Hydrodynamic properties of hydrocephalus shunts: United Kingdom shunt evaluation laboratory

被引:55
作者
Czosnyka, M [1 ]
Czosnyka, Z [1 ]
Whitehouse, H [1 ]
Pickard, JD [1 ]
机构
[1] ADDENBROOKES HOSP, MRC, CAMBRIDGE CTR BRAIN REPAIR, CAMBRIDGE CB2 2QQ, ENGLAND
关键词
hydrocephalus shunts; cerebrospinal fluid; hydrodynamics; laboratory test;
D O I
10.1136/jnnp.62.1.43
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background-Although about 80% of properly diagnosed patients with hydrocephalus improve after implantation of any model of shunt, the remaining 20% may develop further complications because of inadequate shunt performance. Therefore, hydrocephalus shunts require careful independent laboratory evaluation. Method-Computer supported shunt testing, based on the new International Standard Organisation directives, characterises various aspects of pressure-flow performance of shunts such as variability with time, susceptibility to reflux, siphoning, temperature related behaviour, external pressure, the influence of a strong magnetic field (for example, MRI), presence of pulsation in differential pressure, particles in drained fluid, etc. Results-Seven different models of valves, representing most common constructions, have been tested so far. Most contemporary valves have a hydrodynamic resistance which is too low. This may result in overdrainage both related to posture and during nocturnal cerebral vasogenic waves. A long distal catheter increases the resistance of these valves by 100%-200%. Most shunts are very sensitive to the presence of air bubbles and small particles in drained fluid. Few shunt models offer reasonable resistance to negative outlet pressure, preventing complications related to overdrainage. Valves with an antisiphon device may be blocked by raised subcutaneous pressure. All programmable valves are susceptible to overdrainage in an upright position. Conclusion-The behaviour of a valve during such testing is of immediate relevance to the surgeon and may not be adequately described in the manufacturer's product information.
引用
收藏
页码:43 / 50
页数:8
相关论文
共 23 条
  • [1] CHRONIC HEADACHE ASSOCIATED WITH A FUNCTIONING SHUNT - USEFULNESS OF PRESSURE MONITORING
    ABBOTT, R
    EPSTEIN, FJ
    WISOFF, JH
    [J]. NEUROSURGERY, 1991, 28 (01) : 72 - 77
  • [2] OVERDRAINAGE AND SHUNT TECHNOLOGY - A CRITICAL COMPARISON OF PROGRAMMABLE, HYDROSTATIC AND VARIABLE RESISTANCE VALVES AND FLOW-REDUCING DEVICES
    ASCHOFF, A
    KREMER, P
    BENESCH, C
    FRUH, K
    KLANK, A
    KUNZE, S
    [J]. CHILDS NERVOUS SYSTEM, 1995, 11 (04) : 193 - 202
  • [3] ASCHOFF A, 1995, 10 EUR C NEUR BERL
  • [4] BAYSTON R, 1989, HYDROCEPHALUS SHUNT
  • [5] COMPUTER-SUPPORTED MULTIMODAL BED-SIDE MONITORING FOR NEURO INTENSIVE-CARE
    CZOSNYKA, M
    WHITEHOUSE, H
    SMIELEWSKI, P
    KIRKPATRICK, P
    GUAZZO, EP
    PICKARD, JD
    [J]. INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1994, 11 (04): : 223 - 232
  • [6] Testing of cerebrospinal compensatory reserve in shunted and non-shunted patients: A guide to interpretation based on an observational study
    Czosnyka, M
    Whitehouse, H
    Smielewski, P
    Simac, S
    Pickard, JD
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1996, 60 (05) : 549 - 558
  • [7] COMPARISON BETWEEN CLASSIC-DIFFERENTIAL AND AUTOMATIC SHUNT FUNCTIONING ON THE BASIS OF INFUSION TESTS
    CZOSNYKA, M
    MAKSYMOWICZ, W
    BATORSKI, L
    KOSZEWSKI, W
    CZOSNYKA, Z
    [J]. ACTA NEUROCHIRURGICA, 1990, 106 (1-2) : 1 - 8
  • [8] FUNCTIONAL OBSTRUCTION OF AN ANTISIPHON DEVICE BY RAISED TISSUE CAPSULE PRESSURE
    DRAKE, JM
    DASILVA, MC
    RUTKA, JT
    [J]. NEUROSURGERY, 1993, 32 (01) : 137 - 139
  • [9] DRAKE JM, 1995, SHUNT BOOK, P132
  • [10] OVER-DRAINAGE PHENOMENA IN SHUNT TREATED HYDROCEPHALUS
    FAULHAUER, K
    SCHMITZ, P
    [J]. ACTA NEUROCHIRURGICA, 1978, 45 (1-2) : 89 - 101