Pump speed modulations and sub-maximal exercise tolerance in left ventricular assist device recipients: A double-blind, randomized trial

被引:37
作者
Jung, Mette Holme [1 ]
Houston, Brian [2 ]
Russell, Stuart D. [2 ]
Gustafsson, Finn [1 ]
机构
[1] Rigshosp, Univ Hosp, Ctr Heart, Dept Cardiol, Blegdamsvej 9, DK-2100 Copenhagen, Denmark
[2] Johns Hopkins Univ Hosp, Dept Cardiol, Baltimore, MD 21287 USA
关键词
heart failure; LVAD; exercise capacity; pump speed; anaerobic threshold; HEART-FAILURE PATIENTS; QUALITY-OF-LIFE; WALK TEST; FLOW; CAPACITY; PERFORMANCE; PHYSIOLOGY; STANDARDS; DISTANCE; SUPPORT;
D O I
10.1016/j.healun.2016.06.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: The effect of pump speed increase on sub-maximal exercise tolerance, corresponding to activities of daily living (ADLs), is unknown. The aim of this study was to determine the effects of increasing pump speed during exercise at a sub-maximal level below anaerobic threshold (AT). METHODS: Patients each completed 3 exercise sessions on an ergometer cycle. On Day I workload at AT was defined. On Day 2 of the study, 2 sub-maximal tests at a workload below AT were undertaken: one at fixed baseline pump speed (Speed(base)) and the other with baseline pump speed + 800 rpm (Speed(inc)). The sequence of the 2 sub-maximal tests was determined by randomization. Both patient and physician were blinded to the sequence. Exercise duration, oxygen consumption (VO2) and rate of perceived exertion (RPE), using the Borg scale (score 6 to 20), were recorded. RESULTS: Nineteen patients (all with a HeartMate II ventricular assist device) completed 57 exercise tests. Baseline pump speed was 9,326 +/- 378 rpm. At AT, workload was 63 +/- 26 W (25 to 115 W) and VO2 was 79 +/- 14% of maximum. Exercise duration improved by 106 +/- 217 seconds (similar to 13%) in Speed(inc) compared with Speed(base) (837 +/- 358 vs 942 +/- 359 seconds; p = 0.048). The RPE was 13.2 +/- 2.5 in Speed(base) vs 12.7 +/- 2.4 in Speed(inc) (p = 0.2). CONCLUSION: Increasing pump speed by 800 rpm during sustained, low-intensity physical activity is safe and prolongs exercise duration in patients supported with a HeartMate II device. Automated pump speed increase during light exercise may contribute to improved quality of life by facilitating ADLs. (C) 2016 International Society for Heart and Lung Transplantation. All rights reserved.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 23 条
[1]   Quality of life and functional status in patients surviving 12 months after left ventricular assist device implantation [J].
Allen, Jeremiah G. ;
Weiss, Eric S. ;
Schaffer, Justin M. ;
Patel, Nishant D. ;
Ullrich, Susan L. ;
Russell, Stuart D. ;
Shah, Ashish S. ;
Conte, John V. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (03) :278-285
[2]   ON-LINE COMPUTER ANALYSIS AND BREATH-BY-BREATH GRAPHICAL DISPLAY OF EXERCISE FUNCTION TESTS [J].
BEAVER, WL ;
WASSERMAN, K ;
WHIPP, BJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1973, 34 (01) :128-132
[3]   Central and Peripheral Blood Flow During Exercise With a Continuous-Flow Left Ventricular Assist Device Constant Versus Increasing Pump Speed: A Pilot Study [J].
Brassard, Patrice ;
Jensen, Annette S. ;
Nordsborg, Nikolai ;
Gustafsson, Finn ;
Moller, Jacob E. ;
Hassager, Christian ;
Boesgaard, Soren ;
Hansen, Peter Bo ;
Olsen, Peter Skov ;
Sander, Kare ;
Secher, Niels H. ;
Madsen, Per Lav .
CIRCULATION-HEART FAILURE, 2011, 4 (05) :554-560
[4]   The 6-min walk distance in healthy subjects: reference standards from seven countries [J].
Casanova, C. ;
Celli, B. R. ;
Barria, P. ;
Casas, A. ;
Cote, C. ;
de Torres, J. P. ;
Jardim, J. ;
Lopez, M. V. ;
Marin, J. M. ;
Montes de Oca, M. ;
Pinto-Plata, V. ;
Aguirre-Jaime, A. .
EUROPEAN RESPIRATORY JOURNAL, 2011, 37 (01) :150-156
[5]   Reference values for the 6-min walk test in healthy subjects 20-50 years old [J].
Chetta, Alfredo ;
Zanini, Andrea ;
Pisi, Giovanna ;
Aiello, Marina ;
Tzani, Panagiota ;
Neri, Margherita ;
Olivieri, Dario .
RESPIRATORY MEDICINE, 2006, 100 (09) :1573-1578
[6]  
Guazzi M, JACC HEART FAIL
[7]   Usefulness of the Six-Minute Walk Test After Continuous Axial Flow Left Ventricular Device Implantation to Predict Survival [J].
Hasin, Tal ;
Topilsky, Yan ;
Kremers, Walter K. ;
Boilson, Barry A. ;
Schirger, John A. ;
Edwards, Brooks S. ;
Clavell, Alfredo L. ;
Rodeheffer, Richard J. ;
Frantz, Robert P. ;
Joyce, Lyle ;
Daly, Richard ;
Stulak, John M. ;
Kushwaha, Sudhir S. ;
Park, Soon J. ;
Pereira, Naveen L. .
AMERICAN JOURNAL OF CARDIOLOGY, 2012, 110 (09) :1322-1328
[8]   Assessing exercise capacity, quality of life and haemodynamics in heart failure: do the tests tell us the same thing? [J].
Houghton, AR ;
Harrison, M ;
Cowley, AJ ;
Hampton, JR .
EUROPEAN JOURNAL OF HEART FAILURE, 2002, 4 (03) :289-295
[9]   The impact of acute reduction of continuous-flow left ventricular assist device support on cardiac and exercise performance [J].
Jakovljevic, Djordje G. ;
George, Robert S. ;
Nunan, David ;
Donovan, Gay ;
Bougard, Robert S. ;
Yacoub, Magdi H. ;
Birks, Emma J. ;
Brodie, David A. .
HEART, 2010, 96 (17) :1390-1395
[10]   Relation Between Pressure and Volume Unloading During Ramp Testing in Patients Supported with a Continuous-Flow Left Ventricular Assist Device [J].
Jung, Mette H. ;
Hassager, Christian ;
Balling, Louise ;
Russell, Stuart D. ;
Boesgaard, Soeren ;
Gustafsson, Finn .
ASAIO JOURNAL, 2015, 61 (03) :307-312