Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects

被引:24
作者
Guler, Mete [1 ]
Bilak, Semsettin [1 ]
Bilgin, Burak [2 ]
Simsek, Ali [1 ]
Capkin, Musa [1 ]
Reyhan, Ali Hakim [1 ]
机构
[1] Adiyaman Univ, Sch Med, Dept Ophthalmol, TR-02040 Adiyaman, Turkey
[2] Adiyaman Univ, Dept Ophthalmol, Educ & Res Hosp, Adiyaman, Turkey
关键词
central corneal thickness; Goldmann applanation tonometer; intraocular pressure; noncontact tonometer; rebound tonometer; CENTRAL CORNEAL THICKNESS; ACCURACY; RELIABILITY;
D O I
10.1097/IJG.0000000000000132
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To compare intraocular pressure (IOP) measurements obtained by Icare PRO rebound tonometer (RT), Tomey FT-1000 noncontact tonometer (NCT), and Goldmann applanation tonometer (GAT) in healthy subjects, and to investigate the influence of central corneal thickness (CCT) on IOP measurements.Methods:A total of 132 eyes of 66 healthy subjects were included in the study. All IOP measurements were applied in a sitting position and always in the same order (RT, NCT, and GAT). CCT was measured by noncontact anterior segment optical coherence tomography.Results:The mean CCT was 523.531.2 m. The mean IOPs with GAT, NCT, and RT measurements were 14.56 +/- 2.77, 13.42 +/- 2.99, 14.18 +/- 2.55 mm Hg, respectively. A significantly good agreement was found between RT and GAT measurements (P=0.515). The mean difference between RT and GAT measurements was -0.38 mm Hg. An underestimation was found in NCT measurements compared with GAT ones as defined from paired comparisons (P=0.003). The mean difference between NCT and GAT measurements was -1.14 mm Hg. There was no significant difference between NCT and RT readings (P=0.069). The mean difference between NCT and RT measurements was -0.76 mm Hg. There was a significant correlation between GAT-NCT (r=0.740, P<0.001), GAT-RT (r=0.743, P<0.001), NCT-RT (r=0.729, P<0.001) measurement methods. The increment in measured IOP for a 10 m increase in CCT was 0.16, 0.33, and 0.18 mm Hg for the GAT, NCT, and RT, respectively.Conclusions:The IOP readings with the new Icare PRO RT showed good correlation with those obtained by GAT. Tomey FT-1000 NCT underestimated the IOP compared with GAT.
引用
收藏
页码:613 / 618
页数:6
相关论文
共 30 条
[21]   Accuracy and reliability of the Keeler Pulsair EasyEye non-contact tonometer [J].
Ogbuehi, Kelechi C. ;
Almubrad, Turki M. .
OPTOMETRY AND VISION SCIENCE, 2008, 85 (01) :61-66
[22]   Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer [J].
Ogbuehi, Kelechi C. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2006, 89 (05) :310-314
[23]   Clinical comparison of the Icare tonometer and Goldmann applanation tonometry [J].
Pakrou, Nima ;
Gray, Tim ;
Mills, Richard ;
Landers, John ;
Craig, Jamie .
JOURNAL OF GLAUCOMA, 2008, 17 (01) :43-47
[24]   The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness [J].
Poostchi, Ali ;
Mitchell, Robert ;
Nicholas, Simon ;
Purdie, Gordon ;
Wells, Anthony .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 37 (07) :687-691
[25]   Rebound, Applanation, and Dynamic Contour Tonometry in Pathologic Corneas [J].
Rosentreter, Andre ;
Athanasopoulos, Apostolos ;
Schild, Andrea M. ;
Lappas, Alexandra ;
Cursiefen, Claus ;
Dietlein, Thomas S. .
CORNEA, 2013, 32 (03) :313-318
[26]   A new rebound tonometer for home monitoring of intraocular pressure [J].
Rosentreter, Andre ;
Jablonski, Kerstin S. ;
Mellein, Anne C. ;
Gaki, Stergiani ;
Hueber, Arno ;
Dietlein, Thomas S. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 249 (11) :1713-1719
[27]   Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma Patients [J].
Salim, Sarwat ;
Du, Haiming ;
Wan, Jim .
JOURNAL OF GLAUCOMA, 2013, 22 (04) :325-329
[28]   A History of Intraocular Pressure and Its Measurement [J].
Stamper, Robert L. .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (01) :E16-E28
[29]   The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, noncontact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry [J].
Tonnu, PA ;
Ho, T ;
Newson, T ;
El Sheikh, A ;
Sharma, K ;
White, E ;
Bunce, C ;
Garway-Heath, D .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (07) :851-854
[30]   Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients [J].
Vincent, Stephen J. ;
Vincent, Roslyn A. ;
Shields, David ;
Lee, Graham A. .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (04) :E163-E170