A rapid HPLC-MS/MS method for the simultaneous quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human blood samples

被引:96
作者
Seger, Christoph [1 ]
Tentschert, Karin [1 ]
Stoeggl, Wolfgang [2 ]
Griesmacher, Andrea [1 ]
Ramsay, Steven L. [2 ]
机构
[1] Univ Innsbruck Hosp, Inst Med & Chem Lab Diagnost ZIMCL, Div Mass Spectrometry & Chromatog, A-6020 Innsbruck, Austria
[2] Biocrates Life Sci AG, A-6020 Innsbruck, Austria
关键词
TANDEM-MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; LC-MS; PERFORMANCE; TDM; IMMUNOSUPPRESSANTS; EXTRACTION; ASSAYS; IMMUNOASSAYS; SPECIFICITY;
D O I
10.1038/nprot.2009.25
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In the past few years, high-performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) has matured to a true alternative to antibody-based immunoassays in routine therapeutic drug monitoring. In transplantation medicine, mass spectrometry-based assessment of immunosuppressant drug levels is considered a gold standard diagnostic procedure. We describe a fast state-of-the-art routine online solid-phase extraction (SPE) HPLC-MS/MS analysis platform that allows monitoring of cyclosporine A, tacrolimus, sirolimus and everolimus from 50-mu l aliquots of EDTA whole blood specimens within 3.4 min total analysis time. Sample purification is done by offline protein precipitation followed by two automated chromatographic separation steps. Mass spectrometry-based analyte quantification relies on selected reaction monitoring experiments. The assay underwent complete validation and performance evaluation studies and performs very well in several international proficiency testing schemes. In daily routine, it allows reporting of about 75 patient sample results per work shift with a typical total individual sample turnaround time of less than 3 h.
引用
收藏
页码:526 / 534
页数:9
相关论文
共 40 条
[1]   Methanol-associated matrix effects in electrospray ionization tandem mass spectrometry [J].
Annesley, Thomas M. .
CLINICAL CHEMISTRY, 2007, 53 (10) :1827-1834
[2]   Simple extraction protocol for analysis of immunosuppressant drugs in whole blood [J].
Annesley, TM ;
Clayton, L .
CLINICAL CHEMISTRY, 2004, 50 (10) :1845-1848
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Rapid simultaneous quantification of immunosuppressants in transplant patients by turbulent flow chromatography combined with tandem mass spectrometry [J].
Ceglarek, U ;
Lembcke, J ;
Fiedler, GM ;
Werner, M ;
Witzigmann, H ;
Hauss, JP ;
Thiery, J .
CLINICA CHIMICA ACTA, 2004, 346 (02) :181-190
[5]   Mass spectrometry in the clinical laboratory [J].
Chace, DH .
CHEMICAL REVIEWS, 2001, 101 (02) :445-477
[6]   Automated, fast and sensitive quantification of drugs in blood by liquid chromatography-mass spectrometry with on-line extraction: immunosuppressants [J].
Christians, U ;
Jacobsen, W ;
Serkova, N ;
Benet, LZ ;
Vidal, C ;
Sewing, KF ;
Manns, MP ;
Kirchner, GI .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 748 (01) :41-53
[7]   Mass spectrometry-based biochemical assays for enzymeinhibitor screening [J].
de Boer, Arjen R. ;
Lingeman, Henk ;
Niessen, Wilfried M. A. ;
Irth, Hubertus .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2007, 26 (09) :867-883
[8]   Simultaneous quantification of sirolimus, everolimus, tacrolimus and cyclosporine by liquid chromatography-mass spectrometry (LC-MS) [J].
Deters, M ;
Kirchner, G ;
Resch, K ;
Kaever, V .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (03) :285-292
[9]   Reduced exposure to calcineurin inhibitors in renal transplantation [J].
Ekberg, Henrik ;
Tedesco-Silva, Helio ;
Demirbas, Alper ;
Vitko, Stefan ;
Nashan, Bjorn ;
Guerkan, Alp ;
Margreiter, Raimund ;
Hugo, Christian ;
Grinyo, Josep M. ;
Frei, Ulrich ;
Vanrenterghem, Yves ;
Daloze, Pierre ;
Halloran, Philip F. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (25) :2562-2575
[10]   Biochemical applications of mass spectrometry in pharmaceutical drug discovery [J].
Geoghegan, KF ;
Kelly, MA .
MASS SPECTROMETRY REVIEWS, 2005, 24 (03) :347-366