Effect of Bayesian-penalized likelihood reconstruction on [13N]-NH3 rest perfusion quantification

被引:12
作者
O' Doherty, Jim [1 ]
McGowan, Daniel R. [2 ,3 ]
Abreu, Carla [1 ]
Barrington, Sally [1 ]
机构
[1] Kings Coll London, Kings Hlth Partners, St Thomas Hosp, PET Imaging Ctr,Div Imaging Sci & Biomed Engn, 1st Floor,Lambeth Wing, London SE1 7EH, England
[2] Univ Oxford, Dept Oncol, Oxford, England
[3] Oxford Univ Hosp NHS Fdn Trust, Churchill Hosp, Radiat Phys & Protect, Oxford, England
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
N-13; ammonia; PET; physics of imaging; MYOCARDIAL BLOOD-FLOW; CARDIAC SARCOIDOSIS; 18F-FDG PET; N-13-AMMONIA; SPECT/CT; PHANTOM; HEART;
D O I
10.1007/s12350-016-0554-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. Myocardial blood flow (MBF) imaging is used in patients with suspected cardiac sarcoidosis, and also in stress/ rest studies. The accuracy of MBF is dependent on imaging parameters such as new reconstruction methodologies. In this work, we aim to assess the impact of a novel PET reconstruction algorithm (Bayesian-penalized likelihood-BPL) on the values determined from the calculation of [13N]-NH3 MBF values. Methods. Data from 21 patients undergoing rest MBF evaluation [13N]-NH3 as part of sarcoidosis imaging were retrospectively analyzed. Each scan was reconstructed with a range of BPL coefficients (1-500), and standard clinical FBP and OSEM reconstructions. MBF values were calculated via an automated software routine for all datasets. Results. Reconstruction of [13N]-NH3 dynamic data using the BPL, OSEM, or FBP reconstruction showed no quantitative differences for the calculation of territorial or global MBF (P = .97). Image noise was lower using OSEM or BPL reconstructions than FBP and noise from BPL reached levels seen in OSEM images between B = 300 and B = 400. Intrasubject differences between all reconstructions over all patients in respect of all cardiac territories showed a maximum coefficient of variation of 9.74%. Conclusion. Quantitation of MBF via kinetic modeling of cardiac rest MBF by [13N]-NH3 is minimally affected by the use of a BPL reconstruction technique, with BPL images presenting with less noise.
引用
收藏
页码:282 / 290
页数:9
相关论文
共 35 条
[1]  
[Anonymous], 2014, J NUCL MED S1
[2]  
Bouchareb Y, 2005, IEEE NUCL SCI CONF R, P2120
[3]   Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart - A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association [J].
Cerqueira, MD ;
Weissman, NJ ;
Dilsizian, V ;
Jacobs, AK ;
Kaul, S ;
Laskey, WK ;
Pennell, DJ ;
Rumberger, JA ;
Ryan, T ;
Verani, MS .
CIRCULATION, 2002, 105 (04) :539-542
[4]   Perfusion phantom: An efficient and reproducible method to simulate myocardial first-pass perfusion measurements with cardiovascular magnetic resonance [J].
Chiribiri, Amedeo ;
Schuster, Andreas ;
Ishida, Masaki ;
Hautvast, Gilion ;
Zarinabad, Niloufar ;
Morton, Geraint ;
Otton, James ;
Plein, Sven ;
Breeuwer, Marcel ;
Batchelor, Philip ;
Schaeffter, Tobias ;
Nagel, Eike .
MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (03) :698-707
[5]   SNMMI/ASNC/SCCT Guideline for Cardiac SPECT/CT and PET/CT 1.0 [J].
Dorbala, Sharmila ;
Di Carli, Marcelo F. ;
Delbeke, Dominique ;
Abbara, Suhny ;
DePuey, E. Gordon ;
Dilsizian, Vasken ;
Forrester, Joey ;
Janowitz, Warren ;
Kaufmann, Philipp A. ;
Mahmarian, John ;
Moore, Stephen C. ;
Stabin, Michael G. ;
Shreve, Paul .
JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (08) :1485-1507
[6]   Motion Correction in Dual Gated Cardiac PET Using Mass-Preserving Image Registration [J].
Gigengack, Fabian ;
Ruthotto, Lars ;
Burger, Martin ;
Wolters, Carsten H. ;
Jiang, Xiaoyi ;
Schaefers, Klaus P. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (03) :698-712
[7]   NONINVASIVE QUANTIFICATION OF REGIONAL BLOOD-FLOW IN THE HUMAN HEART USING N-13 AMMONIA AND DYNAMIC POSITRON EMISSION TOMOGRAPHIC IMAGING [J].
HUTCHINS, GD ;
SCHWAIGER, M ;
ROSENSPIRE, KC ;
KRIVOKAPICH, J ;
SCHELBERT, H ;
KUHL, DE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (05) :1032-1042
[8]   Absolute Myocardial Blood Flow Emerging Role in Coronary Pathophysiology and Clinical Disease [J].
Klocke, Francis J. ;
Lee, Daniel C. .
JACC-CARDIOVASCULAR IMAGING, 2011, 4 (09) :999-1001
[9]   Evaluation of respiratory and cardiac motion correction schemes in dual gated PET/CT cardiac imaging [J].
Lamare, F. ;
Le Maitre, A. ;
Dawood, M. ;
Schaefers, K. P. ;
Fernandez, P. ;
Rimoldi, O. E. ;
Visvikis, D. .
MEDICAL PHYSICS, 2014, 41 (07)
[10]   Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT [J].
Lee, Young Sub ;
Kim, Jin Su ;
Kim, Kyeong Min ;
Kang, Joo Hyun ;
Lim, Sang Moo ;
Kim, Hee-Joung .
ANNALS OF NUCLEAR MEDICINE, 2014, 28 (04) :340-348