Evaluation of Penalized Maximum-Likelihood PET Image Reconstruction for ROI Quantification

被引:0
|
作者
Yang, Li [1 ]
Zhou, Jian [1 ]
Asma, Evren [1 ]
Wang, Wenli [1 ]
机构
[1] Toshiba Med Res Inst USA Inc, Vernon Hills, IL USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Region of interest (ROI) quantification is one of the important tasks in positron emission tomography (PET) imaging. Statistical reconstruction methods based on the penalized maximum-likelihood (PML) principle have been developed to improve image quality over early stopping with EM-based methods and analytic methods. The penalty function that controls spatial smoothness plays an important role in PML reconstruction. The commonly used quadratic penalty often oversmooths sharp edges, while the edge-preserving penalty functions produce sharp intensity transitions across boundaries at the expense of potential image blockiness. Such blocky images limit the clinical widespread use of edge-prserving penalties. Recently patch-based edge-preserving penalties have been proposed for the purpose of preserving sharp edges and ensuring local smoothness while reducing image blockiness at the same time. The benefits of these penalty functions on clinical tasks had not been fully examined. In this work, we analyzed the ROI quantification performance for pairwise and patch-based penalties and also compared them with the standard ordered subsets expectation maximization (OSEM) reconstructions. We acquired a standard NEMA phantom scan and multiple whole-body patient scans with real lesions and also inserted lesions into the whole-body scans. The hot and cold sphere contrast recovery coefficient (CRC) versus the background variability was used to evaluate performance on the NEMA phantom. Mean lesion standardized uptake value (SUV) versus liver background variability was used for quantifying patient data. The results showed that the patch-based edge-preserving penalty achieved better quantification performance compared to other pairwise penalties and conventional OSEM.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Penalized maximum-likelihood image reconstruction for lesion detection
    Qi, Jinyi
    Huesman, Ronald H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (16): : 4017 - 4029
  • [2] Evaluation of penalty design in penalized maximum-likelihood image reconstruction for lesion detection
    Yang, Li
    Ferrero, Andrea
    Hagge, Rosalie J.
    Badawi, Ramsey D.
    Qi, Jinyi
    JOURNAL OF MEDICAL IMAGING, 2014, 1 (03)
  • [3] Evaluation of penalty design in penalized maximum-likelihood image reconstruction for lesion detection
    Yang, Li
    Ferrero, Andrea
    Hagge, Rosalie J.
    Badawi, Ramsey D.
    Qi, Jinyi
    MEDICAL IMAGING 2014: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2014, 9037
  • [4] Analysis of Penalized Likelihood Image Reconstruction for Dynamic PET Quantification
    Wang, Guobao
    Qi, Jinyi
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (04) : 608 - 620
  • [5] Theoretical Analysis of Penalized Maximum-Likelihood Patlak Parametric Image Reconstruction in Dynamic PET for Lesion Detection
    Yang, Li
    Wang, Guobao
    Qi, Jinyi
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (04) : 947 - 956
  • [6] Regularization design in penalized maximum-likelihood image reconstruction for lesion detection in 3D PET
    Yang, Li
    Zhou, Jian
    Ferrero, Andrea
    Badawi, Ramsey D.
    Qi, Jinyi
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (02): : 403 - 419
  • [7] THEORETICAL ANALYSIS OF LESION DETECTABILITY IN PENALIZED MAXIMUM-LIKELIHOOD PATLAK PARAMETRIC IMAGE RECONSTRUCTION USING DYNAMIC PET
    Yang, Li
    Wang, Guobao
    Qi, Jinyi
    2015 IEEE 12th International Symposium on Biomedical Imaging (ISBI), 2015, : 1188 - 1191
  • [8] Penalized maximum-likelihood image reconstruction for a two-panel PET scanner with image-based resolution modeling.
    Zhang, Hengquan
    Abbaszadeh, Shiva
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [9] Investigation of Image Resolution Effects In Maximum-Likelihood TOF PET Reconstruction
    Ispiryan, Michael
    Matej, Samuel
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2671 - 2678
  • [10] CHARACTERISTICS OF A MAXIMUM-LIKELIHOOD ALGORITHM FOR ITERATIVE PET IMAGE-RECONSTRUCTION
    BELLEMANN, ME
    SCHMIDLIN, P
    BRIX, G
    DOLL, J
    HABERKOM, U
    LORENZ, WJ
    RADIOLOGY, 1995, 197 : 223 - 223