A methodology to develop computational phantoms with adjustable posture for WBC calibration

被引:15
作者
Fonseca, T. C. Ferreira [1 ]
Bogaerts, R. [2 ]
Hunt, John [3 ]
Vanhavere, F. [1 ]
机构
[1] CEN SCK, Belgian Nucl Res Ctr, B-2400 Mol, Belgium
[2] Univ Hosp Gasthuisberg, Dept Radiat Oncol, B-3000 Louvain, Belgium
[3] Brazilian Inst Radioprotect & Dosimetry IRD, Rio De Janeiro, Brazil
关键词
MaMP and FeMP; mesh geometry surface; computational phantoms; whole body counter (WBC); Monte Carlo; calibration; IN-VIVO MEASUREMENTS; VOXEL PHANTOMS; EFFICIENCY; MODELS; FEMALE; AM-241; LIBRARY;
D O I
10.1088/0031-9155/59/22/6811
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Whole Body Counter (WBC) is a facility to routinely assess the internal contamination of exposed workers, especially in the case of radiation release accidents. The calibration of the counting device is usually done by using anthropomorphic physical phantoms representing the human body. Due to such a challenge of constructing representative physical phantoms a virtual calibration has been introduced. The use of computational phantoms and the Monte Carlo method to simulate radiation transport have been demonstrated to be a worthy alternative. In this study we introduce a methodology developed for the creation of realistic computational voxel phantoms with adjustable posture for WBC calibration. The methodology makes use of different software packages to enable the creation and modification of computational voxel phantoms. This allows voxel phantoms to be developed on demand for the calibration of different WBC configurations. This in turn helps to study the major source of uncertainty associated with the in vivo measurement routine which is the difference between the calibration phantoms and the real persons being counted. The use of realistic computational phantoms also helps the optimization of the counting measurement. Open source codes such as MakeHuman and Blender software packages have been used for the creation and modelling of 3D humanoid characters based on polygonal mesh surfaces. Also, a home-made software was developed whose goal is to convert the binary 3D voxel grid into a MCNPX input file. This paper summarizes the development of a library of phantoms of the human body that uses two basic phantoms called MaMP and FeMP (Male and Female Mesh Phantoms) to create a set of male and female phantoms that vary both in height and in weight. Two sets of MaMP and FeMP phantoms were developed and used for efficiency calibration of two different WBC set-ups: the Doel NPP WBC laboratory and AGM laboratory of SCK-CEN in Mol, Belgium.
引用
收藏
页码:6811 / 6825
页数:15
相关论文
共 26 条
[1]  
Bastioni M, 2011, MAKEHUMAN APPL OPEN
[2]  
Blender, 2011, BLEND FREE OP SOURC
[3]   Hybrid computational phantoms for medical dose reconstruction [J].
Bolch, Wesley ;
Lee, Choonsik ;
Wayson, Michael ;
Johnson, Perry .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2010, 49 (02) :155-168
[4]   Construction of an extended library of adult male 3D models: rationale and results [J].
Broggio, D. ;
Beurrier, J. ;
Bremaud, M. ;
Desbree, A. ;
Farah, J. ;
Huet, C. ;
Franck, D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (23) :7659-7692
[5]  
Carlan L D, 2006, RADIAT PROT DOSIM, V125, P477
[6]   FASH and MASH: female and male adult human phantoms based on polygon mesh surfaces: I. Development of the anatomy [J].
Cassola, V. F. ;
de Melo Lima, V. J. ;
Kramer, R. ;
Khoury, H. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :133-162
[7]  
Cristy M., 1980, MATH PHANTOMS REPRES
[8]   CREATION AND USE OF ADJUSTABLE 3D PHANTOMS: APPLICATION FOR THE LUNG MONITORING OF FEMALE WORKERS [J].
Farah, Jad ;
Broggio, David ;
Franck, Didier .
HEALTH PHYSICS, 2010, 99 (05) :649-661
[9]   Adult female voxel models of different stature and photon conversion coefficients for radiation protection [J].
Fill, UA ;
Zankl, M ;
Petoussi-Henss, N ;
Siebert, M ;
Regulla, D .
HEALTH PHYSICS, 2004, 86 (03) :253-272
[10]   MaMP and FeMP: computational mesh phantoms applied for studying the variation of WBC efficiency using a NaI(TI) detector [J].
Fonseca, T. C. Ferreira ;
Bogaerts, R. ;
Lebacq, A. L. ;
Ribeiro, R. M. ;
Vanhavere, F. .
JOURNAL OF RADIOLOGICAL PROTECTION, 2014, 34 (03) :529-543