Limitations of gas filter correlation: A case study on carbon monoxide non-dispersive infrared analyzer

被引:10
|
作者
Dinh, Trieu-Vuong [1 ]
Ahn, Ji-Won [2 ]
Choi, In-Young [1 ]
Song, Kyu-Yong [1 ]
Chung, Chang-Ho [2 ]
Kim, Jo-Chun [1 ]
机构
[1] Konkuk Univ, Dept Environm Engn, 120 Neungdong Ro, Seoul 143701, South Korea
[2] Konkuk Univ, Int Climate & Environm Res Ctr, 120 Neungdong Ro, Seoul 143701, South Korea
关键词
NDIR; GFC; Interference; Artificial formation; Analytical mode; INTERFERENCE CORRECTION; ABSORPTION; EMISSION; SO2;
D O I
10.1016/j.snb.2016.12.036
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two commercial Non-Dispersive Infrared (NDIR) analyzers for carbon monoxide (CO) coupled with a gas filter correlation (GFC) were used to investigate the interference effects by other gases except for CO. A stack level of particular gases including CO2, NO, NO2, SO2, and H2O (g) were mixed with CO to introduce into the analyzers. It was found that the measurement error of two analyzers were the lowest at 50 ppm and 70 ppm of CO, respectively. Moreover, the increase of concentrations of the interfering gases (CO2, NO, NO2, and SO2) induced an interfering effect on the performance of two analyzers. This indicated that the GFC could not cover the entire range of interfering gases. In a special case, the accuracy of the analyzers with humid mixed gases was better than that with dry mixed gases due to the absorption of water-soluble gases by condensate. However, this accuracy was only apparent with respect to low solubility of analytical target gas such as CO. The solubility of the target gas might dramatically reduce the accuracy of the analyzer. (C) 2016 Published by Elsevier B.V.
引用
收藏
页码:684 / 689
页数:6
相关论文
共 50 条
  • [1] A Novel Bandpass Filter for the Analysis of Carbon Monoxide Using a Non-Dispersive Infrared Technique
    Dinh, Trieu-Vuong
    Ahn, Ji-Won
    Choi, In-Young
    Kim, Jo-Chun
    ATMOSPHERE, 2018, 9 (12):
  • [2] Study of SO2 Detecting Using Non-dispersive Infrared Based on Gas Filter Correlation Technology
    Dong, Zhifeng
    Shao, Lian
    Yang, Huijia
    Yao, Rong
    Yang, Xiong
    ADVANCED DESIGN AND MANUFACTURING TECHNOLOGY III, PTS 1-4, 2013, 397-400 : 278 - 281
  • [3] PRESSURE EFFECT OF NON-DISPERSIVE IR GAS ANALYZER
    YONEZAWA, T
    NIKI, E
    BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 1977, 50 (02) : 341 - 344
  • [4] A parallel readout circuit for a space-based non-dispersive infrared gas analyzer
    Tran, Dat
    Gorius, Nicolas
    Quilligan, Gerard T.
    Gong, Qian
    Kolasinski, John R.
    Purser, Todd C.
    Siguelnitzky, Ariel
    Parker, James E.
    Wegel, Donald C.
    Glavin, Daniel P.
    Gerakines, Perry A.
    Aslam, Shahid
    Phan, Thuc
    Nehmetallah, George
    SENSORS AND SYSTEMS FOR SPACE APPLICATIONS XV, 2022, 12121
  • [5] Modeling of Non-Dispersive Infrared Gas Sensors
    Dietachmayr, F. W.
    Zagar, B. G.
    SENSORS AND ELECTRONIC INSTRUMENTATION ADVANCES (SEIA), 2016, : 95 - 99
  • [6] DETERMINATION OF CARBON MONOXIDE IN BLOOD BY MEANS OF A NON-DISPERSIVE INFRARED DIFFERENTIAL ABSORPTION TECHNIQUE
    KRUSMANN, K
    SCHRODER, H
    SCHRODER, J
    ARCHIV FUR TOXIKOLOGIE, 1971, 27 (02): : 99 - &
  • [7] The development system of portable of CO2 non-dispersive infrared detector gas analyzer
    Yeh, T
    TECHNICAL DIGEST OF THE SEVENTH INTERNATIONAL MEETING ON CHEMICAL SENSORS, 1998, : 675 - 685
  • [8] Non-Dispersive Infrared Gas Sensing Technology: A Review
    Jha, Ravindra Kumar
    IEEE SENSORS JOURNAL, 2022, 22 (01) : 6 - 15
  • [9] PROCESS GAS ANALYZERS USING NON-DISPERSIVE INFRARED
    DURHAM, RM
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1983, 411 : 72 - 79
  • [10] Multi-Gas Analyzer Based on Tunable Filter Non-Dispersive Infrared Sensor: Application to the Monitoring of Eco-Friendly Gas Insulated Switchgears
    Kim, Yera
    Goo, Sun-Geun
    Lim, Jeong Sik
    SENSORS, 2022, 22 (22)