Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck

被引:298
作者
Cooper, H. John
Urban, Robert M.
Wixson, Richard L.
Meneghini, R. Michael
Jacobs, Joshua J.
机构
[1] Rush Univ, Med Ctr, Chicago, IL 60612 USA
[2] Northwestern Univ, Feinberg Sch Med, Chicago, IL 60611 USA
[3] Indiana Univ Sch Med, Indianapolis, IN 46202 USA
关键词
TOTAL HIP-ARTHROPLASTY; PSEUDOTUMOR FORMATION; FRETTING CORROSION; METAL RELEASE; PROSTHESES; FAILURE; INTERFACE; WEAR; REPLACEMENT; RETRIEVAL;
D O I
10.2106/JBJS.L.01042
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Femoral stems with dual-taper modularity were introduced to allow additional options for hip-center restoration independent of femoral fixation in total hip arthroplasty. Despite the increasing availability and use of these femoral stems, concerns exist about potential complications arising from the modular neck-body junction. Methods: This was a multicenter retrospective case series of twelve hips (eleven patients) with adverse local tissue reactions secondary to corrosion at the modular neck-body junction. The cohort included eight women and three men who together had an average age of 60.1 years (range, forty-three to seventy-seven years); all hips were implanted with a titanium-alloy stem and cobalt-chromium-alloy neck. Patients presented with new-onset and increasing pain at a mean of 7.9 months (range, five to thirteen months) following total hip arthroplasty. After serum metal-ion studies and metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) revealed abnormal results, the patients underwent hip revision at a mean of 15.2 months (range, ten to twenty-three months). Tissue specimens were examined by a single histopathologist, and the retrieved implants were studied with use of light and scanning electron microscopy. Results: Serum metal levels demonstrated greater elevation of cobalt (mean, 6.0 ng/mL) than chromium (mean, 0.6 ng/mL) or titanium (mean, 3.4 ng/mL). MRI with use of MARS demonstrated adverse tissue reactions in eight of nine patients in which it was performed. All hips showed large soft-tissue masses and surrounding tissue damage with visible corrosion at the modular femoral neck-body junction. Available histology demonstrated large areas of tissue necrosis in seven of ten cases, while remaining viable capsular tissue showed a dense lymphocytic infiltrate. Microscopic analysis was consistent with fretting and crevice corrosion at the modular neck-body interface. Conclusions: Corrosion at the modular neck-body junction in dual-tapered stems with a modular cobalt-chromium-alloy femoral neck can lead to release of metal ions and debris resulting in local soft-tissue destruction. Adverse local tissue reaction should be considered as a potential cause for new-onset pain in patients with these components, and early revision should be considered given the potentially destructive nature of these reactions. A workup including serologic studies (erythrocyte sedimentation rate and C-reactive protein), serum metal levels, and MARS MRI can be helpful in establishing this diagnosis.
引用
收藏
页码:865 / 872
页数:8
相关论文
共 56 条
[1]   A Comparison of Two Implant Systems in Restoration of Hip Geometry in Arthroplasty [J].
Archibeck, Michael J. ;
Cummins, Tamara ;
Carothers, Joshua ;
Junick, Daniel W. ;
White, Richard E., Jr. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) :443-446
[2]   Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[3]  
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[4]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[5]   FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[6]   Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips [J].
Campbell, Pat ;
Ebramzadeh, Edward ;
Nelson, Scott ;
Takamura, Karren ;
De Smet, Koen ;
Amstutz, Harlan C. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (09) :2321-2327
[7]  
COLLIER JP, 1995, CLIN ORTHOP RELAT R, P91
[8]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P305
[9]   CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[10]   CORROSION AND WEAR AT THE MODULAR INTERFACE OF UNCEMENTED FEMORAL STEMS [J].
COOK, SD ;
BARRACK, RL ;
CLEMOW, AJT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :68-72