
    Doezema, Lambert A.
    Longin, Teresa
    Cody, William
    Perraud, Veronique
    Dawson, Matthew L.
    Ezell, Michael J.
    Greaves, John
    Johnson, Kathleen R.
    Finlayson-Pitts, Barbara J.
    [J]. RSC ADVANCES, 2012, 2 (07) : 2930 - 2938
  • [10] Breath ammonia and trimethylamine allow real-time monitoring of haemodialysis efficacy
    Endre, Z. H.
    Pickering, J. W.
    Storer, M. K.
    Hu, W-P
    Moorhead, K. T.
    Allardyce, R.
    McGregor, D. O.
    Scotter, J. M.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2011, 32 (01) : 115 - 130