Towards large-area photon-counting detectors for spectral x-ray imaging

被引:2
|
作者
Thuering, Thomas [1 ]
Gkoumas, Spyridon [1 ]
Zambon, Pietro [1 ]
Trueb, Peter [1 ]
Rissi, Michael [1 ]
Di Prima, Flavio [1 ]
Donath, Tilman [1 ]
Sawall, Stefan [2 ]
Amato, Carlo [2 ]
Kachelriess, Marc [2 ]
Broennimann, Christian [1 ]
机构
[1] DECTRIS Ltd, Tafernweg 1, CH-5405 Baden, Switzerland
[2] German Canc Res Ctr, Neuenheimer Feld 280, D-69120 Heidelberg, Germany
来源
MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING | 2019年 / 10948卷
关键词
Photon-counting detectors; CdTe; spectral X-ray imaging; large-area detector; material decomposition;
D O I
10.1117/12.2507190
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon-counting detector technology using directly converting high-Z sensor materials has recently gained popularity in medical imaging due to its capability to reduce patient dose, increase spatial resolution and provide single shot multi energy information. However, in medical imaging applications, the novel technology is not yet widely used due to technical challenges in manufacturing gap-less detectors with large areas. Here, a nearly gap-less, large-area, multi-energy photon counting detector prototype is presented which was built with existing ASIC and sensor technology. It features an active area of 8x8 cm(2), a 1 mm thick cadmium telluride (CdTe) sensor, four independent energy thresholds and a pixel size of 150 lam. Single and multi-threshold imaging performance of the detector is evaluated by assessing various metrics relevant for conventional (polychromatic) and spectral imaging applications. A high detective quantum efficiency (DQE(0)=0.97), a low dark noise threshold (6.5 keV) and a high count rate capability (up to 3x10(8) counts/s/mm(2)) indicate that the detector is optimally suited for conventional medical X-ray imaging, especially for low dose applications. Spectral performance was assessed by acquiring spectra from fluorescence samples, and the results show a high accuracy of energy peak positions (< 1 keV), precise energy resolution (within a few keV) and decent peak-to-background ratios. Spectral absorption measurements of water and iodinated contrast agent, as well as spectral X-ray radiographs of a human hand phantom, decomposed into bone and soft tissue basis images, demonstrate the multi-energy performance of the detector.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Photon-counting MCP/Timepix detectors for soft X-ray imaging and spectroscopic applications
    Tremsin, Anton S.
    Vallerga, John V.
    Siegmund, Oswald H. W.
    Woods, Justin
    De Long, Lance E.
    Hastings, Jeffrey T.
    Koch, Roland J.
    Morley, Sophie A.
    Chuang, Yi-De
    Roy, Sujoy
    JOURNAL OF SYNCHROTRON RADIATION, 2021, 28 (28) : 1069 - 1080
  • [22] Performance of large-area CZT detectors for hard x-ray imaging
    Pelling, MR
    Heindl, WA
    Matteson, JL
    Rothschild, RE
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 2286 - 2290
  • [23] Spectral Signatures of X-ray Scatter Using Energy-Resolving Photon-Counting Detectors
    Lewis, Cale E.
    Das, Mini
    SENSORS, 2019, 19 (22)
  • [24] Characterization of spectrometric photon-counting X-ray detectors at different pitches
    Jurdit, M.
    Brambilla, A.
    Moulin, V.
    Ouvrier-Buffet, P.
    Radisson, P.
    Verger, L.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [25] Guidelines for the Selection of Scintillators for Indirect Photon-Counting X-ray Detectors
    van Blaaderen, J. Jasper
    van Aarle, Casper
    Leibold, David
    Dorenbos, Pieter
    Schaart, Dennis R.
    CHEMISTRY OF MATERIALS, 2025, 37 (05) : 1716 - 1740
  • [26] Simultaneous X-ray fluorescence and K-edge CT imaging with photon-counting detectors
    Li, Liang
    Li, Ruizhe
    Zhang, Siyuan
    Chen, Zhiqiang
    DEVELOPMENTS IN X-RAY TOMOGRAPHY X, 2016, 9967
  • [27] Single-shot x-ray phase imaging with grating interferometry and photon-counting detectors
    Wang, Zhili
    Gao, Kun
    Wang, Dajiang
    Wu, Zhao
    Chen, Heng
    Wang, Shenghao
    Wu, Ziyu
    OPTICS LETTERS, 2014, 39 (04) : 877 - 879
  • [28] Photon-counting CCD detector as a tool of x-ray imaging
    Liang, Y
    Ida, K
    Kado, S
    Minami, T
    Okamura, S
    Nomura, I
    Watanabe, KY
    Yamada, H
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01): : 717 - 720
  • [29] Practical energy response estimation of photon counting detectors for spectral X-ray imaging
    Kang, Dong-Goo
    Lee, Jongha
    Sung, Younghun
    Lee, SeongDeok
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [30] Photon-counting X-ray imaging at kilohertz frame rates
    Ponchut, Cyril
    Clement, Joel
    Rigal, Jean-Marie
    Papillon, Emmanuel
    Vallerga, John
    LaMarra, Daniel
    Mikulec, Bettina
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01): : 109 - 112