Anterior chamber angle measurement with optical coherence tomography:: Intraobserver and interobserver variability

被引:79
作者
Mueller, Maya
Dahmen, Gerlinde
Poerksen, Erk
Geerling, Gerd
Laqua, Horst
Ziegler, Andreas
Hoerauf, Hans
机构
[1] Univ Eye Clin, D-23538 Lubeck, Germany
[2] Inst Med Biometry & Stat, Lubeck, Germany
[3] Eye Clin, Kiel, Germany
[4] Univ Wurzburg, Eye Clin, Wurzburg, Germany
关键词
D O I
10.1016/j.jcrs.2006.07.014
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess intraobserver and interobserver variability of anterior segment optical coherence tomography (AS-OCT) as an objective diagnostic tool to quantify the anterior chamber angle and opening width. SETTING: University Eye Clinic, Lubeck, Germany. METHODS: The anterior chamber angle and opening width were assessed in 18 eyes of 9 healthy volunteers by 2 observers. Intraobserver reproducibility was evaluated by calculating an intraclass correlation coefficient (ICC) in a mixed model. Each observer had a separate model using information from 5 scans. Interobserver variability was determined by Bland-Altman analysis. The ICC was calculated in a mixed model using a residual maximum likelihood method. The results of 3 and 5 repeated scans were evaluated to indicate the change to 1 measurement application. RESULTS: The mean anterior chamber angle measurement was 35.9 degrees +/-5.7 (SD) for observer A and 36.2 +/- 5.7 degrees for observer B. The ICC was 0.94 and 0.91, respectively. The mean opening width was 315 +/- 62 mu m for observer A and 317 +/- 60 mu m for observer B. The ICC was 0.97 and 0.93, respectively. Interobserver comparisons showed a mean difference between anterior chamber angle measurements of -0.27 +/- 1.6 degrees, a limit of agreement (LOA) interval from -3.52 to 2.98 degrees, and an ICC estimate of 0.96. The mean difference in opening width measurements was 2.40 +/- 12.40 mu m, the LOA from -27.20 to 22.40 mu m, and the estimated ICC 0.96. Using I instead of the mean of 5 measurements, the LOA range increased by 3.46 degrees for the anterior chamber angle and 30.0 mu m for the opening width. CONCLUSION: Anterior chamber angle and opening width measurements by OCT showed low intraobserver and interobserver variability, indicating OCT is a valuable technique for quantitative assessment that provides reproducible measurements and objective documentation by different examiners.
引用
收藏
页码:1803 / 1808
页数:6
相关论文
共 34 条
[1]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
BOKER T, 1995, OPHTHALMIC RES, V27, P104
[4]  
BROWN H, 1999, APPL MIXED MODELS ME, P342
[5]  
FRANKE S, 1994, THESIS U HAMBURG
[6]   Anterior chamber width measurement by high speed optical coherence tomography [J].
Goldsmith, JA ;
Li, Y ;
Chalita, MR ;
Westphal, V ;
Patil, CA ;
Rollins, AM ;
Izatt, JA ;
Huang, D .
OPHTHALMOLOGY, 2005, 112 (02) :238-244
[7]  
HANNEMANN P, 1996, THESIS U HAMBURG
[8]  
Hoerauf H, 2000, OPHTHALMIC SURG LAS, V31, P218
[9]   Transscleral optical coherence tomography - An experimental study in ex-vivo human eyes [J].
Hoerauf, H ;
Winkler, J ;
Scholz, C ;
Wirbelauer, C ;
Gordes, RS ;
Koch, P ;
Engelhardt, R ;
Laqua, H ;
Birngruber, R .
LASERS IN SURGERY AND MEDICINE, 2002, 30 (03) :209-215
[10]  
Hoerauf H, 2002, ARCH OPHTHALMOL-CHIC, V120, P816