Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects

被引:24
作者
Guler, Mete [1 ]
Bilak, Semsettin [1 ]
Bilgin, Burak [2 ]
Simsek, Ali [1 ]
Capkin, Musa [1 ]
Reyhan, Ali Hakim [1 ]
机构
[1] Adiyaman Univ, Sch Med, Dept Ophthalmol, TR-02040 Adiyaman, Turkey
[2] Adiyaman Univ, Dept Ophthalmol, Educ & Res Hosp, Adiyaman, Turkey
关键词
central corneal thickness; Goldmann applanation tonometer; intraocular pressure; noncontact tonometer; rebound tonometer; CENTRAL CORNEAL THICKNESS; ACCURACY; RELIABILITY;
D O I
10.1097/IJG.0000000000000132
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To compare intraocular pressure (IOP) measurements obtained by Icare PRO rebound tonometer (RT), Tomey FT-1000 noncontact tonometer (NCT), and Goldmann applanation tonometer (GAT) in healthy subjects, and to investigate the influence of central corneal thickness (CCT) on IOP measurements.Methods:A total of 132 eyes of 66 healthy subjects were included in the study. All IOP measurements were applied in a sitting position and always in the same order (RT, NCT, and GAT). CCT was measured by noncontact anterior segment optical coherence tomography.Results:The mean CCT was 523.531.2 m. The mean IOPs with GAT, NCT, and RT measurements were 14.56 +/- 2.77, 13.42 +/- 2.99, 14.18 +/- 2.55 mm Hg, respectively. A significantly good agreement was found between RT and GAT measurements (P=0.515). The mean difference between RT and GAT measurements was -0.38 mm Hg. An underestimation was found in NCT measurements compared with GAT ones as defined from paired comparisons (P=0.003). The mean difference between NCT and GAT measurements was -1.14 mm Hg. There was no significant difference between NCT and RT readings (P=0.069). The mean difference between NCT and RT measurements was -0.76 mm Hg. There was a significant correlation between GAT-NCT (r=0.740, P<0.001), GAT-RT (r=0.743, P<0.001), NCT-RT (r=0.729, P<0.001) measurement methods. The increment in measured IOP for a 10 m increase in CCT was 0.16, 0.33, and 0.18 mm Hg for the GAT, NCT, and RT, respectively.Conclusions:The IOP readings with the new Icare PRO RT showed good correlation with those obtained by GAT. Tomey FT-1000 NCT underestimated the IOP compared with GAT.
引用
收藏
页码:613 / 618
页数:6
相关论文
共 30 条
[1]   A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes [J].
Babalola, O. E. ;
Kehinde, A. V. ;
Iloegbunam, A. C. ;
Akinbinu, T. ;
Moghalu, C. ;
Onuoha, I. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2009, 29 (02) :182-188
[2]  
Brandt James D, 2004, Curr Opin Ophthalmol, V15, P85, DOI 10.1097/00055735-200404000-00004
[3]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[4]   Goldmann applanation tonometry compared with corneal-compensated intraocular pressure in the evaluation of primary open-angle Glaucoma [J].
Ehrlich, Joshua R. ;
Radcliffe, Nathan M. ;
Shimmyo, Mitsugu .
BMC OPHTHALMOLOGY, 2012, 12
[5]   Home Tonometry for Management of Pediatric Glaucoma [J].
Flemmons, Meghan S. ;
Hsiao, Ya-Chuan ;
Dzau, Jacqueline ;
Asrani, Sanjay ;
Jones, Sarah ;
Freedman, Sharon F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2011, 152 (03) :470-478
[6]   Icare ONE Rebound Versus Goldmann Applanation Tonometry in Children With Known or Suspected Glaucoma [J].
Gandhi, Nandini G. ;
Prakalapakorn, Sasapin G. ;
El-Dairi, Mays A. ;
Jones, Sarah K. ;
Freedman, Sharon F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 154 (05) :843-849
[7]   Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes [J].
García-Resúa, C ;
González-Meijome, JM ;
Gilino, J ;
Yebra-Pimentel, E .
OPTOMETRY AND VISION SCIENCE, 2006, 83 (02) :102-107
[8]   The use of the Icare tonometer reduced the need for anesthesia to measure intraocular pressure in children [J].
Grigorian, Florin ;
Grigorian, A. Paula ;
Olitsky, Scott E. .
JOURNAL OF AAPOS, 2012, 16 (06) :508-510
[9]   Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects [J].
Halkiadakis, Ioannis ;
Stratos, Aimilianos ;
Stergiopoulos, George ;
Patsea, Eleni ;
Skouriotis, Sotiris ;
Mitropoulos, Panagiotis ;
Papaconstantinou, Dimitrios ;
Georgopoulos, Gerasimos .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 250 (08) :1207-1211
[10]   The Influence of Central Corneal Thickness and Corneal Curvature Radius on The Intraocular Pressure as Measured By Different Tonometers: Noncontact and Goldmann Applanation Tonometers [J].
Harada, Yukinori ;
Hirose, Naofumi ;
Kubota, Toshiaki ;
Tawara, Akihiko .
JOURNAL OF GLAUCOMA, 2008, 17 (08) :619-625