A Robust Random Forest-Based Approach for Heart Rate Monitoring Using Photoplethysmography Signal Contaminated by Intense Motion Artifacts

被引:31
作者
Ye, Yalan [1 ]
He, Wenwen [1 ]
Cheng, Yunfei [1 ]
Huang, Wenxia [2 ]
Zhang, Zhilin [3 ]
机构
[1] Univ Elect Sci & Technol China, Sch Comp Sci & Engn, Chengdu 611731, Peoples R China
[2] Sichuan Univ, West China Hosp, Chengdu 610041, Peoples R China
[3] Samsung Res Amer Dallas, Richardson, TX 75082 USA
基金
中国国家自然科学基金;
关键词
heart rate (HR); photoplethysmography (PPG); motion artifacts (MA); random forest; REDUCTION;
D O I
10.3390/s17020385
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The estimation of heart rate (HR) based on wearable devices is of interest in fitness. Photoplethysmography (PPG) is a promising approach to estimate HR due to low cost; however, it is easily corrupted by motion artifacts (MA). In this work, a robust approach based on random forest is proposed for accurately estimating HR from the photoplethysmography signal contaminated by intense motion artifacts, consisting of two stages. Stage 1 proposes a hybrid method to effectively remove MA with a low computation complexity, where two MA removal algorithms are combined by an accurate binary decision algorithm whose aim is to decide whether or not to adopt the second MA removal algorithm. Stage 2 proposes a random forest-based spectral peak-tracking algorithm, whose aim is to locate the spectral peak corresponding to HR, formulating the problem of spectral peak tracking into a pattern classification problem. Experiments on the PPG datasets including 22 subjects used in the 2015 IEEE Signal Processing Cup showed that the proposed approach achieved the average absolute error of 1.65 beats per minute (BPM) on the 22 PPG datasets. Compared to state-of-the-art approaches, the proposed approach has better accuracy and robustness to intense motion artifacts, indicating its potential use in wearable sensors for health monitoring and fitness tracking.
引用
收藏
页数:16
相关论文
共 32 条
[1]  
Asada HH, 2004, P ANN INT IEEE EMBS, V26, P2157
[2]   Random forests [J].
Breiman, L .
MACHINE LEARNING, 2001, 45 (01) :5-32
[3]   Wavelets: The mathematical background [J].
Cohen, A ;
Kovacevic, J .
PROCEEDINGS OF THE IEEE, 1996, 84 (04) :514-522
[4]  
Fukushima H, 2012, IEEE ENG MED BIO, P2901, DOI 10.1109/EMBC.2012.6346570
[5]   Artifacts in wearable photoplethysmographs during daily life motions and their reduction with least mean square based active noise cancellation method [J].
Han, Hyonyoung ;
Kim, Jung .
COMPUTERS IN BIOLOGY AND MEDICINE, 2012, 42 (04) :387-393
[6]  
Hassani H., 2007, JDS, V5, P239, DOI DOI 10.6339/JDS.2007.05(2).396
[7]   Linear and quadratic time-frequency signal representations [J].
Hlawatsch, F. ;
Boudreaux-Bartels, G. F. .
IEEE SIGNAL PROCESSING MAGAZINE, 1992, 9 (02) :21-67
[8]   SKIN PHOTOPLETHYSMOGRAPHY - A REVIEW [J].
KAMAL, AAR ;
HARNESS, JB ;
IRVING, G ;
MEARNS, AJ .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1989, 28 (04) :257-269
[9]   Detection of Respiratory Arousals Using Photoplethysmography (PPG) Signal in Sleep Apnea Patients [J].
Karmakar, Chandan ;
Khandoker, Ahsan ;
Penzel, Thomas ;
Schoebel, Christoph ;
Palaniswami, Marimuthu .
IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2014, 18 (03) :1065-1073
[10]   A Robust Heart Rate Monitoring Scheme Using Photoplethysmographic Signals Corrupted by Intense Motion Artifacts [J].
Khan, Emroz ;
Al Hossain, Forsad ;
Uddin, Shiekh Zia ;
Alam, S. Kaisar ;
Hasan, Md. Kamrul .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (03) :550-562