A 32-Channel Combined RF and B0 Shim Array for 3T Brain Imaging

被引:112
作者
Stockmann, Jason P. [1 ]
Witzel, Thomas [1 ,2 ]
Keil, Boris [1 ,2 ]
Polimeni, Jonathan R. [1 ,2 ]
Mareyam, Azma [1 ]
LaPierre, Cristen [1 ]
Setsompop, Kawin [1 ,2 ]
Wald, Lawrence L. [1 ,2 ,3 ]
机构
[1] Massachusetts Gen Hosp, Athinoula A Martinos Ctr Biomed Imaging, Dept Radiol, Charlestown, MA 02129 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
[3] MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USA
关键词
multi-coil shimming; RF receive arrays; echo planar imaging; geometric distortion; brain MRI; PARALLEL RECEPTION; 7; TESLA; COIL; ACQUISITION; DESIGN; ORDER;
D O I
10.1002/mrm.25587
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: We add user-controllable direct currents (DC) to the individual elements of a 32-channel radio-frequency (RF) receive array to provide B-0 shimming ability while preserving the array's reception sensitivity and parallel imaging performance. Methods: Shim performance using constrained DC current (+/- 2.5A) is simulated for brain arrays ranging from 8 to 128 elements. A 32-channel 3-tesla brain array is realized using inductive chokes to bridge the tuning capacitors on each RF loop. The RF and B-0 shimming performance is assessed in bench and imaging measurements. Results: The addition of DC currents to the 32-channel RF array is achieved with minimal disruption of the RF performance and/or negative side effects such as conductor heating or mechanical torques. The shimming results agree well with simulations and show performance superior to third-order spherical harmonic (SH) shimming. Imaging tests show the ability to reduce the standard frontal lobe susceptibility-induced fields and improve echo planar imaging geometric distortion. The simulation of 64- and 128-channel brain arrays suggest that even further shimming improvement is possible (equivalent to up to 6th-order SH shim coils). Conclusion: Including user-controlled shim currents on the loops of a conventional highly parallel brain array coil is feasible with modest current levels and produces improved B-0 shimming performance over standard second-order SH shimming. (C) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:441 / 451
页数:11
相关论文
共 36 条
[1]  
Biber S, 2012, P 20 ANN M ISMRM MEL, P2746
[2]  
Breuer FA, 2005, P 13 ANN M ISMRM MIA, P2668
[3]   Dynamic shim updating (DSU) for multislice signal acquisition [J].
de Graaf, RA ;
Brown, PB ;
McIntyre, S ;
Rothman, DL ;
Nixon, TW .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (03) :409-416
[4]   Single Echo MRI [J].
Galiana, Gigi ;
Constable, R. Todd .
PLOS ONE, 2014, 9 (01)
[5]   Simultaneously Driven Linear and Nonlinear Spatial Encoding Fields in MRI [J].
Gallichan, Daniel ;
Cocosco, Chris. A. ;
Dewdney, Andrew ;
Schultz, Gerrit ;
Welz, Anna ;
Hennig, Juergen ;
Zaitsev, Maxim .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (03) :702-714
[6]  
Grissom W.A., 2011, Proc. 19th Ann. Meet. Soc. Mag. Reson. Med, P2898
[7]   Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) [J].
Griswold, MA ;
Jakob, PM ;
Heidemann, RM ;
Nittka, M ;
Jellus, V ;
Wang, JM ;
Kiefer, B ;
Haase, A .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1202-1210
[8]   Integrated parallel reception, excitation, and shimming (iPRES) [J].
Han, Hui ;
Song, Allen W. ;
Trong-Kha Truong .
MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) :241-247
[9]  
Hui H, 2013, P 21 ANN M ISMRM SAL, P664
[10]   Fast, automated, N-dimensional phase-unwrapping algorithm [J].
Jenkinson, M .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :193-197