Is the knowledge of contact force beneficial in pulmonary vein antrum isolation?

被引:9
作者
Borregaard, Rune [1 ]
Jensen, Henrik Kjaerulf [1 ]
Tofig, Bawer Jalal [1 ]
Thrysoe, Samuel Alberg [2 ]
Gerdes, Christian [1 ]
Nielsen, Jens Cosedis [1 ]
Lukac, Peter [1 ]
机构
[1] Aarhus Univ Hosp, Dept Cardiol, Blvd 99, DK-8200 Aarhus N, Denmark
[2] Aarhus Univ, Sch Engn, Aarhus, Denmark
关键词
Atrial fibrillation; pulmonary vein antrum isolation; contact force; implantable cardiac monitor; randomized; RADIOFREQUENCY CATHETER ABLATION; ATRIAL-FIBRILLATION; SENSING CATHETER; RECONNECTION; IMPEDANCE; MODEL;
D O I
10.1080/14017431.2017.1285043
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To investigate the effect of the operator knowing the real-time contact force (CF) on the efficacy of pulmonary vein antrum isolation (PVAI). Methods: Fifty patients with paroxysmal atrial fibrillation (AF) or short lasting persistent AF were randomized to CF guided PVAI (n = 25) or conventional PVAI (n = 25). In the CF guided group, CF between 10 and 40 g was aimed at. Efficacy of PVAI was measured as reduction in AF burden (AFB) and time to AF recurrence detected by implantable cardiac monitor (ICM), inserted three months before PVAI. Blanking period was three months and follow-up 12 months. Results: All pulmonary veins were isolated in the CF guided group and all but one in the conventional group. Mean CF was 25 g in the CF guided group and 24 g in the conventional group (p = 0.75). Compared to pre-ablation, median [IQR] relative reduction in AFB 3-12 months after ablation was 100 [99- 100]% in the CF guided group (p< 0.001) and 99.4 [25- 100]% in the conventional group (p< 0.001), not different between groups (p = 0.09). Nine patients (36%) had AF recurrence in the CF guided group and 13 (52%) in the conventional group (p = 0.21, log-rank test). CF differed between operators. When adjusted for operator by regression analysis, patients without recurrent AF had lower proportion of ablation time with CF < 10 g than recurrent patients (p = 0.034). No complications occurred. Conclusions: Operator knowledge of real-time CF had no significant effect on AFB reduction or time to AF recurrence. Larger trials should be done to study benefit of real-time CF.
引用
收藏
页码:129 / 137
页数:9
相关论文
共 20 条
[1]   Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation [J].
Cappato, R ;
Negroni, S ;
Pecora, D ;
Bentivegna, S ;
Lupo, PP ;
Carolei, A ;
Esposito, C ;
Furlanello, F ;
De Ambroggi, L .
CIRCULATION, 2003, 108 (13) :1599-1604
[2]   General anesthesia reduces the prevalence of pulmonary vein reconnection during repeat ablation when compared with conscious sedation: Results from a randomized study [J].
Di Biase, Luigi ;
Conti, Sergio ;
Mohanty, Prasant ;
Bai, Rong ;
Sanchez, Javier ;
Walton, David ;
John, Annie ;
Santangeli, Pasquale ;
Elayi, Claude S. ;
Beheiry, Salwa ;
Gallinghouse, G. Joseph ;
Mohanty, Sanghamitra ;
Horton, Rodney ;
Bailey, Shane ;
Burkhardt, J. David ;
Natale, Andrea .
HEART RHYTHM, 2011, 8 (03) :368-372
[3]   Long-term Outcomes of Catheter Ablation of Atrial Fibrillation: A Systematic Review and Meta-analysis [J].
Ganesan, Anand N. ;
Shipp, Nicholas J. ;
Brooks, Anthony G. ;
Kuklik, Pawel ;
Lau, Dennis H. ;
Lim, Han S. ;
Sullivan, Thomas ;
Roberts-Thomson, Kurt C. ;
Sanders, Prashanthan .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2013, 2 (02) :e004549
[4]  
HAINES D E, 1991, Journal of Cardiovascular Electrophysiology, V2, P509, DOI 10.1111/j.1540-8167.1991.tb01353.x
[5]   Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: A prospective case control study [J].
Haldar, Shouvik ;
Jarman, Julian W. E. ;
Panikker, Sandeep ;
Jones, David G. ;
Salukhe, Tushar ;
Gupta, Dhiraj ;
Wynn, Gareth ;
Hussain, Wajid ;
Markides, Vias ;
Wong, Tom .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (02) :1160-1166
[6]   Comparison of lesion formation between contact force-guided and non-guided circumferential pulmonary vein isolation: A prospective, randomized study [J].
Kimura, Masaomi ;
Sasaki, Shingo ;
Owada, Shingen ;
Horiuchi, Daisuke ;
Sasaki, Kenichi ;
Itoh, Taihei ;
Ishida, Yuji ;
Kinjo, Takahiko ;
Tomita, Hirofumi ;
Okumura, Ken .
HEART RHYTHM, 2014, 11 (06) :984-991
[7]   Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans [J].
Kumar, Saurabh ;
Haqqani, H. M. ;
Chan, Martin ;
Lee, Justin ;
Yudi, Matias ;
Wong, Michael C. G. ;
Morton, Joseph B. ;
Ling, Liang-Han ;
Robinson, Timothy ;
Heck, Patrick M. ;
Kelland, Nicholas F. ;
Halloran, Karen ;
Spence, Steven J. ;
Kistler, Peter M. ;
Kalman, Jonathan M. .
HEART RHYTHM, 2013, 10 (07) :962-969
[8]   Real-Time Contact Force Sensing for Pulmonary Vein Isolation in the Setting of Paroxysmal Atrial Fibrillation: Procedural and 1-Year Results [J].
Marijon, Eloi ;
Fazaa, Samia ;
Narayanan, Kumar ;
Guy-Moyat, Benoit ;
Bouzeman, Abdeslam ;
Providencia, Rui ;
Treguer, Frederic ;
Combes, Nicolas ;
Bortone, Agustin ;
Boveda, Serge ;
Combes, Stephane ;
Albenque, Jean-paul .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (02) :130-137
[9]   Clinical Impact of an Open-Irrigated Radiofrequency Catheter with Direct Force Measurement on Atrial Fibrillation Ablation [J].
Martinek, Martin ;
Lemes, Christine ;
Sigmund, Elisabeth ;
Derndorfer, Michael ;
Aichinger, Josef ;
Winter, Siegmund ;
Nesser, Hans-Joachim ;
Puererfellner, Helmut .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (11) :1312-1318
[10]   Locations of High Contact Force During Left Atrial Mapping in Atrial Fibrillation Patients Electrogram Amplitude and Impedance Are Poor Predictors of Electrode-Tissue Contact Force for Ablation of Atrial Fibrillation [J].
Nakagawa, Hiroshi ;
Kautzner, Josef ;
Natale, Andrea ;
Peichl, Petr ;
Cihak, Robert ;
Wichterle, Dan ;
Ikeda, Atsushi ;
Santangeli, Pasquale ;
Di Biase, Luigi ;
Jackman, Warren M. .