Reliability of delayed INR determination: Implications for decentralized anticoagulant care with off-site blood sampling

被引:38
作者
Baglin, T
Luddington, R
机构
[1] Department of Haematology, Addenbrooke's NHS Trust, Cambridge
[2] Department of Haematology, Addenbrooke's NHS Trust
关键词
anticoagulation; warfarin; thromboplastin; INR;
D O I
10.1046/j.1365-2141.1997.d01-2079.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In order to develop decentralized anticoagulant care by off-site blood sampling and transport of samples to a centralized laboratory for International Normalized Ratio (INR) determination we have performed a direct comparative study of INR stability Analysis was performed daily for 5 d using nine thromboplastins. The overall mean difference of INR after 3 d was only 0.05 INR units for samples with a therapeutic INR. After 5 d there was a mean difference of 0.11 INR units with 'non-Manchester' reagents and 0.44 INR units with 'Manchester' reagents. With over-anticoagulated samples mean differences of 0.55-0.72 INR units were observed after 3 d and 1.16-2.46 INR units after 5 d. Although there was some variation in stability of results with different thromboplastins, the difference over time with each thromboplastin was much less than the difference between thromboplastins. In conclusion, there is no clinically significant change in INR when analysis is delayed for up to 3 d. Off-site blood sampling can accommodate a large increase in patient workload without a major revenue increase in primary care and with continued total quality management and central expert advice.
引用
收藏
页码:431 / 434
页数:4
相关论文
共 10 条
  • [1] BAGLIN T, 1994, CLIN LAB HAEMATOL, V16, P327
  • [2] BAGLIN T, 1996, BR J HAEMATOL S1, V95, pA115
  • [3] *BRIT SOC HAEM, 1990, J CLIN PATHOL, V43, P177
  • [4] Monitoring oral anticoagulation in primary care - Quality assurance and computerised decision support will help to minimise inconsistencies
    Fitzmaurice, DA
    Hobbs, FDR
    Murray, JA
    [J]. BRITISH MEDICAL JOURNAL, 1996, 312 (7044) : 1431 - 1432
  • [5] KITCHEN S, 1996, BRIT J HAEMATOLOG S2, V93
  • [6] BEHAVIOUR OF FACTORS II, VII, IX, AN X DURING LONG-TERM TREATMENT WITH COUMARIN
    LOELIGER, EA
    MATTERN, MJ
    BRABANDER, AS
    VANDERESCH, B
    [J]. THROMBOSIS ET DIATHESIS HAEMORRHAGICA, 1963, 9 (1-2): : 74 - &
  • [7] THE VALUE OF PLASMA CALIBRANTS IN CORRECTING COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS - AN INTERNATIONAL MULTICENTER STUDY
    POLLER, L
    TRIPLETT, DA
    HIRSH, J
    CARROLL, J
    CLARKE, K
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (03) : 358 - 365
  • [8] POLLER L, 1985, RECENT ADV BLOOD COA, P245
  • [9] EFFECT OF INTERNATIONAL SENSITIVITY INDEX (ISI) OF THROMBOPLASTINS ON PRECISION OF INTERNATIONAL NORMALIZED RATIOS (INR)
    TABERNER, DA
    POLLER, L
    THOMSON, JM
    DARBY, KV
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (01) : 92 - 96
  • [10] VANDENBESSELAAR AMHP, 1993, THROMB HAEMOSTASIS, V70, P794