Repeatability of high-frequency distortion-product otoacoustic emissions in normal-hearing adults

被引:42
作者
Dreisbach, Laura Elizabeth
Long, Kelly Melissa
Lees, Shannon Elizabeth
机构
[1] San Diego State Univ, Sch Speech Language & Hearing Sci, San Diego, CA 92182 USA
[2] Calif Ear Inst, Palo Alto, CA USA
关键词
D O I
10.1097/01.aud.0000233892.37803.1a
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: Distortion-product otoacoustic emissions (DPOAEs) are repeatable over time at lower frequencies (<= 8 kHz). The purpose of this study was to examine the repeatability of DPOAEs measured with high-frequency stimuli in normal-hearing adult subjects. It was hypothesized that high-frequency DPOAEs would be repeatable over a minimum of 4 wk. Design: DPOAEs were measured in 25 subjects (14 female and 11 male) with normal behavioral thresholds, normal middle ear function, and present acoustic reflexes at 1 kHz evoked by contralateral. stimulation. Behavioral thresholds were measured through 16 kHz, using Bekesy tracking. Each subject attended four trials, in which a complete set of data was collected. The four trials were separated by at least one, but no more than 2 wk. At each trial, two different DPOAE paradigms were completed. Frequency and ratio sweeps were used to measure DPOAE levels and calculate DPOAE group delay, respectively. Frequency sweeps were measured with varied stimulus level conditions (L-1/L-2 = 60/45, 60/50, 70/55, 70/60 dB SPL) and an f(2)/f(1), of 1.2 at discrete f(2), frequencies between 2 and 16 kHz. Ratio sweeps were obtained at f(2), frequencies of 2,4,6,8,10,12,14, and 16 kHz, with L, = 60 and L-2 = 45 dB, SPL, and the ratio (f(2)/f(1)) was varied from 1.05 to 1.3. Repeated-measures analysis of variance was performed. Results: Greater variability was found at the higher frequencies (> 8 kHz) for DPOAE level measurements. The average DPOAE level differences-between-trials for the higher and lower frequencies for the four different stimulus level conditions was 5.15 (SD = 4.40 dB) and 2.80 (SD = 2.70 dB) dB, respectively. Individual subject analysis revealed that high-frequency DPOAE levels varied no more than 10 dB for 87.5 and 83.1% of young adult subjects for the 70/55 and 60/50 dB SPL stimulus level conditions, respectively. For low frequencies, repeated DPOAE level variations were within +/- 10 dB; for 98.4 and 96% of young adult subjects for the 70/55 and 60/50 dB SPL stimulus level conditions, respectively. For DPOAE group delay, greater variability was noted at lower frequencies (<= 8 kHz) and at 10 kHz. The average group-delay differences-betweentrials for the higher (10 to 16 kHz) and lower (<= 8 kHz) frequencies was 0.22 (SD = 0.20 msec) and 0.28 (SD = 0.24 msec) msec, respectively. Individual subject analysis revealed that DPOAE group-delay values varied no more than 0.70 msec for the majority of young adult subjects (96.5 and 93.5% at high and low frequencies, respectively). Conclusions: Even though the DPOAE level data obtained at frequencies greater than 8 kHz were more variable than at low frequencies, the higher frequencies were found to be repeatable for both paradigms tested. These results encourage the exploration of high-frequency DPOAE measures to be used as an objective test for monitoring ototoxicity in humans. Testing subjects receiving ototoxic therapies is a necessary step in determining if monitoring high-frequency DPOAEs will successfully predict ototoxic effects.
引用
收藏
页码:466 / 479
页数:14
相关论文
共 60 条
[1]  
*AM SPEECH LANG HE, 1989, ASHA, V31, P71
[2]  
American Speech-Language-Hearing Association, 1994, ASHA, V36, P11
[3]   High-frequency hearing influences lower-frequency distortion-product otoacoustic emissions [J].
Arnold, DJ ;
Lonsbury-Martin, BL ;
Martin, GK .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1999, 125 (02) :215-222
[4]  
ATHERLEY GR, 1963, BRIT J IND MED, V20, P231
[5]   Effects of sample size on the reliability of noise floor and DPOAE [J].
Beattie, RC ;
Bleech, J .
BRITISH JOURNAL OF AUDIOLOGY, 2000, 34 (05) :305-309
[6]   Immediate and short-term reliability of distortion-product otoacoustic emissions [J].
Beattie, RC ;
Kenworthy, OT ;
Luna, CA .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2003, 42 (06) :348-354
[7]  
Biro K, 1997, ONCOLOGY, V54, P387
[8]   ACOUSTIC DISTORTION PRODUCTS CAN BE USED TO MONITOR THE EFFECTS OF CHRONIC GENTAMICIN TREATMENT [J].
BROWN, AM ;
MCDOWELL, B ;
FORGE, A .
HEARING RESEARCH, 1989, 42 (2-3) :143-156
[10]   Individual differences and the reliability of 2F1-F2 distortion product otoacoustic emissions: Effects of time-of-day, stimulus variables, and gender [J].
Cacace, AT ;
McClelland, WA ;
Weiner, J ;
McFarland, DJ .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1996, 39 (06) :1138-1148