Validation of a point of care troponin assay in real life emergency department conditions

被引:8
作者
Loten, Conrad [1 ]
Attia, John [2 ]
Hullick, Carolyn
Marley, John [4 ]
McElduff, Patrick [3 ]
机构
[1] Univ Newcastle, Div Emergency Med, Hunter Reg Mail Ctr, John Hunter Hosp, Newcastle, NSW 2310, Australia
[2] Univ Newcastle, Ctr Clin Epidemiol & Biostat, Newcastle, NSW 2310, Australia
[3] Univ Newcastle, Sch Med & Publ Hlth, Hunter Med Res Inst, Newcastle, NSW 2310, Australia
[4] Univ Queensland, Fac Hlth Sci, Brisbane, Qld, Australia
关键词
emergency medicine; point of care system; troponin; MYOCARDIAL-INFARCTION; OF-CARE;
D O I
10.1111/j.1742-6723.2009.01198.x
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To validate the accuracy of a Point of care (POC) troponin device (Abbott i-Stat) in real life ED conditions. Methods: Design: A three-way comparison between troponin I results obtained by experienced POC operators, inexperienced ED staff and central laboratory criterion standard. Participants: Convenience sample of 332 patients presenting to the ED with possible coronary syndromes. Results: Spearman correlation coefficient for experienced versus laboratory was 0.83 (95% CI 0.78-0.87), occasional users versus laboratory was 0.76 (95% CI 0.71-0.81), and experienced versus occasional users on POC was 0.82 (95% CI 0.76-0.87). Using local troponin cut-off of 0.1 ng/mL, kappa coefficient was 0.94 for occasional users versus laboratory, 0.91 for experienced versus laboratory and 0.94 for experienced versus occasional users. Bland-Altman plots showed good agreement across the range of measured values. The sensitivity of i-Stat (vs laboratory as criterion standard) was 92.2% (95% CI 83.8%-97.0%) with the local cut-off but only 70.1% (95% CI 60.5%-78.6%) using the lowest cut-off associated with acceptable reproducibility (10% coefficient of variation). Conclusions: The i-Stat POC device produces similar results in the hands of experienced and occasional operators in ED. There is good agreement between the POC and laboratory at levels used to diagnose infarction by older, more specific criteria. When compared using new lower cut-offs, the i-Stat had poor sensitivity.
引用
收藏
页码:286 / 292
页数:7
相关论文
共 11 条
  • [1] Ischaemic heart disease - The universal definition of myocardial infarction: a consensus document
    Alpert, Joseph S.
    Thygesen, Kristian
    Jaffe, Allan
    White, Harvey D.
    [J]. HEART, 2008, 94 (10) : 1335 - 1341
  • [2] ACC/AHA 2007 guide lines for the management of patients with unstable Angina/Non-ST-Elevation myocardial infraction - Executive summary
    Anderson, Jeffrey L.
    Adams, Cynthia D.
    Antman, Elliott M.
    Bridges, Charles R.
    Califf, Robert M.
    Casey, Donald E., Jr.
    Chavey, William E., II
    Fesmire, Francis M.
    Hochman, Judith S.
    Levin, Thomas N.
    Lincoff, A. Michael
    Peterson, Eric D.
    Theroux, Pierre
    Wenger, Nanette Kass
    Wright, R. Scott
    Smith, Sidney C., Jr.
    Jacobs, Alice K.
    Adams, Cynthia D.
    Riegel, Barbara
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (07) : 652 - 726
  • [3] Point-of-care i-STAT cardiac troponin I for assessment of patients with symptoms suggestive of acute coronary syndrome
    Apple, FS
    Ler, R
    Chung, AY
    Berger, MJ
    Murakami, MM
    [J]. CLINICAL CHEMISTRY, 2006, 52 (02) : 322 - 325
  • [4] Analytical performance of the i-STAT cardiac troponin I assay
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    [J]. CLINICA CHIMICA ACTA, 2004, 345 (1-2) : 123 - 127
  • [5] European Society of Cardiology and American College of Cardiology guidelines for redefinition of myocardial infarction: How to use existing assays clinically and for clinical trials
    Apple, FS
    Wu, AHB
    Jaffe, AS
    [J]. AMERICAN HEART JOURNAL, 2002, 144 (06) : 981 - 986
  • [6] Comparison of emergency department patient classification by point-of-care and central laboratory methods for cardiac troponin I
    Bock, Jay L.
    Singer, Adam J.
    Thode, Henry C., Jr.
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (01) : 132 - 135
  • [7] Cardiac troponin T in chest pain unit patients without ischemic electrocardiographic changes: Angiographic correlates and long-term clinical outcomes
    deFilippi, CR
    Tocchi, P
    Parmar, RJ
    Rosanio, S
    Abreo, G
    Potter, MA
    Runge, MS
    Uretsky, BF
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (07) : 1827 - 1834
  • [8] Prognostic usefulness of marginal troponin T elevation
    Henrikson, CA
    Howell, EE
    Bush, DE
    Miles, JS
    Meininger, GR
    Friedlander, T
    Bushnell, AC
    Chandra-Strobos, N
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (03) : 275 - 279
  • [9] *INT FED CLIN CHEM, STAND MARK CARD DAM
  • [10] The antibody configurations of cardiac troponin I assays may determine their clinical performance
    James, S
    Flodin, M
    Johnston, N
    Lindahl, B
    Venge, P
    [J]. CLINICAL CHEMISTRY, 2006, 52 (05) : 832 - 837