Evaluation of laboratory scanner accuracy by a novel calibration block for complete-arch implant rehabilitation

被引:15
作者
Pan, Yu [1 ]
Tam, Juliana M. Y. [2 ]
Tsoi, James K. H. [1 ]
Lam, Walter Y. H. [1 ]
Huang, Ruoxuan [3 ]
Chen, Zhuofan [3 ]
Pow, Edmond H. N. [1 ]
机构
[1] Univ Hong Kong, Fac Dent, Hong Kong, Peoples R China
[2] Hong Kong Polytech Univ, Ind Ctr, Hong Kong, Peoples R China
[3] Sun Yat Sen Univ, Guangdong Prov Key Lab Stornatol, Guanghua Sch Stomatol, Hosp Stomatol, Guangzhou, Peoples R China
关键词
Accuracy; Complete-arch; Laboratory scanner; Calibration block; Tilted implant; Inter-implant distance; EXTRAORAL DENTAL SCANNERS; IN-VITRO; 3-DIMENSIONAL ACCURACY; IMPRESSIONS; PRECISION;
D O I
10.1016/j.jdent.2020.103476
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare the accuracies of 4 laboratory scanners using a new custom-made block for complete-arch implant rehabilitation. Materials and Methods: A block comprised 4 cylinders, with 2 in the anterior (0 degrees angulation) and 2 in the posterior region (45 degrees distal angulation) (Experimental group) and a standard block with 2 parallel cones in 16 degrees taper (ISO group), were fabricated. Both blocks were scanned consecutively for 15 times by 4 laboratory scanners: IScan, Zfx, 3Shape, and KaVo. Measurements were also made by a coordinate measuring machine (CMM) as the reference. Acquired digital models were inspected with a metrology software. Linear and angular distortions were computed evaluation of trueness, precision and expanded uncertainty of scanners. Effects of blocks and scanners on the scanning trueness and precision were analyzed by Two-way ANOVA (alpha = 0.05). Results: The linear trueness and precision of scanners was significantly poorer when the Experimental block instead of ISO block was used. Significant greater distortions were noticed at the 45 degrees sites than the 0 degrees sites and no significant effect of inter-implant distance on the scanning accuracy was found. Zfx, IScan and 3Shape exhibited comparable expanded uncertainties (10.6 similar to 11.8) but KaVo showed the greatest (19.3) in complete-arch implant scans. Conclusion: The ISO block might not be more suitable than the Experimental block for evaluating the accuracy of laboratory scanner for complete-arch implant scanning. All the scanners tested except one demonstrated acceptable accuracy for complete-arch implant digitization. Scanning accuracy was compromised by unfavourable angulation of implants. Clinical significance: It is questionable whether the laboratory scanners validated for tooth-borne cases could also perform well for implant cases. It might be necessary to develop a new calibration object specifically for complete-arch implant scenarios to verify the capability of laboratory scanners in the implant workflow.
引用
收藏
页数:9
相关论文
共 24 条
  • [1] Digital Versus Conventional Impressions in Fixed Prosthodontics: A Review
    Ahlholm, Pekka
    Sipila, Kirsi
    Vallittu, Pekka
    Jakonen, Minna
    Kotiranta, Ulla
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2018, 27 (01): : 35 - 41
  • [2] [Anonymous], 2003, AM J PHYS, DOI DOI 10.1119/1.1522703
  • [3] In Vitro Three-Dimensional Accuracy of Digital Implant Impressions: The Effect of Implant Angulation
    Chia, Vanessa A.
    Esguerra, Roxanna J.
    Teoh, Khim Hean
    Teo, Juin Wei
    Wong, Keng Mun
    Tan, Keson B.
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (02) : 313 - 321
  • [4] Dahlmo KI, 2001, INT J PROSTHODONT, V14, P276
  • [5] A Novel Methodology to Validate the Accuracy of Extraoral Dental Scanners and Digital Articulation Systems
    Ellakwa, Ayman
    Elnajar, Sam
    Littlefair, Darren
    Sara, Georges
    [J]. EUROPEAN JOURNAL OF PROSTHODONTICS AND RESTORATIVE DENTISTRY, 2018, 26 (02) : 75 - 84
  • [6] Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
    Ender, Andreas
    Mehl, Albert
    [J]. JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2014, (86):
  • [7] The accuracy of different dental impression techniques for implant-supported dental prostheses: A systematic review and meta-analysis
    Fluegge, Tabea
    van der Meer, Wicher Joerd
    Gonzalez, Beatriz Gimenez
    Vach, Kirstin
    Wismeijer, Daniel
    Wang, Ping
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 : 374 - 392
  • [8] Structured-light 3D surface imaging: a tutorial
    Geng, Jason
    [J]. ADVANCES IN OPTICS AND PHOTONICS, 2011, 3 (02): : 128 - 160
  • [9] Accuracy of a Digital Impression System Based on Active Wavefront Sampling Technology for Implants Considering Operator Experience, Implant Angulation, and Depth
    Gimenez, Beatriz
    Oezcan, Mutlu
    Martinez-Rus, Francisco
    Pradies, Guillermo
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 : E54 - E64
  • [10] An In Vitro Study of Factors Influencing the Performance of Digital Intraoral Impressions Operating on Active Wavefront Sampling Technology with Multiple Implants in the Edentulous Maxilla
    Gimenez-Gonzalez, Beatriz
    Hassan, Bassam
    Ozcan, Mutlu
    Pradies, Guillermo
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (08): : 650 - 655