Reprocessing the bronchoscope: The challenges

被引:9
作者
Mughal, MM [1 ]
Minai, OA [1 ]
Culver, DA [1 ]
Mehta, AC [1 ]
机构
[1] Cleveland Clin Fdn, Dept Pulm & Crit Care Med, Cleveland, OH 44195 USA
关键词
flexible bronchoscope; disinfection; reprocessing; cleaning;
D O I
10.1055/s-2004-832717
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Spread of infection through the flexible bronchoscope is underrecognized and underreported. Prevention of bronchoscopy-induced infection requires increased vigilance by the physician, assiduous implementation of reprocessing protocols, and closer collaboration between bronchoscopy personnel, infection control practitioners, and instrument manufacturers. Patient safety depends on adequate disinfection of bronchoscopes and accessories used, as well as proper training of bronchoscopists, nurses, and ancillary staff. It is important to recognize that microbial transmission may occur via any part of instruments or anything in contact with the instruments including cleaning solutions, automated washers, and rinsing water. Numerous surveys have suggested poor adherence to published preventive guidelines. To address the challenges of reprocessing bronchoscopes, all users must comply with guidelines for cleaning and disinfection and each procedure should be performed with a clean, disinfected bronchoscope.
引用
收藏
页码:443 / 449
页数:7
相关论文
共 37 条
[1]   Importance of lumen flow in liquid chemical sterilization [J].
Alfa, MJ .
AMERICAN JOURNAL OF INFECTION CONTROL, 1999, 27 (04) :373-374
[2]  
Alvarado CJ, 2000, AM J INFECT CONTROL, V28, P138
[3]  
*AM SOC TEST MAT, 1994, STAND PRACT CLEAN DI
[4]  
[Anonymous], 2000, ERR IS HUMAN BUILDIN, DOI DOI 10.17226/9728
[5]   ENDOSCOPE DECONTAMINATION - AUTOMATED VS MANUAL [J].
BRADLEY, CR ;
BABB, JR .
JOURNAL OF HOSPITAL INFECTION, 1995, 30 :537-542
[6]  
*CDCP, 1991, MMWR-MORBID MORTAL W, V40, P675
[7]  
Cetre JC, 2003, NEW ENGL J MED, V348, P2039, DOI 10.1056/NEJM200305153482021
[8]   Bacteria-free water for automatic washer-disinfectors: an impossible dream? [J].
Cooke, RPD ;
Whymant-Morris, A ;
Umasankar, RS ;
Goddard, SV .
JOURNAL OF HOSPITAL INFECTION, 1998, 39 (01) :63-65
[9]   ASTHMA AND RHINITIS AFTER EXPOSURE TO GLUTARALDEHYDE IN ENDOSCOPY UNITS [J].
CORRADO, OJ ;
OSMAN, J ;
DAVIES, RJ .
HUMAN TOXICOLOGY, 1986, 5 (05) :325-327
[10]  
*CTR DEV RAD HLTH, 2002, FDA CLEAR STER HIGH