METAL ARTIFACT REDUCTION IN CONE-BEAM EXTREMITY IMAGES USING GATED CONVOLUTIONS

被引:4
作者
Agrawal, Harshit [1 ,2 ]
Hietanen, Ari [1 ]
Sarkka, Simo [2 ]
机构
[1] Planmeca Oy, Asentajankatu 6, FI-00880 Helsinki, Finland
[2] Aalto Univ, Dept Elect Engn & Automat, FI-02150 Espoo, Finland
来源
2021 IEEE 18TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI) | 2021年
关键词
cone-beam computed tomography; deep learning; metal artifact reduction; orthopedic imaging; CT;
D O I
10.1109/ISBI48211.2021.9434163
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Quality of cone-beam computed tomography (CBCT) images are marred by artifacts in the presence of metallic implants. Metal artifact correction is a challenging problem in CBCT scanning especially for large metallic objects. The appearance of artifacts also change greatly with the body part being scanned. Metal artifacts are more pronounced in orthopedic imaging, when metals are in close proximity of other high density materials, such as bones. Recently introduced mask incorporating deep learning networks for metal inpainting showed improvements over classical methods in CBCT image quality. However, generalization of results for more than one body part is still not investigated. We investigate, the use of gated convolutions for mask guidance inpainting to improve the filling of the corrupt metal area in projection domain. The neural network was trained with eight clinical metal affected datasets by incorporating data augmentation techniques. In the end, we validate our method on six clinical datasets. Our method shows promising results both in projections and reconstructed images.
引用
收藏
页码:1087 / 1090
页数:4
相关论文
共 15 条
[1]  
Dauphin YN, 2017, PR MACH LEARN RES, V70
[2]  
Ghani MU., 2018, Electronic Imaging, V2018, P472, DOI 10.2352/ISSN.2470-1173.2018.15.COIMG-472
[3]   Deep learning methods to guide CT image reconstruction and reduce metal artifacts [J].
Gjesteby, Lars ;
Yang, Qingsong ;
Xi, Yan ;
Zhou, Ye ;
Zhang, Junping ;
Wang, Ge .
MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
[4]   Metal Artifact Reduction in CT: Where Are We After Four Decades? [J].
Gjesteby, Lars ;
De Man, Bruno ;
Jin, Yannan ;
Paganetti, Harald ;
Verburg, Joost ;
Giantsoudi, Drosoula ;
Wang, Ge .
IEEE ACCESS, 2016, 4 :5826-5849
[5]   Deep Learning Based Metal Inpainting in the Projection Domain: Initial Results [J].
Gottschalk, Tristan M. ;
Kreher, Bjorn W. ;
Kunze, Holger ;
Maier, Andreas .
MACHINE LEARNING FOR MEDICAL IMAGE RECONSTRUCTION, MLMIR 2019, 2019, 11905 :125-136
[6]   REDUCTION OF CT ARTIFACTS CAUSED BY METALLIC IMPLANTS [J].
KALENDER, WA ;
HEBEL, R ;
EBERSBERGER, J .
RADIOLOGY, 1987, 164 (02) :576-577
[7]   Generative Mask Pyramid Network for CT/CBCT Metal Artifact Reduction with Joint Projection-Sinogram Correction [J].
Liao, Haofu ;
Lin, Wei-An ;
Huo, Zhimin ;
Vogelsang, Levon ;
Sehnert, William J. ;
Zhou, S. Kevin ;
Luo, Jiebo .
MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2019, PT VI, 2019, 11769 :77-85
[8]  
Mangileva Daria, 2020, 2020 Ural Symposium on Biomedical Engineering, Radioelectronics and Information Technology (USBEREIT). Proceedings, P195, DOI 10.1109/USBEREIT48449.2020.9117624
[9]   Normalized metal artifact reduction (NMAR) in computed tomography [J].
Meyer, Esther ;
Raupach, Rainer ;
Lell, Michael ;
Schmidt, Bernhard ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2010, 37 (10) :5482-5493
[10]   CT sinogram-consistency learning for metal-induced beam hardening correction [J].
Park, Hyoung Suk ;
Lee, Sung Min ;
Kim, Hwa Pyung ;
Seo, Jin Keun ;
Chung, Yong Eun .
MEDICAL PHYSICS, 2018, 45 (12) :5376-5384