Navigated total knee arthroplasty: is it error-free?

被引:14
|
作者
Chua, Kerk Hsiang Zackary [1 ]
Chen, Yongsheng [1 ]
Lingaraj, Krishna [1 ]
机构
[1] Natl Univ Hlth Syst, Div Hip & Knee Surg, Singapore 119228, Singapore
关键词
Computer navigation; Total knee arthroplasty; Alignment deviation; Positioning error; Bone resection; Implant position; COMPUTER-ASSISTED NAVIGATION; CUTTING ERRORS; REPLACEMENT; ALIGNMENT; SURGERY; SYSTEM; IMPLANTATION; TKA;
D O I
10.1007/s00167-013-2641-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The aim of this study was to determine whether errors do occur in navigated total knee arthroplasty (TKAs) and to study whether errors in bone resection or implantation contribute to these errors. Methods A series of 20 TKAs was studied using computer navigation. The coronal and sagittal alignments of the femoral and tibial cutting guides, the coronal and sagittal alignments of the final tibial implant and the coronal alignment of the final femoral implant were compared with that of the respective bone resections. To determine the post-implantation mechanical alignment of the limb, the coronal alignment of the femoral and tibial implants was combined. Results The median deviation between the femoral cutting guide and bone resection was 0 degrees (range -0.5 degrees to +0.5 degrees) in the coronal plane and 1.0 degrees (range -2.0 degrees to +1.0 degrees) in the sagittal plane. The median deviation between the tibial cutting guide and bone resection was 0.5 degrees (range -1.0 degrees to +1.5 degrees) in the coronal plane and 1.0 degrees (range -1.0 degrees to +3.5 degrees) in the sagittal plane. The median deviation between the femoral bone resection and the final implant was 0.25 degrees (range -2.0 degrees to 3.0 degrees) in the coronal plane. The median deviation between the tibial bone resection and the final implant was 0.75 degrees (range -3.0 degrees to +1.5 degrees) in the coronal plane and 1.75 degrees (range -4.0 degrees to +2.0 degrees) in the sagittal plane. The median post-implantation mechanical alignment of the limb was 0.25 degrees (range -3.0 degrees to +2.0 degrees). Conclusions When navigation is used only to guide the positioning of the cutting jig, errors may arise in the manual, non-navigated steps of the procedure. Our study showed increased cutting errors in the sagittal plane for both the femur and the tibia, and following implantation, the greatest error was seen in the sagittal alignment of the tibial component. Computer navigation should be used not only to guide the positioning of the cutting jig, but also to check the bone resection and implant position during TKA. Level of evidence IV.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 50 条
  • [1] Navigated total knee arthroplasty: is it error-free?
    Kerk Hsiang Zackary Chua
    Yongsheng Chen
    Krishna Lingaraj
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 643 - 649
  • [2] Computer Navigated Total Knee Arthroplasty: Rates of Adoption and Early Complications
    Gholson, J. Joseph
    Duchman, Kyle R.
    Otero, Jesse E.
    Pugely, Andrew J.
    Gao, Yubo
    Callaghan, John J.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (07): : 2113 - 2119
  • [3] Navigated versus Conventional Total Knee Arthroplasty
    Moskal, Joseph T.
    Capps, Susan G.
    Mann, John W.
    Scanelli, John A.
    JOURNAL OF KNEE SURGERY, 2014, 27 (03) : 235 - 248
  • [4] Minimally Invasive Computer-Navigated Total Knee Arthroplasty
    Biasca, Nicola
    Schneider, Thomas-Oliver
    Bungartz, Matthias
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2009, 40 (04) : 537 - +
  • [5] Joint line changes after primary total knee arthroplasty: navigated versus non-navigated
    Jawhar, A.
    Shah, V.
    Sohoni, S.
    Scharf, H. P.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) : 2355 - 2362
  • [6] Precision of Ci-navigated extension and flexion gap balancing in total knee arthroplasty and analysis of potential predictive variables
    Fickert, S.
    Jawhar, A.
    Sunil, P.
    Scharf, H. -P
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2012, 132 (04) : 565 - 574
  • [7] Mechanical accuracy of navigated minimally invasive total knee arthroplasty (MIS TKA)
    Biasca, Nicola
    Wirth, Stephan
    Bungartz, Matthias
    KNEE, 2009, 16 (01): : 22 - 29
  • [8] The limits of precision in conventionally instrumented computer-navigated total knee arthroplasty
    da Assuncao, R. E.
    Hancock, N. J.
    Bruce, W. J. M.
    Walker, P.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (12) : 2528 - 2534
  • [9] Training benefits of-computer navigated total knee arthroplasty
    Love, Gavin J.
    Kinninmonth, Andrew W. G.
    KNEE, 2013, 20 (04): : 236 - 241
  • [10] Computer-navigated total knee arthroplasty in haemophilic arthropathy
    Cho, K. -Y.
    Kim, K. -I.
    Khurana, S.
    Cho, S. -W.
    Kang, D. -G.
    HAEMOPHILIA, 2013, 19 (02) : 259 - 266