Influence of guided waves in bone on pulse-inversion contrast-enhanced ultrasound

被引:3
作者
Wang, Diya [1 ,2 ]
Zhang, Xinyu [1 ]
Sang, Yuchao [1 ]
Qu, Zhen [2 ]
Su, Qiang [3 ]
Zhao, Jing [4 ]
Wan, Mingxi [1 ]
机构
[1] Xi An Jiao Tong Univ, Sch Life Sci & Technol, Dept Biomed Engn, Xian 710049, Shaanxi, Peoples R China
[2] Univ Montreal, Hosp Res Ctr, Montreal, PQ H2X 0A9, Canada
[3] Capital Med Univ, Beijing Friendship Hosp, Dept Oncol, Beijing 100050, Peoples R China
[4] Jilin Univ, Hosp 2, Dept Ultrason Diag, Changchun 130041, Jilin, Peoples R China
基金
中国国家自然科学基金;
关键词
contrast-enhanced ultrasound; dispersion; guided waves; imaging quality; pulse-inversion; HIGH-RESOLUTION ULTRASOUND; GENERATION; TISSUE;
D O I
10.1002/mp.13634
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Guided waves generated from bone cortex inevitably act on microbubbles flowing through skeletal muscle capillaries in contrast-enhanced ultrasound (CEUS) and might influence the image quality. However, the action mechanism underlying the guided waves influence is still unknown, especially under contrast pulse-inversion transmission mode. This study aimed to clarify the influence of guided waves on pulse-inversion CEUS, which was investigated via in vitro infusion experiments. Method Tibia guided waves were detected at pulse-inversion transmission and then characterized by using a short-time Fourier transform energy distribution. Using results at normal incidence as a baseline, the influence of guided wave dispersion on the contrast and resolution of pulse-inversion CEUS was investigated at an oblique incidence through continuous microbubbles infusion experiments in a vessel-tibia flow phantom. Results Frequency-dispersive property of tibia guided waves was observed at phases 0 degrees and 180 degrees, which improved the contrast of CEUS and reduced its resolution. Pulse-inversion CEUS balanced the contrast enhancement and resolution degeneration induced by guided waves. By contrast, contrast-to-tissue ratio of pulse-inversion CEUS increased by up to 109.1 +/- 13.2% (P < 0.05) due to guided waves and its resolution was up to 0.9 +/- 0.1 times that of baseline. Conclusions Alterations of contrast and resolution in pulse-inversion CEUS induced by guided waves might provide an additional assessment for the capillary perfusion in the skeletal muscle near the bone cortex.
引用
收藏
页码:3475 / 3482
页数:8
相关论文
共 41 条
[1]  
[Anonymous], MED PHYS
[2]  
[Anonymous], IEEE INT ULTRA SYM
[3]  
[Anonymous], P M AC
[4]  
[Anonymous], P SENS MESSS
[5]   Generation of ultraharmonics in surfactant based ultrasound contrast agents: use and advantages [J].
Basude, R ;
Wheatley, MA .
ULTRASONICS, 2001, 39 (06) :437-444
[6]   Predicting bone strength with ultrasonic guided waves [J].
Bochud, Nicolas ;
Vallet, Quentin ;
Minonzio, Jean-Gabriel ;
Laugier, Pascal .
SCIENTIFIC REPORTS, 2017, 7
[7]  
Chang KV, 2012, J ULTRAS MED, V31, P737
[8]  
de Jong N, 2000, ULTRASON, P1869, DOI 10.1109/ULTSYM.2000.921689
[9]   Detection procedures of ultrasound contrast agents [J].
de Jong, N ;
Frinking, PJA ;
Bouakaz, A ;
Ten Cate, FJ .
ULTRASONICS, 2000, 38 (1-8) :87-92
[10]   Subharmonic imaging of contrast agents [J].
Forsberg, F ;
Shi, WT ;
Goldberg, BB .
ULTRASONICS, 2000, 38 (1-8) :93-98