Purpose The aim of this study was to determine whether errors do occur in navigated total knee arthroplasty (TKAs) and to study whether errors in bone resection or implantation contribute to these errors. Methods A series of 20 TKAs was studied using computer navigation. The coronal and sagittal alignments of the femoral and tibial cutting guides, the coronal and sagittal alignments of the final tibial implant and the coronal alignment of the final femoral implant were compared with that of the respective bone resections. To determine the post-implantation mechanical alignment of the limb, the coronal alignment of the femoral and tibial implants was combined. Results The median deviation between the femoral cutting guide and bone resection was 0 degrees (range -0.5 degrees to +0.5 degrees) in the coronal plane and 1.0 degrees (range -2.0 degrees to +1.0 degrees) in the sagittal plane. The median deviation between the tibial cutting guide and bone resection was 0.5 degrees (range -1.0 degrees to +1.5 degrees) in the coronal plane and 1.0 degrees (range -1.0 degrees to +3.5 degrees) in the sagittal plane. The median deviation between the femoral bone resection and the final implant was 0.25 degrees (range -2.0 degrees to 3.0 degrees) in the coronal plane. The median deviation between the tibial bone resection and the final implant was 0.75 degrees (range -3.0 degrees to +1.5 degrees) in the coronal plane and 1.75 degrees (range -4.0 degrees to +2.0 degrees) in the sagittal plane. The median post-implantation mechanical alignment of the limb was 0.25 degrees (range -3.0 degrees to +2.0 degrees). Conclusions When navigation is used only to guide the positioning of the cutting jig, errors may arise in the manual, non-navigated steps of the procedure. Our study showed increased cutting errors in the sagittal plane for both the femur and the tibia, and following implantation, the greatest error was seen in the sagittal alignment of the tibial component. Computer navigation should be used not only to guide the positioning of the cutting jig, but also to check the bone resection and implant position during TKA. Level of evidence IV.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 50 条
  • [41] Handheld computer-navigated constrained total knee arthroplasty for complex extra-articular deformities
    Pietsch, M.
    Hochegger, M.
    Djahani, O.
    Mlaker, G.
    Eder-Halbedl, M.
    Hofstadter, Th
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (12) : 2245 - 2254
  • [42] A meta-analysis of unicompartmental knee arthroplasty revised to total knee arthroplasty versus primary total knee arthroplasty
    Sun, Xuedong
    Su, Zheng
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2018, 13
  • [43] Do manual cutting guides for total knee arthroplasty introduce systematic error?
    Hohmann, Erik
    Tetsworth, Kevin
    INTERNATIONAL ORTHOPAEDICS, 2016, 40 (02) : 277 - 284
  • [44] Navigated vs Conventional Total Knee Arthroplasty: Is There a Difference in the Rate of Respiratory Complications and Transfusions?
    Liodakis, Emmanouil
    Antoniou, John
    Zukor, David J.
    Huk, Olga L.
    Epure, Laura M.
    Bergeron, Stephane G.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (10) : 2273 - 2277
  • [45] Computer-assisted versus non-navigated total knee arthroplasty: a review
    Martin, A.
    Widemschek, M.
    Von Strempel, A.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2009, 60 (03) : 249 - 254
  • [46] More accurate component alignment in navigated total knee arthroplasty has no clinical benefit at 5-year follow-up
    Hoppe, Sven
    Mainzer, Jens D.
    Frauchiger, Lars
    Ballmer, Peter M.
    Hess, Rolf
    Zumstein, Matthias A.
    ACTA ORTHOPAEDICA, 2012, 83 (06) : 629 - 633
  • [47] Clinical and radiographic outcomes of computer-navigated total knee arthroplasty are not adversely affected by body mass index
    Puah, Ken Lee
    Yeo, William
    Tan, Mann Hong
    JOURNAL OF ORTHOPAEDICS, 2020, 19 : 54 - 58
  • [48] Conventional or navigated total knee arthroplasty affects sagittal component alignment
    Sugama, Ryo
    Minoda, Yukihide
    Kobayashi, Akio
    Iwaki, Hiroyoshi
    Ikebuchi, Mitsuhiko
    Takaoka, Kunio
    Nakamura, Hiroaki
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (12) : 2454 - 2459
  • [49] Computer-navigated revision total knee arthroplasty for failed unicompartmental knee arthroplasty
    Chatain, F.
    Denjean, S.
    Delalande, J. -L.
    Chavane, H.
    Bejui-Hugues, J.
    Guyen, O.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2012, 98 (06) : 720 - 727
  • [50] A vascular complication in computer navigated total knee arthroplasty
    Sandesh Gulhane
    Ian Holloway
    Mathew Bartlett
    Indian Journal of Orthopaedics, 2013, 47 : 98 - 100