Spectral and non-spectral interferences in the determination of selenium by inductively coupled plasma atomic emission spectrometry

被引:45
作者
Machat, J [1 ]
Otruba, V [1 ]
Kanicky, V [1 ]
机构
[1] Masaryk Univ, Fac Sci, Lab Atom Spectrochem, CZ-61137 Brno, Czech Republic
关键词
D O I
10.1039/b202167f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Spectral and non-spectral interferences observed in the determination of selenium by ICP-AES were investigated in this study. The identity of the carbon-induced emission peak that interferes on the Se 196.090 nm analytical line was clarified. It was found that this peak was not a part of a CN band but most likely belongs to a spectral band emitted by the CO molecule. Non-spectral interferences of matrix elements C, S, P and Br assisting in the enhancement of selenium emission were found. Concentration dependences of these matrix effects were measured in the range of 0.5-20 g L-1 of the matrix element and 0.1-100 mg L-1 of Se. No enhancement of the Mg I 285.213 nm line was observed when the influence of these matrix elements on magnesium was monitored for comparison at the same time. The enhancement of selenium emission was about 50% in the most significant case (sulfur, 10 g L-1). A similar effect was observed when gaseous compounds of C and S were introduced into the ICP. The mechanism of this matrix effect was studied and partially clarified as being probably related to ionization of matrix elements.
引用
收藏
页码:1096 / 1102
页数:7
相关论文
共 14 条
[1]   Enzymatic digestion for the determination of trace elements in blood serum by inductively coupled plasma mass spectrometry [J].
AbouShakra, FR ;
Rayman, MP ;
Ward, NI ;
Hotton, V ;
Bastian, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) :429-433
[2]   ELECTRONIC EXCITATION AND IONIZATION TEMPERATURE-MEASUREMENTS IN A HIGH-FREQUENCY INDUCTIVELY COUPLED ARGON PLASMA SOURCE AND THE INFLUENCE OF WATER-VAPOR ON PLASMA PARAMETERS [J].
ALDER, JF ;
BOMBELKA, RM ;
KIRKBRIGHT, GF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1980, 35 (04) :163-175
[3]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[4]   DETERMINATION OF TOTAL ARSENIC CONCENTRATIONS IN BIOLOGICAL MATRICES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
CAMPBELL, MJ ;
DEMESMAY, C ;
OLLE, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1379-1384
[5]   Determination of selenium in urine by inductively coupled plasma mass spectrometry: interferences and optimization [J].
Gammelgaard, B ;
Jons, O .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (05) :867-874
[6]  
Knapp G, 1998, ATOM SPECTROSC, V19, P220
[7]   CARBON-ENHANCED INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRIC DETECTION OF ARSENIC AND SELENIUM AND ITS APPLICATION TO ARSENIC SPECIATION [J].
LARSEN, EH ;
STURUP, S .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (10) :1099-1105
[8]   Determination of selenium in blood serum by inductively coupled plasma atomic emission spectrometry with pneumatic nebulization [J].
Machát, J ;
Kanicky, V ;
Otruba, V .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 372 (04) :576-581
[9]  
MARTIN WC, 2001, NIST ATOMIC SPECTRA
[10]  
Montaser A., 1992, INDUCTIVELY COUPLED