Focused ultrasound treatment of VX2 tumors controlled by local harmonic motion

被引:22
|
作者
Curiel, Laura [1 ,2 ]
Huang, Yuexi [1 ]
Vykhodtseva, Natalia [3 ]
Hynynen, Kullervo [1 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Thunder Bay Reg Res Inst, High Intens Focused Ultrasound HIFU Lab, Toronto, ON P7C 1M5, Canada
[3] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 11期
关键词
IN-VIVO; PROSTATE-CANCER; TEMPERATURE; SURGERY; LESIONS; MRI; ELASTOGRAPHY; THERMOMETRY; DOSIMETRY; ABLATION;
D O I
10.1088/0031-9155/54/11/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to evaluate the feasibility of using localized harmonic motion (LHM) to monitor and control focused ultrasound surgery (FUS) in VX2 tumors in vivo. FUS exposures were performed on 13 VX2 tumors implanted in nine rabbits. The same transducer induced coagulation and generated a localized oscillatory motion by periodically varying the radiation force. A separate diagnostic ultrasound transducer tracked motion by cross-correlating echo signals at different instances. A threshold inmotion amplitude was instituted to cease exposure. Coagulation was confirmed by T2-weighted MR images, thermal dose obtained through MR thermometry and histological examinations. For tumor locations achieving coagulation, the LHM amplitude was 9% (p = 0.04) to 57% (p < 0.0001) lower than that before exposure. Control was successful for 74 (69%) out of 108 cases, with 52 (48%) reaching the threshold and achieving coagulation and 22 (21%) never reaching threshold nor coagulating. For the 34 (31%) unsuccessful exposures, 16 (15%) never reached the threshold but coagulation occurred, and 18 (16%) reached threshold without coagulation confirmed. Noise or radio-frequency signal changes explained motion over- or underestimation in 24 (22%) cases; the remaining 10 (9%) had other causes of error. The control was generally successful, but sudden change or noise in the acquired echo signal caused failure. Coagulation after exposure could be validated by comparing amplitudes before and after exposure.
引用
收藏
页码:3405 / 3419
页数:15
相关论文
共 50 条
  • [41] Magnetic resonance imaging-guided focused ultrasound thermal therapy in experimental animal models: Correlation of ablation volumes with pathology in rabbit muscle and VX2 tumors
    Hazle, JD
    Stafford, RJ
    Price, RE
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (02) : 185 - 194
  • [42] Percutaneous microwave coagulation for eradication of VX2 tumors subcutaneously in rabbits
    Zhou, Wenbin
    Ding, Qiang
    Liu, Xiaoan
    Jiang, Yanni
    Chen, Ling
    Zhang, Yifen
    Xia, Tiansong
    Wang, Shui
    WORLD JOURNAL OF SURGICAL ONCOLOGY, 2012, 10
  • [43] Development of a rabbit pleural cancer model by using VX2 tumors
    Kreuter, Kelly A.
    El-Abbadi, Naglaa
    Shbeeb, Alia
    Tseng, Lillian
    Mahon, Sari Brenner
    Narula, Navneet
    Burney, Tanya
    Colt, Henri
    Brenner, Matthew
    COMPARATIVE MEDICINE, 2008, 58 (03) : 287 - 293
  • [44] Experimental study of applying shear wave elastography for examining effects of high intensity focused ultrasound ablation against subcutaneously implanted VX2 tumors in rabbits.
    Chen, Fei
    Li, Sheng
    Wu, Limei
    Zou, Jianzhong
    Qiu, Shaodong
    Wang, Quanshi
    BIOMEDICAL RESEARCH-INDIA, 2017, 28 (17): : 7398 - 7403
  • [45] Percutaneous microwave coagulation for eradication of VX2 tumors subcutaneously in rabbits
    Wenbin Zhou
    Qiang Ding
    Xiaoan Liu
    Yanni Jiang
    Ling Chen
    Yifen Zhang
    Tiansong Xia
    Shui Wang
    World Journal of Surgical Oncology, 10
  • [46] Radio-frequency ablation of VX2 rabbit tumors:: Assessment of completeness of treatment by using contrast-enhanced harmonic power Doppler US
    Boehm, T
    Malich, A
    Goldberg, SN
    Hauff, P
    Reinhardt, M
    Reichenbach, JR
    Müller, W
    Fleck, M
    Seifert, B
    Kaiser, WA
    RADIOLOGY, 2002, 225 (03) : 815 - 821
  • [47] ARTERIAL SUPPLY OF TRANSPLANTED VX2 + XY TUMORS IN RABBIT LUNGS
    OGILVIE, RW
    KNISELY, WH
    ISRAEL JOURNAL OF EXPERIMENTAL MEDICINE, 1964, 11 (03): : 117 - &
  • [48] ARTERIAL SUPPLY OF TRANSPLANTED VX2 AND XY TUMORS IN RABBIT LUNGS
    KNISELY, WH
    OGILVIE, R
    ANATOMICAL RECORD, 1963, 145 (02): : 250 - &
  • [49] High-intensity Focused Ultrasound Ablation of Soft-tissue Tumors and Assessment of Treatment Response with Multiparametric Magnetic Resonance Imaging: Preliminary Study Using Rabbit VX2 Tumor Model
    Kim, Kyung Won
    Lee, Jae Young
    Lee, Jeong Min
    Jeon, Yong Sik
    Choi, Yoon Seok
    Park, Jisuk
    Kim, Harry
    Han, Joon Koo
    Choi, Byung Ihn
    JOURNAL OF MEDICAL ULTRASOUND, 2014, 22 (02) : 99 - 105
  • [50] Multi-parametric monitoring of High Intensity Focused Ultrasound (HIFU) Treatment Using Harmonic Motion Imaging For Focused Ultrasound (HMIFU)
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa
    12TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1503 : 140 - 145