Multidirectional deformation in fully congruent acetabular components

被引:8
作者
Akisue, T [1 ]
Bauer, TW [1 ]
Yamaguchi, M [1 ]
Matejczyk, MB [1 ]
Stulberg, BN [1 ]
Wilde, AH [1 ]
机构
[1] Cleveland Clin Fdn, Dept Anat Pathol, Cleveland, OH 44195 USA
关键词
total hip arthroplasty; polyethylene; modular acetabular component; wear; retrieved implants;
D O I
10.1016/S0883-5403(99)90018-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Most clinical studies have used femoral head migration as an index of acetabular wear, but a previous study showed multiple wear vectors in 30% of retrieved acetabular components with noncongruent liners. The origin of multiple wear vectors is unclear, and it has been suggested that polyethylene creep in a noncongruent shell might influence deformation on the articular surface. We used shadowgraph and volumetric methods to evaluate the extent and direction of surface deformation of 37 retrieved polyethylene liners that were fully congruent to a single design of metal backing. The results show that multiple deformation vectors are relatively common in retrieved acetabular cups (27% in this study) and are independent of congruency between liner and metal backing, rim impingement, and backside creep. Polyethylene liners with multiple wear vectors were significantly thinner than those of cups with a single vector. The origin of multiple vectors is still unclear, bur clinical and laboratory studies measuring linear wear alone without recognizing multiple vectors underestimate total in vivo volumetric wear.
引用
收藏
页码:1011 / 1018
页数:8
相关论文
共 34 条
[1]  
[Anonymous], CLIN ORTHOPAEDICS
[2]  
[Anonymous], 1992, SOCIOLOGICA, V7
[3]  
Bankston A B, 1994, J Arthroplasty, V9, P533, DOI 10.1016/0883-5403(94)90101-5
[4]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[5]   THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[6]   OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[7]   Early failure of Hylamer acetabular inserts due to eccentric wear [J].
Chmell, MJ ;
Poss, R ;
Thomas, WH ;
Sledge, CB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (03) :351-353
[8]  
Cooper R A, 1992, J Arthroplasty, V7, P285, DOI 10.1016/0883-5403(92)90050-Z
[9]   Wear of polyethylene in artificial hip joints: Superolateral wear of the acetabulum [J].
Fisher, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02) :190-191
[10]   OSTEOLYSIS AND PARTICLE DISEASE IN HIP-REPLACEMENT - A REVIEW [J].
HARRIS, WH .