Effect of fluidics on corneal endothelial cell density, central corneal thickness, and central macular thickness after phacoemulsification with torsional ultrasound

被引:10
作者
Das, Sudeep [1 ]
Nanaiah, Soumya Ganesh [1 ,3 ]
Kummelil, Mathew K. [1 ]
Nagappa, Somshekar [1 ]
Shetty, Rohit [2 ]
Shetty, Bhujang K. [1 ]
机构
[1] Narayana Nethralaya Super Specialty Eye Hosp, Dept Cataract & Refract Surg, Bengaluru, Karnataka, India
[2] Narayana Nethralaya Super Specialty Eye Hosp, Dept Cornea & Refract Surg, Bengaluru, Karnataka, India
[3] Lopamudra Med Ctr, Kodagu 571215, Karnataka, India
关键词
Fluidics; phacoemulsification; torsional; OPTICAL COHERENCE TOMOGRAPHY; CATARACT-SURGERY; EDEMA; PARAMETERS;
D O I
10.4103/0301-4738.169786
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To study the relative effects of high and low fluidic parameters on endothelial cell density (ECD), central corneal thickness (CCT), and central macular thickness (CMT) after phacoemulsification with torsional ultrasound. Settings and Design: Prospective, randomized clinical trial based on a tertiary eye hospital. Subjects and Methods: The study included 65 patients in each group. Patients were randomized to either the high or the low flow group using a computerized random number table. The study was patient and examiner masked. All patients underwent phacoemulsification with torsional ultrasound. Visual acuity, ECD, CCT, and CMT were measured for all patients preoperatively at 2 weeks and 6 weeks postoperatively. Statistical Analysis Used: The Shapiro-Wilks test was used to assess the normality of the data. Mann-Whitney U-test with the P value set at 0.05 was used to compare the two groups. Results: Cumulative dissipated energy was significantly higher in the low flow group (16.44 +/- 9.07 vs. 11.74 +/- 6.68; P = 0.002). No statistically significant difference was noted between the two groups in the ECD, CCT, CMT, or corrected distance visual acuity at the end of 6 weeks. Conclusions: No significant difference was noted in the postoperative outcome between high and low flow groups. Parameters can be modified to suit the surgeon's preference, as both high and low flow parameters were found to have comparable postoperative outcomes.
引用
收藏
页码:641 / 644
页数:4
相关论文
共 23 条
[1]   Effect of hydrodynamic parameters on corneal endothelial cell loss after phacoemulsification [J].
Baradaran-Rafii, Alireza ;
Rahmati-Kamel, Mohsen ;
Eslani, Medi ;
Kiavash, Victoria ;
Karimian, Farid .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (04) :732-737
[2]  
Cheng Bing, 2002, Zhonghua Yan Ke Za Zhi, V38, P265
[3]   Intraoperative performance and postoperative outcome comparison of longitudinal, torsional, and transversal phacoemulsification machines [J].
Christakis, Panos G. ;
Braga-Mele, Rosa M. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) :234-241
[4]   THE LENS OPACITIES CLASSIFICATION SYSTEM-III [J].
CHYLACK, LT ;
WOLFE, JK ;
SINGER, DM ;
LESKE, MC ;
BULLIMORE, MA ;
BAILEY, IL ;
FRIEND, J ;
MCCARTHY, D ;
WU, SY .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) :831-836
[5]   One-handed phacoemulsification with low settings [J].
Corydon, L ;
Krag, S ;
Thim, K .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (08) :1143-1148
[6]  
Debuc DC, 2010, MED SCI MONITOR, V16, pMT15
[7]   CYSTOID MACULAR EDEMA AND PAPILLEDEMA FOLLOWING CATARACT EXTRACTION [J].
GASS, JDM ;
NORTON, EWD .
ARCHIVES OF OPHTHALMOLOGY, 1966, 76 (05) :646-&
[8]   OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA [J].
HEE, MR ;
IZATT, JA ;
SWANSON, EA ;
HUANG, D ;
SCHUMAN, JS ;
LIN, CP ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) :325-332
[9]  
Hsieh FY, 1990, ADEQUACY SAMPLE SIZE, P36