Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting

被引:23
作者
Luukkonen, Antti A. M. [3 ,5 ]
Lehto, Tiina M. [1 ,2 ,3 ,4 ]
Hedberg, Pirjo S. M. [3 ,5 ]
Vaskivuo, Tommy E. [3 ,4 ,5 ]
机构
[1] Univ Oulu, Inst Diagnost, PL 5000, SF-90220 Oulu, Finland
[2] Oulu Univ Hosp, Norldlab Oulu, Oulu, Finland
[3] Univ Oulu, Dept Clin Chem, Inst Diagnost, SF-90220 Oulu, Finland
[4] Paijat Hame Social & Hlth Care Grp, Med Serv Ctr, Dept Clin Chem, Lahti, Finland
[5] Oulu Univ Hosp, Norldlab Oulu, Oulu, Finland
关键词
blood gas analysis; epoc; evaluation study; intensive care unit; point-of-care testing; POINT-OF-CARE; SYSTEM; PERFORMANCE; OUTCOMES;
D O I
10.1515/cclm-2015-0592
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Intensive care units, operating rooms, emergency departments, and neonatology units need rapid measurements of blood gases, electrolytes, and metabolites. These analyses can be performed in a central laboratory or at the clinic with traditional or compact cassettetype blood gas analyzers such as the epoc blood gas testing system for analyzing whole blood samples at the bedside. In this study, the performance and interchangeability of a hand-held epoc blood gas analyzer was evaluated. Methods: The analytical performance of the epoc analyzer was evaluated by determining within-and between-run precisions. The accuracy of the epoc analyzer was assessed by comparing patient results from the device with those obtained with the Siemens Rapidlab 1265 and Rapidpoint RP500 and Siemens Dimension Vista and Sysmex XE-2100 analyzers. The following parameters were measured: pH, pCO(2), pO(2), Hb (calc), Na+, K+, iCa(2+), glucose, and lactate. Results: The CV% of the epoc's between-day imprecision for the various parameters varied from 0.4 to 8.6. The within-run imprecision CV% varied from 0.6 to 5.2. The squared regression coefficient (R-2) between the epoc and RL1265 varied from 0.94 to 0.99, with the exception of Na+ and Ca2+ (R-2 >= 0.82). The correlation (R-2) of Na+ and K+ between epoc and Dimension Vista was 0.73 and 0.89, respectively. The correlation (R-2) of Hb between the epoc and the XE-2100 analyzer was 0.94. Conclusions: With most of the measured blood gas parameters, the epoc analyzer correlated well with reference techniques. The epoc analyzer is suitable for rapid measurement of the blood gases, the electrolytes, and the metabolites in the ICU.
引用
收藏
页码:585 / 594
页数:10
相关论文
共 19 条
[1]  
[Anonymous], EP BLOOD AN COST EFF
[2]  
[Anonymous], TIETZ TXB CLIN CHEM
[3]  
[Anonymous], EP SYST MAN
[4]  
Bland M. J., 1986, LANCET, V327, P307, DOI 10.1016/S0140-6736(86)90837-8
[5]   Analytical performance evaluation of four cartridge-type blood gas analyzers [J].
De Koninck, Anne-Sophie ;
De Decker, Koen ;
Van Bocxlaer, Jan ;
Meeus, Peter ;
Van Hoovels, Lieve .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (06) :1083-1091
[6]  
ERICKSON KA, 1993, CLIN CHEM, V39, P283
[7]   Comparison of lactate values between point-of-care and central laboratory analyzers [J].
Karon, Brad S. ;
Scott, Renee ;
Burritt, Mary F. ;
Santrach, Paula J. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 128 (01) :168-171
[8]   Cost analysis for decision support: The case of comparing centralized versus distributed methods for blood gas testing [J].
Kilgore, ML ;
Steindel, SJ ;
Smith, JA .
JOURNAL OF HEALTHCARE MANAGEMENT, 1999, 44 (03) :207-215
[9]   Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers [J].
Leino, Aila ;
Kurvinen, Kaisa .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (07) :1187-1191
[10]   Evaluation of the Enterprise Point-of-Care (EPOC) System for Blood Gas and Electrolyte Analysis [J].
Nichols, James H. ;
Rajadhyaksha, Aparna ;
Rodriguez, Mirian .
POINT OF CARE, 2008, 7 (01) :7-11