Comparison of 3 nucleic acid isolation methods for the quantification of HIV-1 RNA by Cobas Taqman real-time polymerase chain reaction system

被引:10
作者
Alp, Alpaslan [1 ]
Hascelik, Guelsen [1 ]
机构
[1] Hacettepe Univ, Fac Med, Dept Microbiol & Clin Microbiol, TR-06100 Ankara, Turkey
关键词
HIV-1; RNA; Cobas Taqman; Real-time PCR; MagNA Pure; Cobas AmpliPrep; IMMUNODEFICIENCY-VIRUS TYPE-1; MAGNA PURE; VIRAL LOAD; AUTOMATED QUANTIFICATION; ANTIRETROVIRAL THERAPY; VERSION; 1.5; EXTRACTION; PLASMA; ASSAYS; QUANTITATION;
D O I
10.1016/j.diagmicrobio.2008.12.014
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Quantification of human immunodeficiency virus type I (HIV-1) RNA viral load in blood is a clinically validated tool for the management and follow-up of patients infected with this virus. HIV-1 RNA quantification can be performed by several kinds of commercially available polymerase chain reaction (PCR) assays. The sensitivity of these assays is affected by the nucleic acid isolation procedures that are performed before PCR. Especially for the samples containing low numbers of HIV-1 RNA, ail effective nucleic acid isolation method is very important to obtain a high nucleic acid output. The purpose of this study was to compare and evaluate the conventional Manual extraction method, automated MagNA Pure system, and automated AmpliPrep system for the nucleic acid isolation before TaqMan real-time PCR assay for the quantification of HIV-1 RNA in plasma samples. Plasma samples of 40 patients in which anti-HIV antibodies were found as positive serologically were included in this study. Nucleic acid isolation from these samples was performed by manual High Pure Viral Nucleic Acid Isolation kit (Roche Diagnostics, Mannheim, Germany), automated MagNA Pure LC Total Nucleic Acid Isolation kit (Roche Diagnostics), and automated Cobas AmpliPrep kit (Roche Diagnostics). The nucleic acids obtained by these 3 methods were amplified and detected in the Cobas Taqman 48 (Roche Diagnostics) instrument. Viral RNA copies detected by AmpliPrep system, MagNA Pure system, and manual isolation method were found to be between 40 and 420 000 (average, 100 135), 53 and 1 200 000 (average, 220 554), and 183 and 590 000 (average, 104 761) copies/mL, respectively. The results of isolation methods were evaluated for each sample, and it was found out that the number of HIV-1 RNA obtained by MagNA Pure system was higher than the other 2 methods. All of the 3 methods were found to be reliable in detecting very small amounts of HIV-1 RNA in samples. Manual isolation method is a cheaper alternative, but because the nucleic acid isolation can be performed in a shorter time by automated systems when compared with manual isolation method, these systems can be preferred especially in the laboratories that have high number of samples. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:365 / 371
页数:7
相关论文
共 28 条
[1]  
Beral V, 2004, AIDS, V18, P51, DOI [10.1097/00002030-200401020-00006, 10.1097/01.aids.0000096908.73209.5d]
[2]  
Berger A, 2002, J CLIN VIROL, V25, pS103
[3]   Viral genome quantification as a tool for improving patient management: the example of HIV, HBV, HCV and CMV [J].
Berger, A ;
Preiser, W .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2002, 49 (05) :713-721
[4]   Comparison of four commercial quantitative HIV-1 assays for viral load monitoring in clinical daily routine [J].
Braun, Patrick ;
Ehret, Robert ;
Wiesmann, Frank ;
Zabbal, Frauke ;
Knickmann, Mechthild ;
Kuehn, Rainer ;
Thamm, Sven ;
Warnat, Gerald ;
Knechten, Heribert .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2007, 45 (01) :93-99
[5]   Comparative study of two methods for RNA extraction prior to detection of resistance to human immunodeficiency virus type 1 with the line probe assay [J].
Eiros, JM ;
Labayru, C ;
Hernández, B ;
de Lejarazu, RO ;
Torres, AR .
JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (06) :2257-2259
[6]  
Elbeik T, 2002, J ACQ IMMUN DEF SYND, V29, P330
[7]   Detection of herpes simplex virus DNA in genital and dermal specimens by LightCycler PCR after extraction using the IsoQuick, MagNA pure, and BioRobot 9604 methods [J].
Espy, MJ ;
Rys, PN ;
Wold, AD ;
Uhl, JR ;
Sloan, LM ;
Jenkins, GD ;
Ilstrup, DM ;
Cockerill, FR ;
Patel, R ;
Rosenblatt, JE ;
Smith, TF .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (06) :2233-2236
[8]   Comparison of the LCx human immunodeficiency virus (HIV) RNA quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan assays for quantitation of HIV type 1 RNA in plasma [J].
Foulongne, Vincent ;
Montes, Brigitte ;
Didelot-Rousseau, Marie-Noelle ;
Segondy, Michel .
JOURNAL OF CLINICAL MICROBIOLOGY, 2006, 44 (08) :2963-2966
[9]   Detection of HIV-1 proviral DNA with the AMPLICOR® HIV-1 DNA test, version 1.5, following sample processing by the MagNA Pure LC instrument [J].
Germer, Jeffrey J. ;
Gerads, Tara M. ;
Mandrekar, Jayawant N. ;
Mitchell, P. Shawn ;
Yao, Joseph D. C. .
JOURNAL OF CLINICAL VIROLOGY, 2006, 37 (03) :195-198
[10]   Discontinuation of potent antiretroviral therapy:: predictive value of and impact on CD4 cell counts and HIV RNA levels [J].
Grant, LA ;
Silverberg, MJ ;
Palacio, H ;
Minkoff, H ;
Anastos, K ;
Young, MA ;
Nowicki, M ;
Kovacs, A ;
Cohen, M ;
Muñoz, A .
AIDS, 2001, 15 (16) :2101-2108