Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects
被引:24
作者:
Guler, Mete
[1
]
Bilak, Semsettin
[1
]
Bilgin, Burak
[2
]
Simsek, Ali
[1
]
Capkin, Musa
[1
]
Reyhan, Ali Hakim
[1
]
机构:
[1] Adiyaman Univ, Sch Med, Dept Ophthalmol, TR-02040 Adiyaman, Turkey
[2] Adiyaman Univ, Dept Ophthalmol, Educ & Res Hosp, Adiyaman, Turkey
关键词:
central corneal thickness;
Goldmann applanation tonometer;
intraocular pressure;
noncontact tonometer;
rebound tonometer;
CENTRAL CORNEAL THICKNESS;
ACCURACY;
RELIABILITY;
D O I:
10.1097/IJG.0000000000000132
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose:To compare intraocular pressure (IOP) measurements obtained by Icare PRO rebound tonometer (RT), Tomey FT-1000 noncontact tonometer (NCT), and Goldmann applanation tonometer (GAT) in healthy subjects, and to investigate the influence of central corneal thickness (CCT) on IOP measurements.Methods:A total of 132 eyes of 66 healthy subjects were included in the study. All IOP measurements were applied in a sitting position and always in the same order (RT, NCT, and GAT). CCT was measured by noncontact anterior segment optical coherence tomography.Results:The mean CCT was 523.531.2 m. The mean IOPs with GAT, NCT, and RT measurements were 14.56 +/- 2.77, 13.42 +/- 2.99, 14.18 +/- 2.55 mm Hg, respectively. A significantly good agreement was found between RT and GAT measurements (P=0.515). The mean difference between RT and GAT measurements was -0.38 mm Hg. An underestimation was found in NCT measurements compared with GAT ones as defined from paired comparisons (P=0.003). The mean difference between NCT and GAT measurements was -1.14 mm Hg. There was no significant difference between NCT and RT readings (P=0.069). The mean difference between NCT and RT measurements was -0.76 mm Hg. There was a significant correlation between GAT-NCT (r=0.740, P<0.001), GAT-RT (r=0.743, P<0.001), NCT-RT (r=0.729, P<0.001) measurement methods. The increment in measured IOP for a 10 m increase in CCT was 0.16, 0.33, and 0.18 mm Hg for the GAT, NCT, and RT, respectively.Conclusions:The IOP readings with the new Icare PRO RT showed good correlation with those obtained by GAT. Tomey FT-1000 NCT underestimated the IOP compared with GAT.
引用
收藏
页码:613 / 618
页数:6
相关论文