Electrical Stimulation Mapping With Stereo-EEG Electrodes

被引:39
作者
Britton, Jeffrey W. [1 ,2 ]
机构
[1] Mayo Clin, Epilepsy Div, Rochester, MN 55905 USA
[2] Mayo Clin, Neurol, Rochester, MN 55905 USA
关键词
Stereo-EEG; Cortical stimulation mapping; Epilepsy surgery; PARAMETERS;
D O I
10.1097/WNP.0000000000000443
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The practice of stereo-EEG (SEEG) is expanding. Electrical stimulation mapping remains an important part of the surgical evaluation process in SEEG, as it is in subdural electrocorticography cases. Because the technique and electrodes used in SEEG and electrocorticography are distinct, the clinician needs to be aware of the important differences between these techniques when performing electrical stimulation mapping. In this review, the advantages, disadvantages, and potential safety concerns related to electrical stimulation mapping in SEEG are discussed. A case is presented illustrating the utility of SEEG cortical stimulation and safety concerns that need to be considered when performing these evaluations.
引用
收藏
页码:110 / 114
页数:5
相关论文
共 8 条
[1]   Multimodal responses induced by cortical stimulation of the parietal lobe: a stereo-electroencephalography study [J].
Balestrini, Simona ;
Francione, Stefano ;
Mai, Roberto ;
Castana, Laura ;
Casaceli, Giuseppe ;
Marino, Daniela ;
Provinciali, Leandro ;
Cardinale, Francesco ;
Tassi, Laura .
BRAIN, 2015, 138 :2596-2607
[2]   Tissue damage thresholds during therapeutic electrical stimulation [J].
Cogan, Stuart F. ;
Ludwig, Kip A. ;
Welle, Cristin G. ;
Takmakov, Pavel .
JOURNAL OF NEURAL ENGINEERING, 2016, 13 (02)
[3]   PARAMETERS FOR DIRECT CORTICAL ELECTRICAL-STIMULATION IN THE HUMAN - HISTOPATHOLOGIC CONFIRMATION [J].
GORDON, B ;
LESSER, RP ;
RANCE, NE ;
HART, J ;
WEBBER, R ;
UEMATSU, S ;
FISHER, RS .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 75 (05) :371-377
[4]   Sulcal variability, stereological measurement and asymmetry of Broca's area on MR images [J].
Keller, Simon Sean ;
Highley, John Robin ;
Garcia-Finana, Marta ;
Sluming, Vanessa ;
Rezaie, Roozbeh ;
Roberts, Neil .
JOURNAL OF ANATOMY, 2007, 211 (04) :534-555
[5]   Selection of stimulus parameters for deep brain stimulation [J].
Kuncel, AM ;
Grill, WM .
CLINICAL NEUROPHYSIOLOGY, 2004, 115 (11) :2431-2441
[6]   Electrical Stimulations of the Human Insula: Their Contribution to the Ictal Semiology of Insular Seizures [J].
Mazzola, Laure ;
Mauguiere, Francois ;
Isnard, Jean .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2017, 34 (04) :307-314
[7]   CHARGE-DENSITY AND CHARGE PER PHASE AS COFACTORS IN NEURAL INJURY INDUCED BY ELECTRICAL-STIMULATION [J].
MCCREERY, DB ;
AGNEW, WF ;
YUEN, TGH ;
BULLARA, L .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (10) :996-1001
[8]   A MODEL OF SAFE LEVELS FOR ELECTRICAL-STIMULATION [J].
SHANNON, RV .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (04) :424-426