CT chest abdomen pelvis doses in Scotland: has the DRL had its day?

被引:10
作者
Sutton, D. G. [1 ]
Mcvey, S. [1 ]
Gentle, D. [2 ]
Hince, A. J. [3 ]
Macdonald, N. [4 ]
Mccallum, S. [5 ]
机构
[1] Ninewells Hosp, NHS Tayside, Dept Med Phys, Dundee DD1 9SY, Scotland
[2] NHS Greater Glasgow & Clyde, Gartnavel Hosp, Dept Med Phys, Glasgow, Lanark, Scotland
[3] Raigmore Hosp NHS Highland, Dept Med Phys, Inverness, Scotland
[4] Royal Infirm Edinburgh NHS Trust, NHS Lothian, Dept Med Phys, Edinburgh, Midlothian, Scotland
[5] Aberdeen Royal Infirm, NHS Grampian, Dept Med Phys, Aberdeen, Scotland
关键词
D O I
10.1259/bjr.20140157
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: This article reports on a pilot study designed to collect dose data representative of current CT chest abdomen pelvis (CAP) practice in Scotland, make any immediately obvious interventions and to identify if the current UK diagnostic reference level (DRL) of 940 mGy cm is still appropriate. The aims are to identify if a Scotland-wide picture archiving and communication system (PACS)-based dose audit of a number of CT examinations is likely to have value in terms of optimization of patient doses and to comment on the significance of the results in terms of future optimization strategies. Methods: Dose audit of CT CAP examinations at 32 different scanner sites across Scotland using accepted data collection and analysis methods. The minimum sample size was 30. Results: Results indicate that CT CAP doses are lower than those previously reported (median, 800 mGy cm, 75th percentile 840 mGy cm) but follow a distribution that is not in keeping with the concept of DRLs as presently understood or implemented. Conclusion: There is value in a PACS-based dose audit project to provide serial snapshots of patient doses as optimization efforts take place and to revise current knowledge about CT doses. In our opinion, the results call into question whether DRLs or the concept of "achievable dose" are suitable for devising optimization strategies once a certain degree of optimization has taken place. Advances in knowledge: The results reported here suggest that it may be time to take a different approach to optimization, concentrating on tools that are more refined than the DRL, which may have become more of a compliance tool than an aid to optimization.
引用
收藏
页数:8
相关论文
共 18 条
[1]  
[Anonymous], 1996, Ann ICRP, V26, P1
[2]  
[Anonymous], 2011, RAD PROT SAF RAD S 3
[3]  
Brenner D. J., 2012, Annals of the ICRP, V41, P161, DOI 10.1016/j.icrp.2012.06.004
[4]  
*DEP HLTH, 2000, GUID EST US DIAGN RE
[5]   Establishment of CT diagnostic reference levels in Ireland [J].
Foley, S. J. ;
McEntee, M. F. ;
Rainford, L. A. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1018) :1390-1397
[6]   Cancer risks from diagnostic radiology [J].
Hall, E. J. ;
Brenner, D. J. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (965) :362-378
[7]   Australian diagnostic reference levels for multi detector computed tomography [J].
Hayton, Anna ;
Wallace, Anthony ;
Marks, Paul ;
Edmonds, Keith ;
Tingey, David ;
Johnston, Peter .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2013, 36 (01) :19-26
[8]  
ICRP, 2007, ANN ICRP, V103, P1
[9]  
IPEM, 2004, 88 IPEM
[10]   Dose reduction in oncological staging multidetector CT: effect of iterative reconstruction [J].
Karpitschka, M. ;
Augart, D. ;
Becker, H-C ;
Reiser, M. ;
Graser, A. .
BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1021)