Microprobe x-ray fluorescence with the use of point-focusing diffractors

被引:16
|
作者
Chen, ZW
Wittry, DB
机构
[1] Dept. of Mat. Sci. and Engineering, University of Southern California, Los Angeles
关键词
D O I
10.1063/1.119427
中图分类号
O59 [应用物理学];
学科分类号
摘要
A toroidal point-focusing mica crystal diffractor was used to focus monochromatic x rays from a microfocus x-ray source operated at 0.1 mA and 30 kV, The Cu K-alpha x-ray focal spot of 50 mu m x 85 mu m had 1.6 x 10(4) photons/s/mu m(2). Microprobe x-ray fluorescence (MXRF) spectra were recorded with a Si(Li) energy dispersive detector for bulk specimens of GaAs. Si, and Muscovite. Low background due to monochromatic excitation resulted in predicted detection limits as low as 2 ppm for a measurement time of 500 s. Laboratory MXRF systems based on point-focusing diffractors were shown to provide lower detection limits. larger working distance, and higher theoretical intensity than systems using capillary optics. (C) 1997 American Institute of Physics.
引用
收藏
页码:1884 / 1886
页数:3
相关论文
共 50 条
  • [2] POINT-FOCUSING 2-CRYSTAL X-RAY MONOCHROMATOR FOR X-RAY DIFFRACTION
    FURNAS, TC
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1957, 28 (12): : 1042 - 1048
  • [3] SINGLE QUARTZ CRYSTAL POINT-FOCUSING X-RAY MONOCHROMATOR
    BERREMAN, DW
    SCIENCE, 1955, 122 (3175) : 873 - 873
  • [4] DOUBLY CURVED CRYSTAL POINT-FOCUSING X-RAY MONOCHROMATORS
    BERREMAN, DW
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1976, 66 (10) : 1136 - 1136
  • [5] FOCUSING PROPERTIES OF CURVED X-RAY DIFFRACTORS
    WITTRY, DB
    SUN, SQ
    JOURNAL OF APPLIED PHYSICS, 1990, 68 (02) : 387 - 391
  • [6] A simple short-range point-focusing spatial filter for time-resolved X-ray fluorescence
    Hoehr, C.
    Peterson, E.
    Landahl, E.
    Walko, D. A.
    Dunford, R. W.
    Kanter, E. P.
    Young, L.
    SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 1226 - +
  • [7] Realization of a High-Performance Point-Focusing Monochromator for X-ray Studies
    Okuda, Hiroshi
    Morishita, Kohei
    Nakajima, Kazuo
    Fujiwara, Kozo
    Yonenaga, Ichiro
    Ochiai, Shojiro
    APPLIED PHYSICS EXPRESS, 2010, 3 (04)
  • [8] DOUBLY CURVED CRYSTAL POINT-FOCUSING X-RAY MONOCHROMATORS - GEOMETRICAL AND PRACTICAL OPTICS
    BERREMAN, DW
    STAMATOFF, J
    KENNEDY, SJ
    APPLIED OPTICS, 1977, 16 (08): : 2081 - 2085
  • [9] APPLICATION OF X-RAY FOCUSING OPTICS TO A SYNCHROTRON BASED X-RAY MICROPROBE
    THOMPSON, AC
    UNDERWOOD, JH
    WU, Y
    GIAUQUE, RD
    JONES, KW
    RIVERS, ML
    FUTERNICK, RG
    X-RAY INSTRUMENTATION IN MEDICINE AND BIOLOGY, PLASMA PHYSICS, ASTROPHYSICS, AND SYNCHROTRON RADIATION, 1989, 1140 : 32 - 32
  • [10] DESIGN AND CONSTRUCTION OF A BERREMAN-TYPE POINT-FOCUSING X-RAY MONOCHROMATOR FOR USE IN LOW-ANGLE SCATTERING STUDIES
    DIBBLE, WE
    DUMOND, JWM
    JOURNAL OF ULTRASTRUCTURE RESEARCH, 1963, 9 (5-6): : 419 - &