Glucose Sensing in the Peritoneal Space Offers Faster Kinetics Than Sensing in the Subcutaneous Space

被引:41
作者
Burnett, Daniel R. [1 ]
Huyett, Lauren M. [2 ]
Zisser, Howard C. [2 ,3 ]
Doyle, Francis J., III [2 ]
Mensh, Brett D. [1 ,4 ]
机构
[1] Theranova LLC, San Francisco, CA 94107 USA
[2] Univ Calif Santa Barbara, Dept Chem Engn, Santa Barbara, CA 93106 USA
[3] Sansum Diabet Res Inst, Santa Barbara, CA USA
[4] Howard Hughes Med Inst, Ashburn, VA USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
CARBON-DIOXIDE TENSIONS; NOCTURNAL HYPOGLYCEMIA; ADIPOSE-TISSUE; MICRODIALYSIS; BLOOD; OXYGEN;
D O I
10.2337/db13-1649
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The paramount goal in the treatment of type 1 diabetes is the maintenance of normoglycemia. Continuous glucose monitoring (CGM) technologies enable frequent sensing of glucose to inform exogenous insulin delivery timing and dosages. The most commonly available CGMs are limited by the physiology of the subcutaneous space in which they reside. The very same advantages of this minimally invasive approach are disadvantages with respect to speed. Because subcutaneous blood flow is sensitive to local fluctuations (e. g., temperature, mechanical pressure), subcutaneous sensing can be slow and variable. We propose the use of a more central, physiologically stable body space for CGM: the intraperitoneal space. We compared the temporal response characteristics of simultaneously placed subcutaneous and intraperitoneal sensors during intravenous glucose tolerance tests in eight swine. Using compartmental modeling based on simultaneous intravenous sensing, blood draws, and intraarterial sensing, we found that intraperitoneal kinetics were more than twice as fast as subcutaneous kinetics (mean time constant of 5.6 min for intraperitoneal vs. 12.4 min for subcutaneous). Combined with the known faster kinetics of intraperitoneal insulin delivery over subcutaneous delivery, our findings suggest that artificial pancreas technologies may be optimized by sensing glucose and delivering insulin in the intraperitoneal space.
引用
收藏
页码:2498 / 2505
页数:8
相关论文
共 46 条
[1]   ARTIFICIAL ENDOCRINE PANCREAS [J].
ALBISSER, AM ;
LEIBEL, BS ;
EWART, TG ;
DAVIDOVAC, Z ;
BOTZ, CK ;
ZINGG, W .
DIABETES, 1974, 23 (05) :389-396
[2]   Hypoglycemia in the diabetes control and complications trial [J].
不详 .
DIABETES, 1997, 46 (02) :271-286
[3]  
[Anonymous], DEC CONTR CDC 2013 5
[4]   Sleep-related hypoglycemia-associated autonomic failure in type 1 diabetes - Reduced awakening from sleep during hypoglycemia [J].
Banarer, S ;
Cryer, PE .
DIABETES, 2003, 52 (05) :1195-1203
[5]   Management of occlusion and thrombosis associated with long-term indwelling central venous catheters [J].
Baskin, Jacquelyn L. ;
Pui, Ching-Hon ;
Reiss, Ulrike ;
Wilimas, Judith A. ;
Metzger, Monika L. ;
Ribeiro, Raul C. ;
Howard, Scott C. .
LANCET, 2009, 374 (9684) :159-169
[6]   Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans [J].
Basu, Ananda ;
Dube, Simmi ;
Slama, Michael ;
Errazuriz, Isabel ;
Amezcua, Jose Carlos ;
Kudva, Yogish C. ;
Peyser, Thomas ;
Carter, Rickey E. ;
Cobelli, Claudio ;
Basu, Rita .
DIABETES, 2013, 62 (12) :4083-4087
[7]  
BOQUIST L, 1969, ACTA PATHOL MIC SC, V76, P215
[8]   Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor [J].
Boyne, MS ;
Silver, DM ;
Kaplan, J ;
Saudek, CD .
DIABETES, 2003, 52 (11) :2790-2794
[9]  
Breton Marc, 2008, J Diabetes Sci Technol, V2, P853
[10]   A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring [J].
Cengiz, Eda ;
Tamborlane, William V. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 :S11-S16