Navigated total knee arthroplasty: is it error-free?

被引:14
|
作者
Chua, Kerk Hsiang Zackary [1 ]
Chen, Yongsheng [1 ]
Lingaraj, Krishna [1 ]
机构
[1] Natl Univ Hlth Syst, Div Hip & Knee Surg, Singapore 119228, Singapore
关键词
Computer navigation; Total knee arthroplasty; Alignment deviation; Positioning error; Bone resection; Implant position; COMPUTER-ASSISTED NAVIGATION; CUTTING ERRORS; REPLACEMENT; ALIGNMENT; SURGERY; SYSTEM; IMPLANTATION; TKA;
D O I
10.1007/s00167-013-2641-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The aim of this study was to determine whether errors do occur in navigated total knee arthroplasty (TKAs) and to study whether errors in bone resection or implantation contribute to these errors. Methods A series of 20 TKAs was studied using computer navigation. The coronal and sagittal alignments of the femoral and tibial cutting guides, the coronal and sagittal alignments of the final tibial implant and the coronal alignment of the final femoral implant were compared with that of the respective bone resections. To determine the post-implantation mechanical alignment of the limb, the coronal alignment of the femoral and tibial implants was combined. Results The median deviation between the femoral cutting guide and bone resection was 0 degrees (range -0.5 degrees to +0.5 degrees) in the coronal plane and 1.0 degrees (range -2.0 degrees to +1.0 degrees) in the sagittal plane. The median deviation between the tibial cutting guide and bone resection was 0.5 degrees (range -1.0 degrees to +1.5 degrees) in the coronal plane and 1.0 degrees (range -1.0 degrees to +3.5 degrees) in the sagittal plane. The median deviation between the femoral bone resection and the final implant was 0.25 degrees (range -2.0 degrees to 3.0 degrees) in the coronal plane. The median deviation between the tibial bone resection and the final implant was 0.75 degrees (range -3.0 degrees to +1.5 degrees) in the coronal plane and 1.75 degrees (range -4.0 degrees to +2.0 degrees) in the sagittal plane. The median post-implantation mechanical alignment of the limb was 0.25 degrees (range -3.0 degrees to +2.0 degrees). Conclusions When navigation is used only to guide the positioning of the cutting jig, errors may arise in the manual, non-navigated steps of the procedure. Our study showed increased cutting errors in the sagittal plane for both the femur and the tibia, and following implantation, the greatest error was seen in the sagittal alignment of the tibial component. Computer navigation should be used not only to guide the positioning of the cutting jig, but also to check the bone resection and implant position during TKA. Level of evidence IV.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 50 条
  • [31] Total Knee Arthroplasty With a Computer-navigated Saw: A Pilot Study
    Garvin, Kevin L.
    Barrera, Andres
    Mahoney, Craig R.
    Hartman, Curtis W.
    Haider, Hani
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (01) : 155 - 161
  • [32] Multi-factorial analysis of time efficiency in total knee arthroplasty
    Stiehl, James B.
    Jackson, Scott
    Szabo, Aniko
    COMPUTER AIDED SURGERY, 2009, 14 (1-3) : 58 - 62
  • [33] New on minimally invasive computer-assisted navigated total knee arthroplasty (MIS CA-TKA)
    Biasca, N.
    Bungartz, M.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2010, 61 (03) : 203 - 219
  • [34] NAVIGATED TOTAL KNEE ARTHROPLASTY FOR OSTEOARTHRITIS WITH EXTRA-ARTICULAR DEFORMITY
    Tani, Izumi
    Nakano, Naoki
    Takayama, Koji
    Ishida, Kazunari
    Kuroda, Ryosuke
    Matsumoto, Tomoyuki
    ACTA ORTOPEDICA BRASILEIRA, 2018, 26 (03): : 170 - 174
  • [35] Computer-navigated minimally invasive total knee arthroplasty for patients with retained implants in the femur
    Lin, Sung-Yen
    Chen, Chung-Hwan
    Huang, Peng-Ju
    Fu, Yin-Chih
    Huang, Hsuan-Ti
    KAOHSIUNG JOURNAL OF MEDICAL SCIENCES, 2014, 30 (08) : 415 - 421
  • [36] Experiences with computer navigated total knee arthroplasty
    Zoltán Bejek
    László Sólyom
    Miklós Szendrõi
    International Orthopaedics, 2007, 31 : 617 - 622
  • [37] Experiences with computer navigated total knee arthroplasty
    Bejek, Zoltan
    Solyom, Laszlo
    Szendroi, Miklos
    INTERNATIONAL ORTHOPAEDICS, 2007, 31 (05) : 617 - 622
  • [38] Current Role of Computer Navigation in Total Knee Arthroplasty
    Jones, Christopher W.
    Jerabek, Seth A.
    JOURNAL OF ARTHROPLASTY, 2018, 33 (07) : 1989 - 1993
  • [39] Navigated total knee arthroplasty in a patient with severe diaphyseal deformities
    Patai, Janos
    Janositz, Gabor
    Mecs, Laszlo
    Toth, Kalman
    ACTA ORTHOPAEDICA BELGICA, 2007, 73 (04): : 536 - 540
  • [40] Systematic review of computer-navigated total knee arthroplasty
    Zamora, Luis A.
    Humphreys, Karen J.
    Watt, Amber M.
    Forel, Deanne
    Cameron, Alun L.