Analysis of nanoliter samples of electrolytes using a flow-through microfluorometer

被引:20
作者
Zhelyaskov, VR [1 ]
Liu, SY [1 ]
Broderick, MP [1 ]
机构
[1] World Precis Instruments Inc, Int Trade Ctr, Sarasota, FL 34240 USA
关键词
transport; renal tubules; nanoliter sample; flow-through microfluorometer; intracellular dye; electrolyte; calcium; sodium; chloride; urea; ammonia; lactate; glucose; carbon dioxide;
D O I
10.1046/j.1523-1755.2000.00022.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Several techniques have been developed to study the transport properties of nanoliter samples of renal tubule segments, such as continuous now colorimetry and continuous fluorometry. We have extended the capability of the NANOFLO(TM), a how-through microfluorometer, designed for measurement of carbon dioxide, urea, ammonia. glucose, lactate, etc., to analyze sodium, calcium and chloride ions, using three commercially available fluorescent indicators for intracellular and extracellular measurements. The selection of fluorescent indicator for each electrolyte was dependent on the optimal match of the dissociation constant and the analyte concentration range of interest. Using Fluo-3 dye we achieved a detection limit for Ca2+ of 0.1 pmol and selectivity over Mg2+ of between 7:1 to 10:1. Using sodium green dye we achieved detection limit for Na+ of 12 pmol and a selectivity over K+ of 40:1. The detection limit for Cl- using lucigenin dye was 10 pmol. This technique can be readily adapted for the measurement of other physiologically important ultralow volume.
引用
收藏
页码:1764 / 1769
页数:6
相关论文
共 18 条
[1]  
BAGNASCO S, 1985, AM J PHYSIOL, V258, pF522
[2]   LONG-WAVELENGTH CHLORIDE-SENSITIVE FLUORESCENT INDICATORS [J].
BIWERSI, J ;
TULK, B ;
VERKMAN, AS .
ANALYTICAL BIOCHEMISTRY, 1994, 219 (01) :139-143
[3]   Fluorescent determination of chloride in nanoliter samples [J].
García, NH ;
Plato, CF ;
Garvin, JL .
KIDNEY INTERNATIONAL, 1999, 55 (01) :321-325
[4]   AMMONIUM REPLACES POTASSIUM IN SUPPORTING SODIUM-TRANSPORT BY THE NA-K-ATPASE OF RENAL PROXIMAL STRAIGHT TUBULES [J].
GARVIN, JL ;
BURG, MB ;
KNEPPER, MA .
AMERICAN JOURNAL OF PHYSIOLOGY, 1985, 249 (05) :F784-F788
[5]   PICOMOLE QUANTITATION OF AMMONIA BY FLOW-THROUGH FLUOROMETRY [J].
GOOD, DW ;
VUREK, GG .
ANALYTICAL BIOCHEMISTRY, 1983, 130 (01) :199-202
[6]  
HAUGLAND RP, 1996, HDB FLUORESECENT PRO
[7]  
Liu S., 1995, Patent No. [US 5,444,807, 5444807]
[8]  
Liu S.Y., 1995, US Patent No, Patent No. [5,416.879, 5416879]
[9]   Colonic H+-K+-ATPase is induced and mediates increased HCO3 - reabsorption in inner medullary collecting duct in potassium depletion [J].
Nakamura, S ;
Amlal, H ;
Galla, JH ;
Soleimani, M .
KIDNEY INTERNATIONAL, 1998, 54 (04) :1233-1239
[10]   HCO3- reabsorption in renal collecting duct of NHE-3- deficient mouse:: a compensatory response [J].
Nakamura, S ;
Amlal, H ;
Schultheis, PJ ;
Galla, JH ;
Shull, GE ;
Soleimani, M .
AMERICAN JOURNAL OF PHYSIOLOGY-RENAL PHYSIOLOGY, 1999, 276 (06) :F914-F921