Fracture of a titanium femoral neck after revision total hip arthroplasty

被引:1
作者
Zimmerer, Alexander [1 ,2 ]
Ogunleye, Philips [1 ]
Navas, Luis [1 ]
Hauschild, Matthias [1 ]
机构
[1] ARCUS Kliniken, Dept Orthopaed, Pforzheim, Germany
[2] Univ Med Greifswald, Dept Orthopaed & Orthopaed Surg, Greifswald, Germany
关键词
Prosthesis failure; Hip implants; DAMAGE; IMPLANTS; FAILURE;
D O I
10.1136/bcr-2021-247293
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a case of a fractured titanium alloy (Ti6Al4V) femoral stem 2 years after revision (debridement, antibiotics and implant retention due to periprosthetic joint infection) total hip arthroplasty (THA). Material analysis revealed a superficial abnormality on the upper neck surface caused by local melting with electrocautery probably acting as a stress amplifier. This lesion presumably was the trigger of a fatigue fracture with typical features on the fracture surface. Electrocautery should therefore be avoided in case of revision THA or used with utmost care to avoid contact with the femoral stem
引用
收藏
页数:4
相关论文
共 12 条
[1]  
Aepli Martin, 2019, JBJS Case Connect, V9, pe0418, DOI 10.2106/JBJS.CC.18.00418
[2]   Standardized Loads Acting in Hip Implants [J].
Bergmann, Georg ;
Bender, Alwina ;
Dymke, Joern ;
Duda, Georg ;
Damm, Philipp .
PLOS ONE, 2016, 11 (05)
[3]  
Depuy, 2017, AI EDAM, V14, P437
[4]   A modified transfemoral approach using modular cementless revision stems [J].
Fink, Bernd ;
Grossmann, Alexandra ;
Schubring, Svenja ;
Schulz, Martin S. ;
Fuerst, Martin .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (462) :105-114
[5]  
Gibon Emmanuel, 2020, Arthroplast Today, V6, P176, DOI 10.1016/j.artd.2020.04.004
[6]   Trends in Total Hip Arthroplasty in Germany from 2007 to 2016: What Has Changed and Where Are We Now? [J].
Klug, Alexander ;
Gramlich, Yves ;
Hoffmann, Reinhard ;
Pfeil, Joachim ;
Drees, Philipp ;
Kutzner, Karl Philipp .
ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2021, 159 (02) :173-180
[7]  
Konrads C, 2014, ORTHOPADE, V43, P1106, DOI 10.1007/s00132-014-3023-6
[8]   Factors influencing taper failure of modular revision hip stems [J].
Krull, A. ;
Morlock, M. M. ;
Bishop, N. E. .
MEDICAL ENGINEERING & PHYSICS, 2018, 54 :65-73
[9]   Electrosurgery Induced Damage to Ti-6Al-4V and CoCrMo Alloy Surfaces in Orthopedic Implants In Vivo and In Vitro [J].
Kubacki, Gregory W. ;
Sivan, Shiril ;
Gilbert, Jeremy L. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (11) :3533-3538
[10]   Electrocautery Damage Can Reduce Implant Fatigue Strength Cases and in Vitro Investigation [J].
Sonntag, Robert ;
Gibmeier, Jens ;
Pulvermacher, Samuel ;
Mueller, Ulrike ;
Eckert, Johannes ;
Braun, Steffen ;
Reichkendler, Markus ;
Kretzer, J. Philippe .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2019, 101 (10) :868-878