Comparison of metabolic monitors in critically ill, ventilated patients

被引:27
作者
Singer, Pierre
Pogrebetsky, Ira
Attal-Singer, Joelle
Cohen, Jonathan
机构
[1] Rabin Med Ctr, Dept Gen Intens Care, Petah Tiqwa, Israel
[2] Tel Aviv Univ, Sackler Sch Med, IL-69978 Tel Aviv, Israel
[3] Rabin Med Ctr, Inst Endocrinol, Petah Tiqwa, Israel
关键词
indirect calorimetry; metabolic monitors; critically ill;
D O I
10.1016/j.nut.2006.06.007
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: We compared the Deltatrac II, the M-COVX, and the Evita 4 metabolic monitoring devices under clinical conditions. Methods: A prospective simultaneous clinical comparison was performed in a general intensive care department of a tertiary university hospital in 43 ventilated, critically ill patients. The monitors were compared simultaneously. After 30 min of steady state, oxygen consumption per unit time, carbon dioxide consumption per unit time, resting energy expenditure, and respiratory quotient were recorded for the Deltatrac II; the same parameters in addition to end-tidal carbon dioxide and fraction of inspired oxygen were recorded for the M-COVX; and carbon dioxide consumption per unit time, end-tidal carbon dioxide, and fraction of inspired oxygen were recorded for the Evita 4. Pulmonary gas-exchange measurements from the Deltatrac II and resting energy expenditure and respiratory quotient from the M-COVX were obtained after 30 min. The other parameters were calculated from the last five measurements obtained at the end of the study period. Results: A good correlation was found between oxygen consumption per unit time and resting energy expenditure as obtained from the Deltatrac II and the M-COVX (r = 0.76 and 0.75, respectively; P < 0.001), but the correlation was lower between carbon dioxide consumption per unit time as obtained from the Deltatrac II and the M-COVX or Evita 4 (r = 0.67 and 0.48, respectively). Agreement between the different methods did not reach clinical acceptability, exceeding a 20% difference using the Bland-Altman statistical methods. Conclusion: Poor agreement was found between the Deltatrac II and M-COVX or Evita 4 metabolic monitors, despite a good correlation between measurements, leading to the conclusion that the M-COVX and Evita 4 provide less accurate measurements of metabolic gas exchange in stable ventilated patients. These devices can be used for daily nutritional assessment and continuous monitoring, but the Deltatrac II remains the method of choice for metabolic measurement. (C) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:1077 / 1086
页数:10
相关论文
共 25 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Indirect calorimetry in critically ill patients: Clinical applications and practical advice [J].
Brandi, LS ;
Bertolini, R ;
Calafa, M .
NUTRITION, 1997, 13 (04) :349-358
[3]  
*BRIT STAND I, 1979, PREC TEST METH, V1
[4]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[5]   Clinical evaluation of a continuous oxygen consumption monitor in mechanically ventilated patients [J].
Donaldson, L ;
Dodds, S ;
Walsh, TS .
ANAESTHESIA, 2003, 58 (05) :455-460
[6]   The use of oxygen consumption and delivery as endpoints for resuscitation in critically ill patients [J].
Durham, RM ;
Neunaber, K ;
Mazuski, JE ;
Shapiro, MJ ;
Baue, AE .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1996, 41 (01) :32-39
[7]   Computerized energy balance and complications in critically ill patients: An observational study [J].
Dvir, D ;
Cohen, J ;
Singer, P .
CLINICAL NUTRITION, 2006, 25 (01) :37-44
[8]  
FERNANDEZ JP, 1999, NUTR HOSP, V14, P23
[9]   Effect of low-dose propofol infusion on total-body oxygen consumption after coronary artery surgery [J].
Hammarén, E ;
Scheinin, M ;
Hynynen, M .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1999, 13 (02) :154-159
[10]   ELEVATION OF SYSTEMIC OXYGEN DELIVERY IN THE TREATMENT OF CRITICALLY ILL PATIENTS [J].
HAYES, MA ;
TIMMINS, AC ;
YAU, EHS ;
PALAZZO, M ;
HINDS, CJ ;
WATSON, D .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 330 (24) :1717-1722