Flexible Needle Steering in Moving Biological Tissue With Motion Compensation Using Ultrasound and Force Feedback

被引:31
作者
Chevrie, Jason [1 ,2 ]
Shahriari, Navid [3 ,4 ]
Babel, Marie [2 ,5 ]
Krupa, Alexandre [1 ,2 ]
Misra, Sarthak [3 ,4 ,6 ]
机构
[1] Univ Rennes, INRIA, CNRS, F-35000 Rennes, France
[2] IRISA, F-35000 Rennes, France
[3] Univ Twente, Dept Biomech Engn, Surg Robot Lab, NL-7522 LW Enschede, Netherlands
[4] Univ Groningen, Univ Med Ctr Groningen, Ctr Med Imaging North East Netherlands, NL-9713 AV Groningen, Netherlands
[5] Univ Rennes, INSA, CNRS, INRIA, F-35000 Rennes, France
[6] Univ Groningen, Univ Med Ctr Groningen, Dept Biomed Engn, NL-9713 AV Groningen, Netherlands
基金
欧洲研究理事会;
关键词
Surgical robotics; steerable catheters/needles; computer vision for medical robotics; medical robots and systems; SOFT-TISSUES; TRACKING; IMAGES; SURGERY; ROBOT;
D O I
10.1109/LRA.2018.2809484
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Needle insertion procedures under ultrasound guidance are commonly used for diagnosis and therapy. It is often critical to accurately reach a targeted region, and this can be difficult to achieve due to intraoperative tissue motion. In this letter, we present a method to steer a beveled-tip flexible needle toward a target embedded in moving tissue. Needle steering is performed using a needle insertion device attached to a robot arm. Closed-loop 3-D steering of the needle is achieved using tracking of an artificial target in 2-D ultrasound images and tracking of the needle tip position and orientation with an electromagnetic tracker. Tissue motion compensation is performed using force feedback to reduce targeting error and forces applied to the tissue. The method uses a mechanics-based interaction model that is updated online. A novel control law using task functions is proposed to fuse motion compensation, steering via base manipulation and tip-based steering. Validation of the tracking and steering algorithms are performed in gelatin phantom and bovine liver. Tissue motion up to 15 mm is applied and average targeting error is 1.2 +/- 0.8 mm and 2.5 +/- 0.7 mm in gelatin and liver, respectively, which is sufficiently accurate for commonly performed needle insertion procedures.
引用
收藏
页码:2338 / 2345
页数:8
相关论文
共 26 条
[1]   Experimental evaluation of ultrasound-guided 3D needle steering in biological tissue [J].
Abayazid, Momen ;
Vrooijink, Gustaaf J. ;
Patil, Sachin ;
Alterovitz, Ron ;
Misra, Sarthak .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2014, 9 (06) :931-939
[2]   Integrating Deflection Models and Image Feedback for Real-Time Flexible Needle Steering [J].
Abayazid, Momen ;
Roesthuis, Roy J. ;
Reilink, Rob ;
Misra, Sarthak .
IEEE TRANSACTIONS ON ROBOTICS, 2013, 29 (02) :542-553
[3]   Image-guided control of a robot for medical ultrasound [J].
Abolmaesumi, P ;
Salcudean, SE ;
Zhu, WH ;
Sirouspour, MR ;
DiMaio, SP .
IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 2002, 18 (01) :11-23
[4]  
Atashzar SF, 2013, IEEE INT CONF ROBOT, P1682, DOI 10.1109/ICRA.2013.6630796
[5]   Lung biopsy: Special techniques [J].
Cham, Matthew D. ;
Lane, Maureen E. ;
Henschke, Claudia I. ;
Yankelevitz, David F. .
SEMINARS IN RESPIRATORY AND CRITICAL CARE MEDICINE, 2008, 29 (04) :335-349
[6]  
Chatelain P, 2015, IEEE INT CONF ROBOT, P2250, DOI 10.1109/ICRA.2015.7139497
[7]  
Chatelain P, 2013, IEEE INT CONF ROBOT, P1676, DOI 10.1109/ICRA.2013.6630795
[8]  
Chevrie J, 2016, 2016 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS (IROS 2016), P2375, DOI 10.1109/IROS.2016.7759370
[9]  
Chevrie J, 2016, IEEE INT CONF ROBOT, P4450, DOI 10.1109/ICRA.2016.7487645
[10]   Needle steering and motion planning in soft tissues [J].
DiMaio, SP ;
Salcudean, SE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (06) :965-974